TP2-5W<br />

Compact Touch Trigger Probe<br />

• Compact touch trigger probes<br />

This touch trigger probe has an outside diameter as small as ø13 mm, which greatly<br />

contributes to probing complex portions of a workpiece. With the combined use of an<br />

appropriate probe extension it can probe even deeper locations.<br />

• Enhancing the setup and measurement efficiency through the automatic<br />

change of probe orientation<br />

Since the TP2-5W can be mounted on a probe head such as the PH10M/PH10MQ that<br />

automatically changes the probe orientation, it can drastically reduce the time required to<br />

prepare <strong>for</strong> measurement and <strong>for</strong> actual measurement in comparison to a conventionaltype<br />

scanning probe that has a fixed position downward. With the combined use of an<br />

appropriate probe extension it can probe even deeper locations (when it is mounted on<br />

the CNC coordinate measuring machine).<br />

TP2-5W Specifications<br />

TP2-5W<br />

<strong>Measuring</strong> direction<br />

Repeatability (2 )<br />

Directionality (XY: 2D)<br />

Directionality (XYZ: 3D)<br />

Required <strong>for</strong>ce to generate<br />

trigger signal<br />

Amount of over-travel<br />

Required <strong>for</strong>ce to achieve<br />

over-travel<br />

Stylus mounting method<br />

Mass of single unit<br />

Durability<br />

Probe head<br />

Applicable models<br />

XY<br />

Z<br />

XY<br />

Z<br />

XY<br />

Z<br />

±X, ±Y, +Z<br />

0.35µm or less<br />

±0.8µm (with 10mm stylus), ±2.5µm (with the 50mm stylus)<br />

±1µm (with 10mm stylus), ±4µm (with the 50mm stylus)<br />

0.07 to 0.15N (Variable. Standard: 0.07N)<br />

0.7N (when the X and Y axis are set to standard)<br />

XY±14º, Z+4mm (when the required <strong>for</strong>ce to generate trigger signal is 0.07N)<br />

±4mm (when the required <strong>for</strong>ce to generate trigger signal is 0.07N), +3mm (at 0.15N)<br />

0.2 to 0.4N<br />

4N (when the required <strong>for</strong>ce to generate trigger signal is 0.07N)<br />

M2 screw<br />

22g<br />

1,000,000 times<br />

Essential: PH10M/PH10MQ/MIH/PH1<br />

Manual/CNC coordinate measuring machines<br />