Scanning Electron Microscopy
Scanning Electron Microscopy
Scanning Electron Microscopy
Transform your PDFs into Flipbooks and boost your revenue!
Leverage SEO-optimized Flipbooks, powerful backlinks, and multimedia content to professionally showcase your products and significantly increase your reach.
<strong>Scanning</strong> <strong>Electron</strong> <strong>Microscopy</strong><br />
Field<br />
emitting tip<br />
Grid<br />
≤2kV<br />
≤100kV<br />
Anode<br />
ZEISS SUPRA<br />
Variable Pressure FESEM<br />
Analytical<br />
Workshop 2012<br />
Dr Heath Bagshaw – CMA<br />
bagshawh@tcd.ie
Why use an SEM?<br />
Analytical<br />
Workshop 2012<br />
Fig 1. Examples of features resolvable using different imaging techniques
Improving Resolution<br />
• Firstly, the wavelength of the imaging source is important.<br />
In an optical microscope white light is used (λ – 380-700-nm)<br />
• In an <strong>Electron</strong> Microscope the imaging source is a beam of electrons which has a<br />
shorter wavelength (λ ~0.0025nm at 200kV) .<br />
• This is approximately five times smaller than visible light and twice as small as a<br />
typical atom – this is why electrons can ‘see’ atoms but white light can’t :-<br />
Analytical<br />
Workshop 2012<br />
‘the analysis probe must be smaller than the feature being analysed’<br />
• The wavelength of electrons is dependent on the accelerating voltage, i.e.:-<br />
kV<br />
Wavelength λ (pm)<br />
20 8.588<br />
100 3.702<br />
200 2.508<br />
300 1.968<br />
• The higher the accelerating voltage the shorter the wavelength.
The Parts of an EM<br />
• <strong>Electron</strong> Microscopes (EMs) are similar in operation to optical microscopes except<br />
that they use a focused beam of electrons instead of light to "image" the specimen and<br />
gain information about its structure and composition.<br />
• There are four major regions in an <strong>Electron</strong> Microscope:-<br />
Analytical<br />
Workshop 2012<br />
(1) A stream of electrons is formed (by the electron source/gun) and<br />
accelerated toward the specimen using a positive electrical potential<br />
(2) This stream is confined and focused using metal apertures and magnetic<br />
lenses into a thin, focused, monochromatic beam.<br />
(3) This beam is focused onto the sample using a magnetic lens. In an SEM<br />
the beam is then also scanned across the surface of the sample.<br />
(4) Interactions occur inside the irradiated sample, affecting the electron<br />
beam which are detected and transformed into an image or signal.<br />
• The above happens in all EMs regardless of type.
Layout of a Generic SEM<br />
Gun<br />
1<br />
Aperture<br />
Holder<br />
2<br />
Deflection<br />
coils<br />
3<br />
Objective Lens<br />
4<br />
Analytical<br />
Workshop 2012
<strong>Electron</strong> Gun<br />
• There are two main types of gun – Thermionic and Field Emission (FEG).<br />
• Thermionic gun :- Simplistically, a material (often a piece of twisted tungsten) is heated<br />
to a high temperature so that it will emit electrons.<br />
• Can also use LaB 6 crystal grown to a tip – gives a brighter beam than W for same kV.<br />
Analytical<br />
Workshop 2012<br />
Tungsten filament Tungsten filament assembly LaB 6 filament tip
Thermionic Gun<br />
Filament<br />
Wehnelt<br />
cylinder<br />
-<br />
10-10000kV<br />
+<br />
Anode<br />
earth<br />
•Filament is heated and begins to produce electrons.<br />
•<strong>Electron</strong>s leave the filament tip with a negative potential so accelerate towards the<br />
earthed anode and into the microscope column.<br />
•A small negative bias on the Wehnelt then focuses the beam to a crossover which acts as<br />
the electron source.<br />
Analytical<br />
Workshop 2012
Field Emission Gun (FEG)<br />
FEG source – W tip<br />
• A very strong magnetic field (~10 9 Vm -1 ) draws electrons from a very fine metallic tip<br />
(usually W).<br />
•An extraction voltage of around 2kV is applied to the first anode to create an intense<br />
electric field to allow electrons to escape from the tip.<br />
•The second anode is then used to accelerate the electrons into the microscope at the<br />
required energy.<br />
•Combination of the two anodes focuses the beam into a crossover creating a fine<br />
beam source.
