T2 Ankle Arthrodesis Nail - Stryker
T2 Ankle Arthrodesis Nail - Stryker
T2 Ankle Arthrodesis Nail - Stryker
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>T2</strong><br />
<strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
Operative Technique<br />
1
<strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong>ing System<br />
Contributing Surgeons<br />
Joseph D. DiCicco D.O.<br />
Chief Orthopedic Trauma<br />
Good Samaritan Hospital<br />
Chairman, Dept of Orthopaedic Surgery<br />
Grandview Hospital, Dayton, Ohio<br />
Clinical Professor Orthopaedic Surgery<br />
Ohio University & Wright State University<br />
Dayton, Ohio<br />
Thomas Mückley, M.D.<br />
Department of Trauma, Hand and<br />
Reconstructive Surgery<br />
University Hospital Jena<br />
Germany<br />
Anthony T. Sorkin, M.D.<br />
Rockford Orthopaedic Associates, LLP<br />
Clinical Instructor, Department of Surgery<br />
University of Illinois, College of Medicine<br />
Director, Orthopaedic Traumatology<br />
Rockford Memorial Hospital<br />
Rockford, Illinois, USA<br />
This publication sets forth detailed<br />
recommended procedures for using<br />
<strong>Stryker</strong> Osteosynthesis devices and<br />
instruments.<br />
It offers guidance that you should<br />
heed, but, as with any such technical<br />
guide, each surgeon must consider<br />
the particular needs of each patient<br />
and make appropriate adjustments<br />
when and as required. A workshop<br />
training is required prior to first<br />
surgery.<br />
See package insert (L22000007) for<br />
a complete list of potential adverse<br />
effects, contraindications, warnings<br />
and precautions. The surgeon must<br />
discuss all relevant risks, including the<br />
finite lifetime of the device, with the<br />
patient, when necessary.<br />
2<br />
Warning:<br />
All bone screws referenced in<br />
this document here are not<br />
approved for screw attachment or<br />
fixation to the posterior elements<br />
(pedicles) of the cervical, thoracic<br />
or lumbar spine.
Contents<br />
Page<br />
1. Introduction 4<br />
Implant Features 4<br />
Design Rational 4<br />
Technical Details 5<br />
Instrument Features 6<br />
2. Relative Indications & Contraindications 7<br />
3. Pre-operative Planning 7<br />
4. Locking Options 7<br />
5. Operative Technique 8<br />
Patient Positioning and Joint Surface Preparation 8<br />
Incision and Entry Point 9<br />
Reaming 11<br />
Target Device Assembly 12<br />
<strong>Nail</strong> Insertion 13<br />
Guided Locking via Targeting Device 14<br />
Approsition/Compression Locking Mode 14<br />
Static Locking Mode 20<br />
Freehand Proximal Locking of Long <strong>Nail</strong>s 23<br />
End Cap Insertion 24<br />
<strong>Nail</strong> Removal 25<br />
Case Resport 26<br />
References 27<br />
Ordering Information – Implants 28<br />
Ordering Information – Instruments 29<br />
3
Introduction<br />
Introduction<br />
The <strong>T2</strong> <strong>Nail</strong>ing System represents<br />
<strong>Stryker</strong>´s latest and most<br />
comprehensive development of the<br />
original intramedullary principles<br />
presented by Prof. Gerhard Küntscher<br />
in 1940. With a new generation of<br />
<strong>T2</strong> <strong>Nail</strong>s (e.g. <strong>T2</strong> Proximal Humeral<br />
<strong>Nail</strong>, <strong>T2</strong> Supracondylar <strong>Nail</strong>, <strong>T2</strong><br />
Recon <strong>Nail</strong>), the indications have<br />
extended from shaft fractures to<br />
fractures of the metaphyseal regions.<br />
In addition to the <strong>T2</strong> <strong>Nail</strong>ing systems<br />
for Femoral, Tibial, and Humeral<br />
fractures, <strong>Stryker</strong> Osteosynthesis has<br />
developed the <strong>T2</strong> Knee <strong>Arthrodesis</strong><br />
<strong>Nail</strong> and, more recently, the <strong>T2</strong> <strong>Ankle</strong><br />
<strong>Arthrodesis</strong> <strong>Nail</strong> to provide the<br />
option for tibiotalocalcaneal fusion<br />
with a retrograde intramedullary nail.<br />
The main advantages of the technique<br />
are limited soft tissue damage in the<br />
ankle area, high primary stability<br />
allowing early weight bearing,<br />
(3) as well as compression of the<br />
subtalar and tibiotalar joints (5).<br />
Severe arthrosis and deformity<br />
of the ankle and subtalar joints<br />
are debilitating problems that<br />
can be difficult to treat. The<br />
tibiotalocalcaneal fusion with a<br />
retrograde intramedullary nail can<br />
be considered a salvage procedure<br />
for severe arthrosis and deformity<br />
of the ankle and subtalar joints (1).<br />
<strong>Ankle</strong> arthrodesis is a challenging<br />
procedure due to poor host conditions<br />
(e.g. bad skin, deformity, avascular<br />
necrosis), inability to get adequate<br />
fixation for this slow healing process,<br />
and the inability to get adequate<br />
compression across the fusion.<br />
Like other nails in the <strong>T2</strong> family,<br />
the <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
enables the surgeon to create an<br />
inherently stable construct by<br />
providing internal compression to<br />
the fusion mass and distal crosslock<br />
holes within the nail to generate an<br />
intramedullary fixed-angle device.<br />
Performing an ankle arthrodesis can<br />
be technically demanding because of<br />
the shape and small size of the talus.<br />
Therefore, preoperative planning is<br />
an absolute necessity to determine<br />
placement and number of screws (2).<br />
The <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong>ing<br />
System is based on the established<br />
<strong>T2</strong> instrument platform and locking<br />
screws. It offers the advantages of a<br />
unique locking configuration allowing<br />
for tibiotalocalcaneal fixation.<br />
Implant Features<br />
The <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong> offers<br />
strong biomechanical* intramedullary<br />
stabilization using cannulated<br />
implants for the tibiotalocalcaneal<br />
fusion.<br />
All implants of the <strong>T2</strong> <strong>Ankle</strong><br />
<strong>Arthrodesis</strong> <strong>Nail</strong>ing System are<br />
made of Type II anodized titanium<br />
alloy (Ti6Al4V) for enhanced<br />
biomechanical and biomedical<br />
performance*.<br />
Standard 5mm cortical screws<br />
