<strong>Operative</strong> <strong>Technique</strong><br />
Note:<br />
If very hard bone is encountered,<br />
it may be necessary to alternate<br />
between using the cannulated<br />
rasp and the intramedullary<br />
reamer (Ref XFR003001;<br />
Maximum speed limit of the<br />
centromedullar reamer: 100<br />
000 min- 1 ) or guide wire (Ref<br />
XVIPR02001) in order to<br />
successfully create the housing.<br />
It is advisable to take an X-Ray<br />
to check the progress of the rasp.<br />
Fitting an XS stem<br />
For an XS-sized stem (usually the 5th<br />
digital ray) the housing is prepared<br />
directly using the XS proximal rasp<br />
(Ref XRP006001).<br />
Withdraw the wire, then rasp using<br />
the proximal progressive rasp<br />
(Ref XRP007001) until successfully<br />
obtaining a properly adapted seat.<br />