Comparison of Sources<br />
•W filaments are very simple and inexpensive.<br />
•LaB 6 filaments give greater brightness than W (approximately X10), but cost more.<br />
•FEG’s give much more brightness than thermionic systems.<br />
•FEG’s give a more monochromatic electron source and finer probe (i.e. better resolution).<br />
•Temperatures used are much lower than for thermionic sources (particularly cold cathode<br />
FEG’s).<br />
•FEG’s require better vacuum systems and are more expensive.<br />
Analytical<br />
Workshop 2012<br />
Comparison of the three types of source<br />
operating at 100kV
Focusing the Beam<br />
• After the beam is formed it is focused by a condenser lens system to form a ‘probe’.<br />
• The lenses are electromagnetic – the focal length changes as current in the coil changes.<br />
• After focusing, the beam is passed through an aperture which excludes electrons which are<br />
not on the optical axis – improving resolution.<br />
• Inconsistencies in the beam are corrected by stigmators and the beam focused onto the<br />
sample.<br />
A typical Electro<br />
Magnetic Lens<br />
Analytical<br />
Workshop 2012
<strong>Scanning</strong> the Beam\Beam Interactions<br />
• Deflector coils move the beam back and forth over the sample and the signal<br />
generated from each area is collected simultaneously, building up the final image<br />
shown on the monitor.<br />
• Many signals are generated at the surface of the<br />
sample and many different forms of analysis may be<br />
performed.<br />
• The ‘interaction volume’ is the area of the sample<br />
excited by the electron beam to produce a signal.<br />
Incident <strong>Electron</strong>s<br />
(<strong>Electron</strong> Probe)<br />
Auger <strong>Electron</strong>s<br />
Secondary <strong>Electron</strong>s<br />
Backscattered <strong>Electron</strong>s<br />
Continuum X-Rays<br />
Fluorescence X-Rays<br />
Characteristic X-Rays<br />
Signals generated in the interaction volume<br />
Analytical<br />
Workshop 2012
Interaction Volume<br />
•The ‘interaction volume’ is the area of the sample excited by the electron beam to<br />
produce a signal.<br />
•The penetration of the electron beam into the sample is affected by the accelerating<br />
voltage used, the higher the kV the greater the penetration.<br />
•The effective interaction volume can be calculated using the electron range, R:-<br />
1.67<br />
0.0276<br />
A E0<br />
R =<br />
( µ m)<br />
0.89<br />
ρ Z<br />
Where A is the atomic weight (g/mole), Z is the atomic number, ρ is the density (in g/cm 3 ) and E o is the energy of<br />
the primary electron beam (in kV).<br />
Take the example of iron:<br />
A=55.85, Z=26, r=7.87 g/cm 3<br />
Accelerating voltage (kV) Primary <strong>Electron</strong> Range (µm)<br />
30 3.1<br />
15 0.99<br />
5 0.16<br />
1 0.01 (10nm!)<br />
Signals generated in the interaction volume<br />
Analytical<br />
Workshop 2012
Signal Detection<br />
• The Everhard – Thornley Detector (ETD) is the most common detector used to detect<br />
secondary electrons to image surface topography.<br />
• <strong>Electron</strong>s are attracted to a +ve charge on a grid in front of the detector. The captured<br />
electrons are then amplified by a photo-multiplier before being digitised and sent to a screen.<br />
• The signal detected is transferred to a viewing screen as the beam is scanned building up the<br />
image.<br />
Everhard Thornley Detector<br />
<strong>Scanning</strong> to produce an image<br />
Analytical<br />
Workshop 2012
Biological Samples<br />
•Biological samples are not conductive and are particularly vulnerable to beam damage and<br />
other heating effects when examined in an <strong>Electron</strong> Microscope.<br />
•The level of exposure is referred to as ‘<strong>Electron</strong> Dose’ and is a measure of the number of<br />
electrons per unit area (e/nm 2 ).<br />
•Samples are either stained with conductive materials (e.g. OsO 4 ) or coated with Au or C.<br />
•Samples are viewed under vacuum, so they are dried to remove all water.<br />
a) b)<br />
•a) SEM image of Pneumonia, and b) SEM image of Diatom (Pictures from University of Iowa)<br />
•Preparing the samples ‘fixes’, and alters them – need a way to look at samples whilst they are<br />
still ‘wet’.<br />
Analytical<br />
Workshop 2012
Variable Pressure SEM<br />