simplify the surgical procedure and<br />
offer the advantages of both reduced<br />
insertion torque and low profile<br />
heads*. Fully Threaded Locking<br />
Screws are available for standard<br />
locking procedures. Partially<br />
Threaded Locking Screws (Shaft<br />
Screws) are designed for use if<br />
apposition/compression is applied.<br />
The proximal locking configuration<br />
features a round and an oblong hole to<br />
allow for static or dynamic locking.<br />
Controlled apposition/compression<br />
up to 5mm can be applied at the<br />
tibio-talar joint by introducing a<br />
Compression Screw from the driving<br />
end of the nail against the 5mm<br />
Shaft Screw placed in the talus. The<br />
Compression Screw is cannulated<br />
therefore allowing nail insertion over<br />
the guide-wire with the compression<br />
screw pre-loaded.<br />
<strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong>s come in<br />
10, 11 and 12mm diameters and 150,<br />
200 and 300mm lengths. The driving<br />
end diameter is 12mm for all nails.<br />
* Data on file at <strong>Stryker</strong>:<br />
- Test Report No 130505CG1 Strength calculation<br />
(FEA) of 10mm <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
- White Paper: Ti6Al4V with Anodization Type II<br />
Biological Behavior and Biomechanical Effects<br />
Axel Baumann, Dipl.-Ing. DOT GmbH, Rostock,<br />
Germany, Nils Zander, Dipl.-Ing. <strong>Stryker</strong><br />
Trauma GmbH, Schönkirchen / Kiel, Germany<br />
- Test Report 080103HK1 <strong>T2</strong> / S2: Modified screw<br />
head design of cross screws Ø4 / 5mm<br />
Design Rationale<br />
The design of the <strong>T2</strong> <strong>Ankle</strong><br />
<strong>Arthrodesis</strong> <strong>Nail</strong> features a 5°<br />
lateral (valgus) bend providing an<br />
anatomical fit with better purchase<br />
through the calacaneal bone.<br />
Proximal locking is performed<br />
from medial to lateral in order to:<br />
• Avoid damage of muscular and<br />
neuro-vascular stuctures located<br />
on the lateral side of the tibia<br />
• Avoid interference with the fibula.<br />
The low profile design of <strong>T2</strong><br />
Locking Screws helps reduce<br />
prominence under the soft tissue<br />
on the medial cortex of the tibia.<br />
Two 5mm Fully Threaded Locking<br />
Screws can be placed in the<br />
calcaneus: the proximal one in a<br />
L/M direction through a threaded<br />
locking hole, and the distal one in<br />
P/A direction with a 10° angle from<br />
postero-medial to antero-lateral<br />
(Fig. 1). This design dictates the<br />
need for left and right nails.<br />
Internal compression techniques are<br />
preferred because of higher union<br />
rates, shorter fusion times, and fewer<br />
complications (2).<br />
End Caps in different sizes are<br />
available to provide the improved<br />
fit for every indication and lock<br />
down on the PA calcaneal screw.<br />
This feature creates a fixed angle<br />
between the nail and Locking<br />
Screw.<br />
See the detailed chart on the next<br />
page for the design specifications<br />
and size offerings of the implants.<br />
4
Features<br />
Technical Details - <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
Diameter<br />
Sizes<br />
10, 11 and 12mm<br />
(Left and Right)<br />
150, 200 and 300mm<br />
Note:<br />
Driving end diameter is 12mm<br />
for all nails.<br />
40mm<br />
15mm<br />
5mm<br />
5.0mm Partially Threaded<br />
Locking Screws (Shaft Screws)<br />
L=25mm – 120mm<br />
Note:<br />
Screw length is measured<br />
from top of head to tip.<br />
lateral<br />
Compression Screw<br />
medial<br />
5mm<br />
5.0mm Fully Threaded<br />
Locking Screws<br />
L=25mm – 120mm<br />
47mm<br />
19.5mm<br />
7.5mm<br />
Endcap<br />
anterior<br />
10°<br />
Compression Screw<br />
(cannulated)<br />
lateral<br />
medial<br />
End Caps<br />
Standard +5mm +10mm +15mm<br />
Fig 1<br />
View from distal<br />
5
Features<br />
Instrument Features<br />
The major advantage of the<br />
instrument system is the integration<br />
of a core instrument platform<br />
which can be used not only for the<br />
complete <strong>T2</strong> <strong>Nail</strong>ing System, but<br />
also represent the platform for all<br />
future <strong>Stryker</strong> Osteosynthesis nailing<br />
and further reduces complexity<br />
and inventory. The <strong>T2</strong> instrument<br />
platform offers advanced precision<br />
and usability, as well as ergonomically<br />
styled targeting devices. Except<br />
for the addition of a small number<br />
of dedicated instruments, the <strong>T2</strong><br />
Femoral Instrument platform is used<br />
for the <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong>.<br />
Targeting Device<br />
Apposition Ring<br />
Apposition Sleeve<br />
Dedicated instruments for the<br />
<strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
include the <strong>Ankle</strong> <strong>Arthrodesis</strong><br />
Targeting Device and a special<br />
Compression Screwdriver.<br />
Apposition Handle<br />
A pre-assembled Apposition Handle<br />
and Ring/Sleeve allow for applying<br />
additional external compression.<br />
As with all <strong>T2</strong> <strong>Nail</strong>ing Systems, the <strong>T2</strong><br />
<strong>Ankle</strong> <strong>Arthrodesis</strong> Targeting Device<br />
is made of carbon fiber and features<br />
a friction locking mechanism to lock<br />
the Tissue Protection Sleeves in place.<br />
Both proximal and distal holes of<br />
the 150mm and 200mm nails can be<br />
locked with the Targeting Device. For<br />
proximal locking of the 300mm nails,<br />
free hand technique must be applied.<br />
An additional Aiming Adapter can<br />
be attached to the Targeting Device<br />
to help achieve the appropriate<br />
rotation based on the position of<br />
the PA calcaneal screw aligned with<br />
the anatomic calcaneal body axis.<br />
Aiming Adapter<br />
6
Relative Indications & Contraindications<br />
Relative<br />
Indications and<br />
Contraindications<br />
The <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
may be used for:<br />
• Posttraumatic and primary<br />
Arthrosis<br />
• Neuromuscular deformity<br />
• Revision of Failed <strong>Ankle</strong><br />
<strong>Arthrodesis</strong><br />
• Failed Total <strong>Ankle</strong> Replacement<br />
• Avascular Necrosis of the<br />
Talus (requiring tibiocalcaneal<br />
