Environmental Conditions<br />
This unit, and operations, will comply with DIN EN 61010-1:2001, only when it is<br />
located <strong>in</strong> suitable environmental conditions.<br />
10<br />
- for <strong>in</strong>door use only;<br />
- <strong>in</strong>stallation site ≤ 2000 m altitude;<br />
- <strong>in</strong>stalled on a level, even, non flammable surface;<br />
- ma<strong>in</strong>ta<strong>in</strong> a clearance above and around the unit of 10 cm for water-cooled units,<br />
and 20cm for air-cooled units, to allow air to circulate around the unit;<br />
- for ambient temperature conditions please refer to the technical data sheet;<br />
rema<strong>in</strong><strong>in</strong>g with<strong>in</strong> these ambient conditions is imperative <strong>in</strong> ensur<strong>in</strong>g accurate<br />
operation;<br />
- maximum relative humidity of 80% up to 32°C, decreas<strong>in</strong>g l<strong>in</strong>early to 50%<br />
relative humidity at 40°C<br />
- use only as long a power cord as necessary;<br />
- the unit should be located so as not to restrict access to the ma<strong>in</strong>s power<br />
switch;<br />
- ma<strong>in</strong>s voltage should be ±10% of the rated value;<br />
- avoid voltage spikes;<br />
- transient voltage surges as they occur normally <strong>in</strong> the supply grid;<br />
- clean rat<strong>in</strong>g 2;<br />
- overvoltage category II