Description<br />

Voltage ranges for the canned motors of the RUVAC WS/WSU pumps<br />

Pump model<br />

Nominal power / Nominal current at<br />

50 Hz<br />

200 V<br />

KW<br />

A<br />

230 V<br />

KW<br />

A<br />

400 V<br />

KW<br />

A<br />

Nominal power / Nominal current at<br />

60 Hz<br />

200-208 V<br />

KW A<br />

265 V<br />

KW<br />

A<br />

460 V<br />

KW<br />

A<br />

WS 151/251<br />

WSU 151/251<br />

WS/WSU PFPE 151/251<br />

0,9 4,8<br />

1,1 5,5<br />

1,1 3,2<br />

0,9 4,8<br />

1,4 5,5<br />

1,4 3,2<br />

WS 501<br />

WSU 501<br />

WS/WSU PFPE 501<br />

1,6 7,8<br />

2,2 9<br />

2,2 5,4<br />

1,6 7,8<br />

2,4 10<br />

2,4 5,8<br />

WS 1001<br />

WSU 1001<br />

WS PFPE 1001<br />

3 13<br />

4 15,7<br />

4 9,1<br />

3 13<br />

4,4 14,7<br />

4,4 8,5<br />

WS 2001<br />

WSU 2001<br />

WS PFPE 2001<br />

5 21<br />

7,5 26<br />

7,5 15<br />

5 21<br />

8,5 26<br />

8,5 15<br />

Caution<br />

The motor overload protection switch must be set to the nominal current stated in the table in each case.<br />

The separate fan current may only be operated within the permissible voltage range (230 V, 50 Hz and<br />

265 V, 60 Hz). In the case of deviating voltages you must use a transformer.<br />

with<br />

with<br />

with<br />

with<br />

with<br />

S v<br />

= pumping speed of the backing pump<br />

70<br />

(mbar)<br />

Fig. 6 Pumping speed characteristics for operation at 50 Hz<br />

Fig. 7 Power consumption of the<br />

RUVAC WS/WSU<br />

8<br />

GA 03.108/7.02 - 11/96