arthrodesis)<br />
• Neuroarthropathy (Charcot)<br />
• Rheumatoid Arthritis with severe<br />
deformity<br />
• Osteoarthritis<br />
• Pseudarthrosis<br />
The <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong> should<br />
NOT be used if following conditions<br />
are present:<br />
• Tibial malalignment of > 10˚ in<br />
any plane<br />
• Severe vascular deficiency<br />
• Osteomyelitis or soft tissue<br />
infection<br />
Pre-operative<br />
Planning<br />
Preoperative clinical and radiological<br />
assessments are very important for the<br />
surgical outcome.<br />
• Clinical assessment comprises:<br />
evaluation of pain, quality and<br />
viability of soft tissue at the<br />
surgical site, neurological and<br />
vascular status.<br />
• Radiological assessment of the<br />
ankle includes: weight bearing<br />
anteroposterior and lateral views.<br />
A lateral hindfoot and Broden’s<br />
view are useful in evaluating the<br />
subtalar and transverse tarsal<br />
joints.<br />
• Appropriate implant size can be<br />
selected with the <strong>T2</strong> <strong>Ankle</strong> X-Ray<br />
Template (1806-3217).<br />
Locking Options<br />
Based on the clinical and radiological<br />
assessment, different locking<br />
options can be used to obtain<br />
the Tibiotalocalcaneal fusion:<br />
Apposition/Compression<br />
Locking Mode:<br />
- Tibio-talo internal compression<br />
with or without additional talocalcaneal<br />
external compression<br />
(static locking proximal)<br />
- Tibio-talo-calcaneal external<br />
compression (static locking<br />
proximal and distal)<br />
Static Locking Mode:<br />
- Talo-calcaneal static locking<br />
with proximal static locking<br />
Dynamic Locking Mode:<br />
- The proximal oblong hole allows<br />
for secondary dynamization<br />
Note:<br />
Please see package insert for<br />
warnings, precautions, adverse<br />
effects and other essential product<br />
information.<br />
7
Operative Technique<br />
Patient Positioning and Joint Surface Preparation<br />
Positioning<br />
Place the patient supine on a<br />
radiolucent table (Fig. 2). Care should<br />
be taken to assure neutral alignment<br />
of the knee and ankle. Prepare the<br />
entire foot and ankle and drape<br />
the limb free from above the knee<br />
to allow intraoperative assessment<br />
of lower limb alignment to avoid<br />
malalignment later in the procedure.<br />
The lower limbs should hang over the<br />
operating table about 15−20cm and<br />
the affected limb should be elevated<br />
by placing a bolster under the calf.<br />
This position will allow:<br />
- Easy exposure to the C-Arm<br />
for the X-Ray control<br />
- Proximal locking from medial side<br />
and distal locking from lateral side<br />
- Convenient access for posterior<br />
locking of the PA calcaneal screw.<br />
Fig 2<br />
Place the C-Arm on the opposite<br />
side and make sure that both lateral<br />
and anterioposterior views of the<br />
lower limb can be obtained.<br />
Exposure<br />
Make a 5−6 cm lateral incision in<br />
line with the distal lateral malleolus.<br />
To gain access to the tibiotalar joint,<br />
resection of the most distal portion<br />
of the fibula just above the tibiotalar<br />
joint might be required (Fig. 3).<br />
This allows adequate exposure<br />
of the tibiotalar joint and may<br />
provide source of bone graft<br />
if required. Resect the distal<br />
fibula at an angle superolateral<br />
to inferomedial to prevent<br />
prominence after healing (Fig. 4).<br />
Fig 3<br />
Any exposure (lateral or medial) may<br />
be used as long as it allows adequate<br />
access to the tibiotalar and subtalar<br />
joints. In complex cases, exposure<br />
of both sides might be required.<br />
Fig 4<br />
8
Operative Technique<br />
Joint Preparation<br />
Correction of any deformity should<br />
be addressed at this time. Generally,<br />
the contours of the tibiotalar and<br />
subtalar joints are maintained with<br />
denuding of any articular cartilage.<br />
Sometimes a “flat on flat” surface<br />
can be used depending on surgeon<br />
preference. Tibiotalar joint<br />
preparation may be aided by a<br />
laminar spreader or distraction of<br />
some kind. Care should be taken<br />
to avoid excessive bony resection<br />
which may later result in limb<br />
shortening or loss of talar fixation.<br />
<strong>Ankle</strong> positioning for fusion<br />
Several authors have attempted<br />
to define the optimal position for<br />
ankle arthrodesis without objective<br />
multiplanar radiographic analysis<br />
and consistent reference points (4).<br />
Position the foot with neutral ankle<br />
dorsi-plantar flexion, 5−10˚ external<br />
rotation in relation to the tibial crest<br />
and 5˚ of hindfoot valgus seems to be<br />
the most accepted<br />
(Fig. 5). An assistant should<br />
maintain this position for proper<br />
entry point determination.<br />
Incision and Entry Point<br />
Incision:<br />
After joint preparation and<br />
confirmatory X-Ray evaluation<br />
of fusion position, the incision<br />
point is determined as follows:<br />
Place a K-Wire (1806-0050S) on the<br />
plantar surface (1/3 lateral) and take<br />
an axial heel view to align it with<br />
the longitudinal axis of the calcaneus<br />
(Fig. 6). Mark this line with a pen on<br />
the skin. Next, place the K-Wire on<br />
the lateral side of the ankle aligning<br />
the wire along the tibial axis on a<br />
lateral fluoroscopic view. Mark this<br />
line with a pen on the skin extending<br />
the line onto the plantar surface<br />
(Fig. 7).<br />
The starting point for the incision<br />
is determined by the intersection of<br />
the two lines on the plantar surface.<br />
The line marked on the plantar<br />
surface which is aligned with the<br />
longitudinal axis of the calcaneus<br />
will also help align the Aiming<br />
Adapter after <strong>Nail</strong> insertion.<br />
A longitudinal incision approximately<br />
2−3cm should be made at this<br />
intersection. Careful dissection is<br />
then utilized to gain access to the<br />
plantar surface of calcaneus.<br />
Fig 6<br />
5° – 10°<br />
Fig 7<br />
Fig 5<br />
5°<br />
9
Operative Technique<br />
Entry Point<br />
The entry point is made under<br />
lateral and axial heel fluoroscopy<br />
control (Fig. 8) by using one of the<br />
following options:<br />