•Localised charging is removed by the presence of gas in the sample chamber, effectively allowing the<br />
examination of non conductive samples<br />
•In Low vacuum mode the chamber is isolated from the high vacuum system(A) and is instead pumped<br />
by the additional rotary pump system(B).<br />
Analytical<br />
Workshop 2012
Variable Pressure SEM (2)<br />
• This allows analysis of ‘non conducting’ samples as charge is compensated by gas in the<br />
chamber.<br />
• Use of an SEM in VP mode does lead to some limitations in it’s operation:-<br />
• Because the vacuum is lower in a VPSEM chamber, some resolution of the instrument is<br />
lost due to scattering of the electron beam by the gas particles in the chamber.<br />
• In situ heating and or cooling (with the appropriate sample stage) is possible in VPSEM<br />
to allow direct observation of sample changes.<br />
• Compositional analysis is still possible.<br />
Analytical<br />
Workshop 2012
Compositional Analysis – Back Scattered Imaging<br />
• As mentioned previously, when the electron beam hits the sample a number of signals<br />
are generated. Secondary electrons are used for looking at surface detail (topography).<br />
• EM is also a very powerful technique for analysing composition and compositional<br />
distribution in a material\sample.<br />
B<br />
Analytical<br />
Workshop 2012<br />
Signals generated in the interaction volume<br />
• Back Scattered electrons are produced just below the surface of the sample (B) and are<br />
scattered more by heavier elements than by lighter elements.<br />
• The backscattered coefficient, η = (Z-1.5)/6 So, as Z increases, so does the degree of<br />
backscatter.
Back Scattered <strong>Electron</strong>s<br />
0.5 Backscattered<br />
electrons<br />
0.4<br />
<strong>Electron</strong> yield<br />
0.3<br />
0.2<br />
0.1<br />
Secondary electrons<br />
0 20 40 60 80<br />
Atomic number (Z)<br />
<strong>Electron</strong> yield (i.e. intensity) as a function of atomic number for backscattered and Secondary<br />
electrons.<br />
• Back Scattered electron have approximately the same energy as the primary electron<br />
beam and are therefore easy to detect - simply by a semiconductor placed above the<br />
sample :-<br />
Analytical<br />
Workshop 2012<br />
Schematic of a backscattered electron detector.
BackScattered Imaging<br />
• Back scattered electrons are deflected more by heavier atoms leading to a brighter<br />
contrast in BEI images – the lighter the region the heavier the element present.<br />
a)<br />
b)<br />
White<br />
region<br />
Dark<br />
region<br />
Grey<br />
region<br />
• a)Secondary image of a cement showing surface morphology<br />
• b)Backscattered image of same area showing compositional inhomogeneity<br />
• Three distinct regions in b), EDS analysis can then be used to find the different<br />
compositions of the these regions.<br />
Analytical<br />
Workshop 2012
Example EDS of a Cement<br />
• There were 3 distinct regions in the Backscattered Image<br />
‘Light’ region is made up predominantly of<br />
Fe. (i.e. the heaviest element)<br />
‘Grey’ region is made up predominantly of<br />
Ca.<br />
‘Dark’ region is made up predominantly of Si<br />
and Al. (i.e. the lightest elements)<br />
Analytical<br />
Workshop 2012
Examples Images<br />
Imaged using ET Detector<br />
Imaged using InLens Detector<br />
Analytical<br />
Workshop 2012<br />
Low kV image of platelets
Examples Images<br />
SE image of nano tubes<br />
BE image of nano tubes<br />
Analytical<br />
Workshop 2012<br />
SE Image of nano structure
Conclusions<br />
•<strong>Scanning</strong> <strong>Electron</strong> Microscopes (SEM’s) are very useful tools for looking at a range of<br />
samples\materials.<br />
•Surface detail, homogeneity and elemental composition can be determined in one<br />
experiment on the same sample.<br />
•Newer Variable Pressure SEM’s allow the imaging of non conducting samples.<br />
•ESEM’s, with cold stages and other peripherals allow imaging at 100% relative humidity<br />
allowing imaging of ‘wet’ samples<br />
•<strong>Electron</strong> <strong>Microscopy</strong> based analysis when used with other analysis techniques can<br />
assist in complete characterisation\identification of materials.<br />
•<strong>Electron</strong> Microscopes provide a very powerful analysis tool in both Materials and<br />
biological fields.<br />
Analytical<br />
Workshop 2012