Coating Thickness Software for Eagle µ-EDXRF Systems
Coating Thickness Software for Eagle µ-EDXRF Systems
Coating Thickness Software for Eagle µ-EDXRF Systems
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Figure 13a. Default dialogue box <strong>for</strong> layer<br />
definition - Step #2<br />
Figure 13b. Example of completed Step #2<br />
dialogue box <strong>for</strong> a single layer<br />
If the appropriate scatter spectrum had not been identified in Step 1 of 3, this will be indicated by the traffic<br />
lights at the bottom left of the Step 2 of 3 window being in the red-alert state<br />
as in Figure 13.<br />
When the system’s scatter spectrum has not been measured and<br />
identified, the calibration will refer to a default spectrum supplied with the<br />
software.<br />
instead of<br />
The check box alongside the layer element (Fig 13 with detail to the<br />
left) allows a maximum measuring range to be set (i.e. limited). This<br />
is intended to improve calibration accuracy particularly in the case of<br />
multi-layer situations where the prior knowledge of individual<br />
maximum thicknesses simplifies the correction algorithms. With<br />
single layer systems it is generally useful to ensure that the measure<br />
range includes the thickest standard to be used in any ReCalib<br />
samples in order to ensure that its data is not ignored.<br />
However, the program will compute what it considers to be the maximum range<br />
when this box is left unchecked based on limits imposed by the counting statistics<br />
and other considerations. Such limits are defined in the “Preselections” region at<br />
bottom right of the Step 2 window. Counting statistics are based upon in<strong>for</strong>mation<br />
gained from the pure element intensity together with the measuring time (here<br />
Au(La) 1264cps – to be introduced at Step 3 of 3 - and 30sec). The [%]<br />
“preselection” value defines the maximum relative error to be tolerated in the<br />
thickness determination per unit change in intensity (i.e. layer/bulk – I/I o - intensity ratio). Increasing the<br />
value of allowable error increases the maximum thickness range. The default value is 5%.<br />
The drop-down selection box to the right of the Base graphic is used to define<br />
which Recalibration mode is to be used. The default is “standard”. This<br />
function will be discussed within the section “Monolayer Recalibration”.<br />
Clicking on the<br />
button opens the panel <strong>for</strong> the third and final stage.<br />
3.1 Procedures in FunMaster (ex. Au on Cu base) 22