- A center-tipped Drill<br />
Ø4.2×340mm (1806-4260S).<br />
- A Stepped Reamer, Ø8/12mm<br />
(1806-2013), over a Ø3×285mm<br />
(1806-0050) K-Wire.<br />
The Wire should be inserted to<br />
the level of the superior aspect of<br />
the talar cut or prepared surface.<br />
Once this position has been verified<br />
as center/center in the talus, the<br />
Stepped reamer is inserted over the<br />
wire.<br />
It is recommended in this case to<br />
use the Protection Sleeve Retrograde<br />
(703165).<br />
Note:<br />
Do not use bent K-Wires.<br />
The axial heel view can help center<br />
and assure good position within the<br />
calcaneal body.<br />
Stop the Drill or Stepped Reamer<br />
after passing through the tibial<br />
articular surface gaining access into<br />
the tibial canal.<br />
Fig 8<br />
10
Operative Technique<br />
Reaming<br />
Insert the Ø3×800mm Ball Tipped<br />
Guide Wire (1806-0080S) with the<br />
Guide Wire Handle (1806-1095 and<br />
1806-1096) through the talocalcaneal<br />
and tibiotalar joints. Reaming is then<br />
performed with the Bixcut Reamers<br />
in 0.5mm increments until cortical<br />
contact is made within the tibia. For<br />
easier nail insertion, the medullary<br />
canal should be reamed 0.5−1.0mm<br />
more than the nail diameter selected<br />
(Fig. 9).<br />
The Ball Tip at the end of the Guide<br />
Wire will stop the Bixcut Reamer<br />
(Fig. 10).<br />
Prior to reaming, it is important to<br />
check the centered intramedullary<br />
position of the Guide Wire with<br />
image intensifier.<br />
Prior to nail insertion, the<br />
Ø3×800mm Ball Tip Guide Wire<br />
must be exchanged for a Ø3×800mm<br />
Smooth Tip Guide Wire (1806-<br />
0090S).<br />
Fig 9<br />
Use the Teflon Tube (1806-0073S)<br />
to facilitate guide wire exchange.<br />
Fig 10<br />
11
Operative Technique<br />
Target Device Assembly<br />
• Pre-load the Compression Screw<br />
(1818-0001S).<br />
Use the Compression Screwdriver<br />
(1806-3210) to insert the<br />
Compression Screw into the nail (Fig.<br />
11a).<br />
Make sure the screw is set between<br />
the round and the oblong hole.<br />
Fig 11a<br />
Prior to nail insertion, the Ball Tip<br />
Guide Wire must be exchanged for a<br />
Smooth Tip Guide Wire.<br />
The pre-loaded Compression Screw is<br />
cannulated but does not allow the ball<br />
tip to pass through.<br />
• Assemble the Apposition Handle<br />
(1806-3215) onto the <strong>Nail</strong> Adapter<br />
(1806-3211). Turn the Apposition<br />
Handle until the end of the threads<br />
in order not to influence the insertion<br />
depth of the nail (Fig. 11b).<br />
• Attach the selected nail to the<br />
<strong>Nail</strong> Adapter (Fig. 11c) until it´s 3<br />
connection teeth engage into the<br />
corresponding slots of the <strong>Nail</strong>.<br />
The <strong>Nail</strong> Holding Screw (1806-<br />
3203) is placed through the <strong>Nail</strong><br />
Adapter and tightened securely<br />
with the Insertion Wrench (1806-<br />
0135) and Wrench 8/10mm (1806-<br />
0130) to avoid loosening during<br />
<strong>Nail</strong> insertion. Engravings on the<br />
<strong>Nail</strong> Adapter will indicate lateral<br />
direction.<br />
Fig 11b<br />
Fig 11c<br />
• Insert the Target Arm (1806-3212)<br />
over the <strong>Nail</strong> Adapter and lock it<br />
in the “Lateral Locking” position.<br />
Attach the Aiming Adapter (1806-<br />
3216) and secure the whole assembly<br />
by tightening the Nut (1806-3213)<br />
(Fig. 12).<br />
Prior to nail insertion please check<br />
correct alignment of the Targeting<br />
Device by inserting a Ø4.2 × 340mm<br />
Drill (1806-4260S) through the<br />
assembled Tissue Protection (1806-<br />
0185) and Drill Sleeve, Long, (1806-<br />
0215) placed into the Targeting Arm<br />
and targeting all “Lateral Locking”<br />
holes of the implant.<br />
Note:<br />
If the Apposition Sleeve (1806-<br />
3214) is to be used, slide it<br />
over the nail and <strong>Nail</strong> Adapter<br />
prior to nail insertion.<br />
Note:<br />
The Aiming Adapter should<br />
be attached only when the<br />
Target Arm is mounted on the<br />
<strong>Nail</strong> Adapter in the “Lateral<br />
Locking” position. Check<br />
alignment of the P/A calcaneal<br />
hole by passing a K-Wire<br />
through the Aiming Adapter.<br />
Apposition<br />
Sleeve<br />
Nut<br />
Fig 12<br />
12
Operative Technique<br />
<strong>Nail</strong> Insertion<br />
Insert the nail over the Smooth Tip<br />
Guide Wire (Fig. 13) to the desired<br />
depth.<br />
A chamfer is located on the medial<br />
side of the nail driving end to<br />
avoid soft tissue impingement after<br />
insertion.<br />
Verify correct position of the nail<br />
by checking the correct depth and<br />
rotation.<br />
Depth of insertion is determined by<br />
CORRECT PLACEMENT OF THE<br />
DISTAL OBLONG HOLE IN THE<br />
CENTER OF THE TALAR BODY.<br />
This should be approximately the<br />
mid-talar region to ensure satisfactory<br />
purchase of the locking screw (Fig.<br />
13a).<br />
Two circumferential grooves are<br />
located on the insertion post at 2mm<br />
and 7mm from the driving end of the<br />
nail (Fig. 13b). Depth of insertion<br />
may be visualized with the aid of<br />
fluoroscopy.<br />
Additionally, the 3×285mm K-Wire<br />
can be inserted through the Targeting<br />
Device to identify the junction of the<br />
nail and insertion post (Fig. 13).<br />
Fig 13a<br />
2mm<br />
7mm<br />
Fig 13<br />
Rotational alignment is determined<br />
by a K-Wire placed into the Aiming<br />
Adapter. This indicates the position<br />
of the P/A calcaneal screw and aligns<br />
the screw with the anatomic calcaneal<br />
body axis (Fig. 14).<br />
Fig 13b<br />
Correct position is achieved when the<br />
K-Wire is in line with the vertical line<br />
marked on the plantar surface (used<br />
for determining the entry point) (Fig.<br />
14a).<br />
The K-Wire may be inserted 1cm<br />
into the calcaneus to help maintain<br />
position.<br />
Fig 14<br />
Remove the guide wire and proceed<br />
with locking screw placement.<br />
Fig 14a<br />
13
Operative Technique<br />
Guided Locking via Target Device<br />
Apposition/Compression Locking<br />
Mode<br />
The <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
provides the option to achieve active<br />
mechanical apposition/compression.<br />
