chiller ® or Unichiller ®<br />

Unichiller ® with CC-Pilot ®<br />

The tower models offer power with small footprints.<br />

These models are fitted with the exchangeable Compatible<br />

Control CC-Pilot, and are used in both research<br />

and production, the range of cooling powers<br />

available is from 1,6 to 100 kW.<br />

Features<br />

K Space saving tower design: small dimensions,<br />

high powers<br />

K Robust stainless steel construction<br />

K Reliable continuous operation with alarm and<br />

early warning functions<br />

K CC-Pilot with <strong>Plug</strong> & <strong>Play</strong> Technology<br />

K Splash protection of display and function keys<br />

K Large and bright TFT display<br />

K Digital level indicator<br />

K Simple to fill and drain<br />

K Simple to use EASY Control, with rotary input<br />

and function keys<br />

K All functions listed alphabetically<br />

K RS232 interface and connection for optional<br />

Com.G@te (NAMUR Standard)<br />

K Strong pumps for systems with large pressure<br />

drops<br />

K High flow rates for optimal heat transfer<br />

K External Pt100 sensor via 4-wire Lemo<br />

connector<br />

K 5-Point calibration<br />

K IP-class to IEC EN 60529: 21<br />

K Options (factory fitted)<br />

– Heater and adjustable over temperature<br />

protection<br />

– VPC (Variable Pressure Control) with steplessly<br />

variable bypass and pressure sensor<br />

– Winter operation for use in low temperature<br />

external environments<br />

– Weather protection<br />

– Tropical versions for environmental temperature<br />

up to 45 °C<br />

– Stronger pumps<br />

Introduction | Unichiller ® • Minichiller ® • RotaCool ®<br />

Unichiller ® with MPC ® -Controller<br />

Compact, value-for-money units; Unichillers are available with<br />

cooling powers up to 2,5 kW for applications in the lab. The<br />

models ranging from the Minichiller to UC025w are suitable for<br />

on or under the lab bench.<br />

Options for Heating and<br />

Temperatures to +100 °C<br />

With the exception of the two models UC006Tw-NR and<br />

UC009Tw-NR all units can be factory fitted with optional heating<br />

and independent over-temperature protection. Increasing the<br />

maximum working temperature to 100 °C and the temperature<br />

stability to ± 0,2 K. The new construction allows constant operation<br />

in ambient temperatures up to 40 °C. The water-cooled models<br />

are especially quiet and require little cooling water even at<br />

full cooling power. Despite the increasing costs of water the ROI<br />

is exceptionally short. All models with maximum pump pressure<br />

of 2,5 bar have an adjustable bypass and a pressure gauge.<br />

Pump Options<br />

We offer alternative pumps for applications with higher pressure<br />

drops.<br />

Models with Pump A are only suitable for externally closed systems.<br />

Models with Pumps B, C and D can also be configured for<br />

operation with external open systems.<br />

The quoted technical data is indicative, and will vary slightly<br />

depending on the model.<br />

Pump Pump duty (l/min) at (bar)<br />

0,2 0,5 1,0 2,0 2,5 3,0 4,0<br />

A 20 10 – – – – –<br />

B* – 21 17 6 – – –<br />

B1* – 40 35 28 24 20 10<br />

C3** – – 65 60 45 40 25<br />

C4** – – 80 75 75 60 30<br />

D3** – – 170 160 140 130 90<br />

* Stronger pump options available from UC ® 017T / UC ® 017Tw<br />

** Pump suitable for UC ® 055T / UC ® 055Tw and above<br />

Note: The fitting of a more powerful pump reduces the cooling<br />

power by the following amounts: Pump B1 for B = 150 W; C4 for<br />

C3 = 400 W; D3 for C3 = 750 W; D3 for C4 = 350 W.<br />

In a few cases it may be necessary to use larger machine housing.<br />

45<br />