Tactys Operative Technique - Stryker
Tactys Operative Technique - Stryker
Tactys Operative Technique - Stryker
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Operative</strong> <strong>Technique</strong><br />
Seat Preparation<br />
Fitting a stem S, M or L<br />
Start to prepare the seat by using<br />
the distal cannulated rasp<br />
(Ref XRP005002).<br />
Note:<br />
If very hard bone is encountered,<br />
it may be necessary to alternate<br />
between using the cannulated<br />
rasp and the intramedullary<br />
reamer (Ref XFR003001;<br />
Maximum speed limit of the<br />
centromedullar reamer: 100<br />
000 min- 1 ) or guide wire (Ref<br />
XVIPR02001) in order to<br />
successfully create the housing.<br />
It is advisable to take an X-Ray to<br />
check the progress of the rasp.<br />
Fitting an XS stem<br />
In the case of an XS-sized stem<br />
(5th digital ray mainly) the seat<br />
is prepared directly using the<br />
“XS distal rasp”.<br />
Withdraw the wire, then rasp using<br />
the distal progressive rasp (Ref<br />
XRP007002), until successfully<br />
obtaining a properly adapted seat.<br />
13