Note:<br />
Proximal static locking with two<br />
Fully Threaded Locking Screws<br />
must be performed prior to<br />
applying active, controlled tibiotalar<br />
apposition/compression.<br />
Fig 15a<br />
Step 1:<br />
Guided Dynamic Locking<br />
of the Talar Screw<br />
If clinical and radiological assessment<br />
allow for applying tibio-talar<br />
compression, a 5mm Shaft Screw<br />
should be placed in the Dynamic<br />
position of the oblong hole. This<br />
will allow for a maximum of 5mm<br />
of active, controlled apposition/<br />
compression. Make sure the Target<br />
Arm is locked in the “Lateral<br />
Locking” position to place the screw<br />
from the lateral side of the talus.<br />
• Insert the Tissue Protection Sleeve,<br />
Long, (1806-0185) together with<br />
the Drill Sleeve, Long, (1806-0215)<br />
and the Trocar, Long, (1806-0315)<br />
into the “Talus Dyn./Compr.”<br />
hole of the Targeting Arm by<br />
pressing the Safety Clip (Fig. 15a).<br />
This mechanism will keep the<br />
sleeve in place and prevent it from<br />
falling out. It will also prevent<br />
the sleeve from sliding during<br />
screw measurement. To release<br />
the Tissue Protection Sleeve, the<br />
Safety Clip must be pressed again.<br />
• Advance the assembly through<br />
the skin incision that was used<br />
for joint preparation until it<br />
is in contact with the lateral<br />
cortex of the Talus (Fig. 15).<br />
Fig 15<br />
14
Operative Technique<br />
• The Trocar is removed, with the<br />
Tissue Protection Sleeve and the<br />
Drill Sleeve remaining in position<br />
(Fig. 16).<br />
• To ensure accurate drilling and<br />
determination of the screw<br />
length, use the centered tipped<br />
4.2×340mm calibrated Drill<br />
(1806-4260S). After drilling both<br />
cortices, the screw length may be<br />
read directly from the calibrated<br />
Drill at the end of the Drill Sleeve<br />
(Fig. 16a).<br />
• Next, drill the near cortex only,<br />
with the Ø5×230mm Drill (1806-<br />
5000S).<br />
• If measurement with the Screw<br />
Gauge, Long, (1806- 0325) is<br />
preferred, first remove the Drill<br />
Sleeve and read the screw length<br />
directly at the end of the Tissue<br />
Protection Sleeve.<br />
The position of the tip of the Drill as<br />
it relates to the far cortex is equal to<br />
where the tip of the screw will end.<br />
Therefore, if the tip of the Drill is<br />
3mm beyond the far cortex, the tip of<br />
the screw will also be 3mm beyond.<br />
The Screw Gauge is calibrated so that<br />
when the bend at the end is pulled<br />
back flush with the far cortex, the<br />
screw tip will end 3mm beyond the<br />
far cortex.<br />
Note:<br />
Make sure the Tissue Protection<br />
Sleeve/Drill Sleeve Assembly is seated<br />
on bone prior to selecting final screw<br />
length. (according to this picture)<br />
Fig 16a<br />
Fig 17a<br />
Fig 16<br />
• Remove the Drill Sleeve and insert<br />
the appropriate Shaft Screw length<br />
through the Tissue Protection<br />
Sleeve using the Screwdriver, Long,<br />
(1806-0232) (Fig. 17). The screw is<br />
advanced through both cortices.<br />
The screw is near its proper seated<br />
position when the groove around<br />
the shaft of the screwdriver is<br />
approaching the end of the Tissue<br />
Protection Sleeve (see Fig. 17a).<br />
• Remove the Tissue Protection<br />
Sleeve and proceed with proximal<br />
locking.<br />
Note:<br />
The Adapter and K-Wire must<br />
be removed before turning the<br />
Target Arm to the medial side for<br />
proximal locking.<br />
15<br />
Fig 17
Operative Technique<br />
Step 2:<br />
Guided Locking of the<br />
Proximal Screws<br />
Note:<br />
Guided Locking of the Proximal<br />
Screws must be performed with<br />
the Target Arm locked in the<br />
“Medial Locking” position.<br />
Do not attempt to use the Target<br />
Arm in the “Lateral Locking”<br />
position for proximal locking as<br />
this will lead to miss-drilling.<br />
The 300mm <strong>Nail</strong>s can be locked<br />
proximally only with the<br />
free-hand technique.<br />
• Release the Nut and turn the<br />
Target Arm around the <strong>Nail</strong><br />
Adapter until it can be locked in<br />
the “Medial Locking” position<br />
(Fig. 18).<br />
Before locking the proximal screws,<br />
check with the image intensifier<br />
the gap between the tibial and talar<br />
surface. If this is more than 5mm, try<br />
to reduce the gap by applying gentle<br />
pressure on the <strong>Nail</strong> Adapter.<br />
Fig 18<br />
• Insert the Tissue Protection<br />
Sleeve, Long, together with the<br />
Drill Sleeve, Long, and the Trocar,<br />
Long, into the appropriate hole<br />
for locking the static proximal<br />
hole of the selected <strong>Nail</strong> length<br />
(150mm or 200mm are marked on<br />
the Target Arm).<br />
• Make a small skin incision in<br />
front of the Trocar and push<br />
the assembly until the Tissue<br />
Protection Sleeve is in contact with<br />
the medial cortex of the tibia (Fig.<br />
19)<br />
Before starting to drill for the first<br />
proximal locking screw, check correct<br />
rotational position for the fusion; an<br />
imaginary sagital line drawn down<br />
from the tibia tuberosity, along the<br />
tibial crest, should align with the<br />
second ray of the foot (Fig. 19).<br />
Fig 19<br />
16
Operative Technique<br />
• The Trocar is removed, with<br />
the Tissue Protection Sleeve<br />
and the Drill Sleeve remaining<br />
in position (Fig. 20).<br />
• To ensure accurate drilling,<br />
it is recommended to use the<br />
Ø4.2×250, Drill oblique (1806-<br />
8018) to open the first cortex.<br />
• Use the centered tipped<br />
4.2×340mm calibrated Drill<br />
(1806-4260S). After drilling<br />
both cortices, the screw length<br />
may be read directly from the<br />
calibrated Drill at the end of<br />
the Drill Sleeve (Fig. 20a).<br />
The position of the tip of the Drill, as<br />
it relates to the far cortex, is equal to<br />
where the tip of the screw will end.<br />
Fig 20a<br />
• When the Drill Sleeve is removed,<br />
the correct Fully Threaded<br />
Locking Screw is inserted through<br />
the Tissue Protection Sleeve using<br />
the Screwdriver, Long. The screw<br />
is advanced through both cortices.<br />
The screw is near its proper seating<br />
position when the groove around<br />
