Probes for Coordinate Measuring Machines - Mitutoyo America ...
Probes for Coordinate Measuring Machines - Mitutoyo America ...
Probes for Coordinate Measuring Machines - Mitutoyo America ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
TP2-5W<br />
Compact Touch Trigger Probe<br />
• Compact touch trigger probes<br />
This touch trigger probe has an outside diameter as small as ø13 mm, which greatly<br />
contributes to probing complex portions of a workpiece. With the combined use of an<br />
appropriate probe extension it can probe even deeper locations.<br />
• Enhancing the setup and measurement efficiency through the automatic<br />
change of probe orientation<br />
Since the TP2-5W can be mounted on a probe head such as the PH10M/PH10MQ that<br />
automatically changes the probe orientation, it can drastically reduce the time required to<br />
prepare <strong>for</strong> measurement and <strong>for</strong> actual measurement in comparison to a conventionaltype<br />
scanning probe that has a fixed position downward. With the combined use of an<br />
appropriate probe extension it can probe even deeper locations (when it is mounted on<br />
the CNC coordinate measuring machine).<br />
TP2-5W Specifications<br />
TP2-5W<br />
<strong>Measuring</strong> direction<br />
Repeatability (2 )<br />
Directionality (XY: 2D)<br />
Directionality (XYZ: 3D)<br />
Required <strong>for</strong>ce to generate<br />
trigger signal<br />
Amount of over-travel<br />
Required <strong>for</strong>ce to achieve<br />
over-travel<br />
Stylus mounting method<br />
Mass of single unit<br />
Durability<br />
Probe head<br />
Applicable models<br />
XY<br />
Z<br />
XY<br />
Z<br />
XY<br />
Z<br />
±X, ±Y, +Z<br />
0.35µm or less<br />
±0.8µm (with 10mm stylus), ±2.5µm (with the 50mm stylus)<br />
±1µm (with 10mm stylus), ±4µm (with the 50mm stylus)<br />
0.07 to 0.15N (Variable. Standard: 0.07N)<br />
0.7N (when the X and Y axis are set to standard)<br />
XY±14º, Z+4mm (when the required <strong>for</strong>ce to generate trigger signal is 0.07N)<br />
±4mm (when the required <strong>for</strong>ce to generate trigger signal is 0.07N), +3mm (at 0.15N)<br />
0.2 to 0.4N<br />
4N (when the required <strong>for</strong>ce to generate trigger signal is 0.07N)<br />
M2 screw<br />
22g<br />
1,000,000 times<br />
Essential: PH10M/PH10MQ/MIH/PH1<br />
Manual/CNC coordinate measuring machines<br />
20