Flexible solutions for calibration<br />

in production<br />

High Precision Calibration<br />

Calibration is a comparison between a measurement<br />

system and a reference or standard. During the comparison<br />

it is established how large the deviation between<br />

the two values or if the value lies within the<br />

specified limits. Calibration is normally carried out in<br />

accordance with rigorous national or international<br />

standards.<br />

Meaningful and comparable measurements around<br />

the world require calibrated instruments. The quality<br />

of measurements is defined in terms of tolerance<br />

and repeatability, and is only achievable with the use<br />

of calibrated measurement devices or by adjusting<br />

sensors.<br />

Calibration baths are used in quality management<br />

departments of industry and research. The modular<br />

concept based on the combination of a calibration<br />

bath with a Unistat, which dictates the temperature<br />

range and speed of temperature change. The stainless<br />

steel calibration bath is designed in a similar format<br />

to a calorimeter to ensure temperature homogeneity.<br />

Baths with a 118 mm diameter and depth of<br />

Models | Accelerated Forced ageing test • Calibration<br />

384 mm are offered for calibration of measurement<br />

and control sensors. The calibration space is freely accessible<br />

and symmetrical. The upper edge is designed<br />

to allow exact reading of the temperature measured<br />

by glass thermometers and also offers a tight seal for<br />

the customer specific bath lid. Special calibration<br />

software in the circulator and the self-optimising controller<br />

with TAC Technology mean short times between<br />

the different calibration temperatures. The calibration<br />

space of the baths can be customised to suit<br />

specific customer requirements.<br />

Advantages<br />

K Highest temperature stability up to ± 0,002 °C<br />

K Temperature homogeneity better than ± 0,01 °C<br />

K External overflow vessel<br />

K 5-point calibration of the control sensor<br />

See page 93 for the calibration inserts for our bath<br />

thermostats.<br />

Model Temperature Pump Dimensions Bath Cat.No. G Price<br />

Range Connection W x D x H Opening Depth Volume<br />

(°C) (mm) (mm) (mm) (ltr)<br />

Cal 700 -100...400 M30x1,5 300 (440*) x 300 x 566 Ø118 384 7,0 9623 3<br />

* with external overflow vessel (140 mm)<br />

Cal 700 – Calibration bath for<br />

measurement and control sensors<br />

81<br />