the shaft of the screwdriver is<br />
approaching the end of the Tissue<br />
Protection Sleeve (see Fig. 21).<br />
Fig 20<br />
The countersink (1806-2015) can be<br />
used through the Tissue Protection<br />
Sleeve to help sinking the proximal<br />
screw head. If this is used, undersize<br />
the screw length by 5mm.<br />
Fig 21<br />
17
Operative Technique<br />
• Repeat the locking procedure for<br />
the second Locking Screw<br />
(Fig. 22). This one can only be<br />
placed in the dynamic position of<br />
the proximal oblong hole.<br />
• Remove the Tissue Protection<br />
Sleeve and proceed with the tibiotalar<br />
compression.<br />
Step 3:<br />
Tibio-talar apposition/compression<br />
• Insert the Compression<br />
Screwdriver (1806-3210) through<br />
<strong>Nail</strong> Holding screw until the tip of<br />
the Screwdriver engages into the<br />
Compression Screw.<br />
• Start turning the Compression<br />
Screwdriver clockwise. As the<br />
Compression Screw is advanced<br />
against the 5.0mm Partially<br />
Threaded Locking Screw (Shaft<br />
Screw), it draws the talus towards<br />
the proximal tibial segment,<br />
employing active apposition/<br />
compression (Fig. 23).<br />
Fig 22<br />
Note:<br />
Caution should be taken when<br />
actively compressing across<br />
the tibiotalar fusion site in<br />
osteoporotic bone to avoid<br />
iatrogenic talus fractures due<br />
to overcompression. Tibio-talar<br />
active compression must be<br />
carried out under fluoroscopy<br />
control.<br />
Before proceeding with the guided<br />
locking of the Lateral Calcaneal<br />
Screw, external talo-calcaneal<br />
apposition/compression can be<br />
applied, if needed.<br />
Fig 23<br />
18
Operative Technique<br />
Step 4 (optional):<br />
Talo-calcaneal external compression<br />
External compression is achieved by<br />
inserting the Apposition Ring (1806-<br />
3204) over the Apposition Handle<br />
(1806-3215). This will protect the soft<br />
tissues by applying compression forces<br />
on a larger surface.<br />
Turn the Apposition Handle<br />
Clockwise until the Apposition Ring<br />
is in contact with the soft tissues.<br />
Continue turning the Apposition<br />
Handle to apply talo-calcaneal<br />
apposition/compression (Fig. 24).<br />
Alternatively, the Apposition Sleeve<br />
(1806-3214) can be used to apply<br />
external compression directly on the<br />
calcaneal cortex in case of poor soft<br />
tissue condition.<br />
Fig 24<br />
Note:<br />
The Apposition Sleeve must be<br />
inserted over the <strong>Nail</strong> Adapter<br />
before nail insertion.<br />
Step 5:<br />
Guided Locking of the<br />
Lateral Calcaneal Screw<br />
• Release the Nut and turn the Target<br />
Arm around the <strong>Nail</strong> Adapter<br />
until it can be locked again in the<br />
“Lateral Locking” position (Fig. 25).<br />
• Insert the Tissue Protection Sleeve,<br />
Long, together with the Drill<br />
Sleeve, Long, and the Trocar, Long,<br />
into the “Calcaneus” hole of the<br />
Target Arm by pressing the Safety<br />
Clip.<br />
• Make a small skin incision in<br />
front of the Trocar and push<br />
the assembly until the Tissue<br />
Protection Sleeve is in contact with<br />
the lateral calcaneal cortex.<br />
• The Trocar is removed, with the<br />
Tissue Protection Sleeve and the<br />
Drill Sleeve remaining in position.<br />
• Use the centered tipped<br />
4.2×340mm calibrated Drill<br />
(1806-4260S). After drilling both<br />
cortices, the screw length may be<br />
read directly from the calibrated<br />
Drill at the end of the Drill Sleeve.<br />
• When the Drill Sleeve is removed,<br />
the correct Fully Threaded Locking<br />
Screw is inserted through the<br />
Tissue Protection Sleeve using the<br />
Screwdriver, Long (Fig. 26).<br />
Step 6:<br />
Guided Locking of the<br />
Posterior Calcaneal Screw<br />
• Release the external compression<br />
• Release the Nut and turn the Target<br />
Arm around the <strong>Nail</strong> Adapter until<br />
it can be locked in the “Posterior<br />
Locking” position.<br />
• Insert the Tissue Protection<br />
Sleeve, Long, together with the<br />
Drill Sleeve, Long, and the Trocar,<br />
Long, into the Calcaneus hole of<br />
the Targeting Arm by pressing the<br />
Safety Clip (Fig. 27).<br />
• Repeat the locking procedure as<br />
described for the Lateral Calcaneal<br />
Locking.<br />
The countersink (1806-2015) can be<br />
used through the Tissue Protection<br />
Sleeve to assist in sinking the P/A<br />
Calcaneus screw head. If this is used,<br />
undersize the screw length by 5mm.<br />
19<br />
Fig 25<br />
Fig 26<br />
Fig 27
Operative Technique<br />
Static Locking Mode<br />
Step 1:<br />
Guided Static Locking of the<br />
Talar Screw<br />
If clinical and radiological<br />
assessment does not allow for<br />
applying tibio-talar compression, a<br />
5mm Shaft Screw should be placed<br />
in the Static position of the oblong<br />
hole. Make sure the Target Arm<br />
is locked in the “Lateral Locking”<br />
position to place the screw from<br />
the lateral side of the talus.<br />
• Insert the Tissue Protection<br />
Sleeve, Long, together with<br />
the Drill Sleeve, Long, and the<br />
Trocar, Long, into the “Talus<br />
Static” hole of the Target Arm by<br />
pressing the Safety Clip (Fig. 28).<br />
• Follow the same locking<br />
procedure as described on page<br />
14 and 15 for the insertion of the<br />
Talar Screw in dynamic position.<br />
Fig 28<br />
20
Operative Technique<br />
• After the Talar Screw is inserted,<br />
remove the Tissue Protection<br />
Sleeve and proceed with advancing<br />
the compression screw against<br />
the Talar Screw (Fig. 29).<br />
• Insert the Compression<br />
Screwdriver (1806-3210) through<br />
<strong>Nail</strong> Holding screw until the<br />
tip of the Screwdriver engages<br />
into the Compression Screw.<br />
• Start turning the Compression<br />
Screwdriver clockwise. The<br />
Compression Screw will advance<br />
until it locks down onto the<br />
Talar Screw providing axial<br />
stability of the construct.<br />
Caution:<br />
The coupling of Elastosil Handles<br />
contains a mechanism with<br />
one or multiple ball bearings.<br />
In case of applied axial stress<br />
on the Elastosil handle, those<br />
components are pressed into the<br />
surrounding cylinder resulting in<br />
a complete blockage of the device<br />
and possible bending. To avoid<br />
intra-operative complications and<br />
secure long-term functionality, we<br />
mandate that Elastosil handles be<br />
used only for their intended use.<br />
DO NOT HIT hit on them.<br />
Fig 29<br />
Step 2:<br />
Guided Locking of the<br />
Lateral Calcaneal Screw<br />
• After locking the Talar screw<br />
in place with the compression<br />
screw, leave the Target Arm in<br />
the “Lateral Locking” position<br />
and proceed with the Lateral<br />
Calcaneal screw insertion.<br />
• Follow the locking procedure<br />
as described on page 19 for the<br />
Lateral Calcaneal Screw.<br />
Fig 30<br />
21
Operative Technique<br />
Step 3:<br />
Guided Locking of the<br />
Proximal Screws<br />
Note:<br />
Guided Locking of the Proximal<br />
Screws must be performed with<br />
the Target Arm locked in the<br />
“Medial Locking” position (Fig.<br />
30).<br />
The 300mm <strong>Nail</strong>s can be locked<br />
proximally only with the freehand<br />
technique.<br />
• Proceed with the locking<br />
procedure as described on page<br />
16 and 17 for the Apposition/<br />
Compression Locking Mode.<br />
Fig 31<br />
Step 4:<br />
Guided Locking of the<br />
Posterior Calcaneal Screw<br />
• Release the Nut and turn the<br />
Target Arm around the <strong>Nail</strong><br />
Adapter until it can be locked in<br />
the “Posterior Locking” position<br />
(Fig. 32).<br />
• Insert the Tissue Protection Sleeve,<br />
Long, together with the Drill<br />
Sleeve, Long, and the Trocar, Long,<br />
into the Calcaneus hole of the<br />
Target Arm by pressing the Safety<br />
Clip.<br />
• Repeat the locking procedure as<br />
described on page 19.<br />
Fig 32<br />
22
Operative Technique<br />
Freehand Proximal Locking of Long <strong>Nail</strong>s<br />
The freehand technique is used to<br />
insert locking screws into both M/L<br />
proximal holes of the <strong>T2</strong> <strong>Ankle</strong><br />
<strong>Arthrodesis</strong> Long <strong>Nail</strong>s (300mm).<br />
Multiple locking techniques and<br />
radiolucent drill devices are available<br />
for freehand locking. The critical<br />
step with any freehand locking<br />
technique, proximal or distal,<br />
is to visualize a perfectly round<br />
locking hole or perfecty oblong<br />
locking hole with the C-Arm.<br />
• Hold the center-tipped<br />
Ø4.2×130mm Drill (1806-4280)<br />
at an oblique angle to the center<br />
of the locking hole (Fig. 33).<br />
Upon fluoroscopic verification,<br />
the Drill is placed perpendicular<br />
to the nail and drilled through<br />
the medial and lateral cortex<br />
of the tibia. Confirm that the<br />
Drill passes through the hole<br />
in the nail in both the A/P<br />
and M/L fluoroscopy views.<br />
• After drilling both cortices,<br />
the screw length may be read<br />
directly from Screw Scale,<br />
Short, (1806-0360) at the green<br />
ring on the center-tipped<br />
Ø4.2×130mm Drill(Fig. 34).<br />
Alternatively, the Screw Gauge<br />
(1806-0480) can be used<br />
instead of the Screw Scale to<br />
determine the screw length.<br />
• Routine locking screw<br />
insertion is employed with the<br />
assembled Screwdriver Shaft,<br />
Short, (1806-0294) and the<br />
Teardrop Handle (702429).<br />
• Repeat the locking procedure<br />
to insert the second<br />
proximal locking screw.<br />
Fig 33<br />
Fig 34<br />
23
Operative Technique<br />
End Cap Insertion<br />
The End Cap (1826-0003S) can be<br />
inserted<br />
• Either through the <strong>Nail</strong> Adapter,<br />
with the Screwdriver, Long, (after<br />
removal of the <strong>Nail</strong> Holding Screw)<br />
or<br />
• With the Screwdriver Shaft, Short,<br />
and the Teardrop Handle (Fig. 35),<br />
after removal of the Target Device<br />
(Fig. 35).<br />
Note:<br />
This is the same End Cap used<br />
for the <strong>T2</strong> SCN. The End Cap will<br />
lock on the distal P/A calcaneal<br />
screw providing additional axial<br />
stability.<br />
Fig 35<br />
Extension End Caps of +5, +10 and<br />
+15mm are also available to adjust<br />
nail length and lock down on the<br />
distal P/A calcaneal screw. These End<br />
Caps cannot be inserted through<br />
the <strong>Nail</strong> Adapter due to the larger<br />
diameter of the head.<br />
Extension End Caps are not<br />
cannulated.<br />
+5mm +10mm +15mm<br />
24
Operative Technique<br />
<strong>Nail</strong> Removal<br />
<strong>Nail</strong> removal is an elective procedure.<br />
• If used, remove first the End Cap<br />
and the most distal Screw with the<br />
Screwdriver, Long (Fig. 36).<br />
• Remove the Lateral Calcaneal<br />
Screw.<br />
• Release the Compression Screw to<br />
allow removal of the Talar screw.<br />
• Insert the Universal Rod into the<br />
driving end of the nail.<br />
• Remove all other Locking Screws<br />
and use the Slottet Hammer<br />
(1806-0170) to extract the nail in a<br />
controlled manner (Fig. 37).<br />
Fig 36<br />
Fig 37<br />
25
Operative Technique<br />
Case Reports<br />
Provided by Anthony T. Sorkin, M.D.<br />
The patient is a 58 year old male<br />
with a pilon fracture 30 years<br />
prior to her first visit to the<br />
office, complaining of severe<br />
pain. Advanced post-traumatic<br />
arthritis of both sub-talar and<br />
tibio-talar joints can be seen on the<br />
preoperative X-Rays (Fig. 38 and<br />
Fig. 39).<br />
Incisions were made both laterally<br />
and medially to fully debride the<br />
tibiotalar and subtalar joints.<br />
The <strong>T2</strong> <strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong><br />
used as definitive treatment.<br />
Tibio-talar compression was<br />
applied with a preloaded<br />
compression screw. Additional<br />
talo-calcaneal compression<br />
was achieved with the external<br />
compression device. Axially<br />
stable locking of the transverse<br />
calcaneal screw (passing through<br />
a threaded hole of the <strong>T2</strong><br />
AAN) and P/A calcaneal screw<br />
(locked in place by the end cap)<br />
providing increased stability.<br />
Postoperative X-Rays (Fig.<br />
40a, 40b, 40c) show perfect<br />
alignment sustained by the<br />
5° valgus of the <strong>T2</strong> AAN.<br />
Patient weight bearing at 8 weeks without any pain.<br />
Fig 38<br />
Fig 39<br />
Fig 40a<br />
Fig 40b<br />
Fig 40c<br />
26
References<br />
1. Tibiotalocalcaneal fusion with a retrograde intramedullary nail:<br />
clinical and functional outcomes.<br />
Millett PJ, O‘Malley MJ, Tolo ET, Gallina J, Fealy S, Helfet DL.<br />
Harvard Medical School, Brigham & Women‘s Hospital, Boston,<br />
Massachusetts, USA. Am J Orthop. 2002 Sep; 31(9):531-6.<br />
2. The effect of bone quality on the stability of ankle arthrodesis.<br />
A finite element study.<br />
Alonso-Vazquez A, Lauge-Pedersen H, Lidgren L, Taylor M.<br />
Bioengineering Sciences Research Group, School of Engineering Sciences,<br />
University of Southampton, Southampton SO17 1BJ, UK. Foot <strong>Ankle</strong> Int.<br />
2004 Nov; 25(11):840-50.<br />
3. <strong>Ankle</strong> arthrodesis with intramedullary compression nailing<br />
Muckley T, Schutz T, Srivastava S, Goebel M, Gonschorek O, Bühren V.<br />
Berufsgenossenschaftliche Unfallklinik, Murnau. Unfallchirurg.<br />
2003 Sep; 106(9):732-40.<br />
4. Realignment arthrodesis of the rearfoot and ankle:<br />
a comprehensive evaluation.<br />
Mendicino RW, Lamm BM, Catanzariti AR, Statler TK, Paley D.<br />
Division of Foot and <strong>Ankle</strong> Surgery, Western Pennsylvania Hospital,<br />
Pittsburgh, PA 15224, USA.<br />
5. Primary stiffness of different arthrodesis techniques for the upper ankle<br />
joint considering the compression nail: a biomechanical study<br />
Muckley T, Eichhorn S, Steinhauser E, von Oldenburg G, Speitling A,<br />
Hofmann G.O., J. Ortop. Trauma 2004 #18,<br />
6. Intramedullary nailing in tibiocalcaneal arthrodesis<br />
Goebel M, Muckley T, Gerdesmeyer L, Militz M, Bühren V.<br />
Unfallchirurg. 2003 Aug; 106(8):633-41<br />
27
Ordering Information – Implants<br />
<strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong>, left<br />
REF Diameter Length<br />
mm<br />
mm<br />
1818-1015S 10 150<br />
1818-1020S 10 200<br />
1818-1030S 10 300<br />
1818-1115S 11 150<br />
1818-1120S 11 200<br />
1818-1130S 11 300<br />
1818-1215S 12 150<br />
1818-1220S 12 200<br />
1818-1230S 12 300<br />
<strong>Ankle</strong> <strong>Arthrodesis</strong> <strong>Nail</strong>, right<br />
REF Diameter Length<br />
mm<br />
mm<br />
1819-1015S 10 150<br />
1819-1020S 10 200<br />
1819-1030S 10 300<br />
1819-1115S 11 150<br />
1819-1120S 11 200<br />
1819-1130S 11 300<br />
1819-1215S 12 150<br />
1819-1220S 12 200<br />
1819-1230S 12 300<br />
Compression Screw (cannulated)<br />
REF Diameter Length<br />
mm<br />
mm<br />
1818-0001S 8.0 14.5<br />
End Caps<br />
REF Diameter Length<br />
mm<br />
mm<br />
1826-0003S 8.0 4.0<br />
1818-0005S 12.0 +5mm<br />
1818-0010S 12.0 +10mm<br />
1818-0015S 12.0 +15mm<br />
+5mm<br />
+10mm<br />
+15mm<br />
28
Ordering Information – Instruments<br />
REF<br />
Description<br />
Standard Instruments<br />
1806-3217 <strong>T2</strong> AAN X-Ray Template<br />
1806-0050 K-Wire 3×285mm (2)<br />
703165 Protection Sleeve Retrograde<br />
1806-0073S Teflon Tube, sterile<br />
1806-0080S Guide Wire, Ball Tip, Ø3×800mm, Sterile<br />
1806-0090S Guide Wire, Smooth Tip, Ø3×800mm, Sterile<br />
1806-1095 Guide Wire Handle<br />
1806-1096 Guide Wire Handle Chuck<br />
1806-0110 Universal Rod<br />
1806-0130 Wrench, 8mm / 10mm<br />
1806-0135 Insertion Wrench, 10mm<br />
1806-0150 Strike Plate<br />
1806-0170 Slotted Hammer<br />
1806-0185 Tissue Protection Sleeve, Long<br />
1806-0215 Drill Sleeve, Long<br />
1806-0232 Screwdriver, Long<br />
1806-0294 Screwdriver Shaft, 3,5×85 mm, conical tip<br />
702429 Teardrop Handle<br />
1806-0315 Trocar, Long<br />
1806-0325 Screw Gauge, Long<br />
1806-0360 Screw scale short (only for 300mm nails)<br />
1806-0480 Screw Gauge, Femur (only for 300mm nails)<br />
1806-2013 Stepped Reamer, Ø8/12mm<br />
1806-2015 Countersink<br />
29
Ordering Information – Instruments<br />
REF<br />
Description<br />
Standard Instruments<br />
1806-3213 Nut, <strong>T2</strong> AAN<br />
1806-3218 Nut Tightening Pin<br />
1806-3211 <strong>Nail</strong> Adapter<br />
1806-3212 Targeting Arm<br />
1806-3203 <strong>Nail</strong> Holding Screw<br />
1806-3204 Apposition-ring<br />
1806-3214 Apposition Sleeve<br />
1806-3215 Apposition-handle<br />
1806-3216 Aiming Adapter<br />
1806-3210 Compression Screwdriver<br />
1806-4260S Drill Ø4.2×340mm, AO, Sterile (2)<br />
1806-4280S Drill Ø4.2×130mm, Sterile (2) (only for 300mm nails)<br />
1806-5000S Drill Ø5×230mm, AO, Sterile (2)<br />
1806-8018 Drill oblique, Ø4.2×250mm<br />
1806-9263 <strong>T2</strong> AAN Dedicated Instrument Tray<br />
1806-9265 <strong>T2</strong> AAN Add-on Instrument Tray<br />
30<br />
Note:<br />
Outside of the U. S., Locking<br />
Screws and other specific<br />
products may be ordered nonsterile<br />
without the “S” at the end<br />
of the corresponding Reference<br />
Number.
Notes<br />
31
<strong>Stryker</strong> Trauma GmbH<br />
Prof.-Küntscher-Straße 1–5<br />
D - 24232 Schönkirchen<br />
Germany<br />
www.osteosynthesis.stryker.com<br />
This document is intended solely for the use of healthcare professionals. A surgeon must always rely on his or her own<br />
professional clinical judgment when deciding whether to use a particular product when treating a particular patient.<br />
<strong>Stryker</strong> does not dispense medical advice and recommends that surgeons be trained in the use of any particular product<br />
before using it in surgery. The information presented in this brochure is intended to demonstrate a <strong>Stryker</strong> product.<br />
Always refer to the package insert, product label and/or user instructions including the instructions for Cleaning and<br />
Sterilization (if applicable) before using any <strong>Stryker</strong> products. Products may not be available in all markets. Product<br />
availability is subject to the regulatory or medical practices that govern individual markets. Please contact your <strong>Stryker</strong><br />
representative if you have questions about the availability of <strong>Stryker</strong> products in your area.<br />
<strong>Stryker</strong> Corporation or its divisions or other corporate affiliated entities own, use or have applied for the following<br />
trademarks or service marks: <strong>Stryker</strong>, <strong>T2</strong>, Elastosil, Bixcut.<br />
All other trademarks are trademarks of their respective owners or holders.<br />
The products listed above are CE marked.<br />
Literature Number: B1000044<br />
LOT C3009<br />
Copyright © 2009 <strong>Stryker</strong>