Photometry<br />

Technical data<br />

Optics<br />

temperature-compensated LEDs and<br />

photosensor amplifier in water proof<br />

sample chamber<br />

Ranges 0 – 150 mg / l O 2 ±3.5% *)<br />

0 – 1500 mg / l O 2 ±3.5% *)<br />

0 –15000 mg / l O 2 ±3.5% *)<br />

Power<br />

supply<br />

Auto - OFF<br />

Display<br />

Storage<br />

Additional<br />

feature<br />

Calibration<br />

9 V powerpack battery<br />

providing 40 hours operation<br />

automatic switch-off<br />

backlit LCD (on keypress)<br />

internal ring memory for 16 data sets<br />

real time clock<br />

and date<br />

factory calibration and<br />

user calibration. Reset to factory<br />

calibration possible<br />

Dimensions<br />

190 x 110 x 55 mm<br />

(L x W x H)<br />

Weight<br />

0.4 kg (base unit)<br />

Environmental Temperature: 0 – 40 °C<br />

conditions rel. humidity: 30 – 90%<br />

(non condensing)<br />

CE- DIN EN 50081-1<br />

conformity VDE 0839 part 81-1: 1993-03<br />

DIN EN 50082-2<br />

VDE 0839 part 82-2: 1996-02<br />

* based on the use of potassium-hydrogenephthalate standards (DIN 38 409)<br />

COD VARIO tube tests<br />

The AQUALYTIC ® COD VARIO tube tests are available<br />

for the measuring ranges 0-150 mg/l O 2 , 0-1500 mg/l O 2 and<br />

0-15000 mg/l O 2 . Their chemical properties and a 16 mm tube<br />

diameter is suitable also for use with Hach photometers.<br />

Tube tests Quantity Order code<br />

0 - 150 mg/l O 2 (25 pc.) mercury free 420710<br />

(25 pc.) 420720<br />

(150 pc.) 420725<br />

0 - 1500 mg/l O 2 (25 pc.) mercury free 420711<br />

(150 pc.) mercury free 420716<br />

(25 pc.) 420721<br />

(150 pc.) 420726<br />

0 - 15000 mg/l O 2 (25 pc.) mercury free 420712<br />

(25 pc.) 420722<br />

(150 pc.) 420727<br />

Standard solutions<br />

Standard solutions are solutions with a defined concentration<br />

and are provided to check the operation methods and<br />

devices of the cuvette tests as well as the condition of optical<br />

filters and the instrument.<br />

Standard solutions Quantity Order code<br />

100 mg/l COD 30 ml 420803<br />

500 mg/l COD 30 ml 420804<br />

5000 mg/l COD 10 ml 420805<br />

COD - reactors<br />

ET 108 / ET 125 / ET 125 SC**<br />

Chemical digestion of the samples is a prerequisite for the<br />

calculation of COD levels.<br />

The required temperatures and reaction time are programmed<br />

using the membrane keypad on the front of the<br />

AQUALYTIC ® reactors ET 108 / ET 125 / ET 125 SC**. All units are<br />

equipped with four temperature ranges (70/100/120/150 °C).<br />

At the end of the reaction time, the reactors automatically<br />

switch off and emit an acoustic alarm.<br />

Reactor<br />

Code<br />

COD-reactor ET 108 418930<br />

COD-reactor ET 125 418938<br />

COD-reactor ET 125 SC** 418939<br />

Technical data for COD-reactors<br />

Temperatur ET 108 / ET 125 / ET 125 SC**:<br />

70/100/120/150 °C ± 0,3 °C<br />

Temperature ET 108 / ET 125 / ET 125 SC**:<br />

70/100/120/150 °C ± 0.3 °C<br />

Timer<br />

30, 60, 120 minutes and<br />

continuous operation, automatic<br />

switch-off with alarm<br />

Heating ET 108: 270 Watts<br />

ET 125/ET 125 SC**: 400 Watts<br />

electronically controlled,<br />

with protection against overheating<br />

Heating-up time approx. 10 minutes<br />

Power supply ET 108: 230/115 V / 50-60 Hz<br />

ET 125 / ET 125 SC**:<br />

230 V / 50-60 Hz<br />

Weight ET 108: approx. 2.15 kg<br />

ET 125 / ET 125 SC**:<br />

approx. 3.60 kg<br />

Dimensions ET 108: 214 x 133 x 110 mm<br />

(L x W x H) ET 125 / AL25 SC**:<br />

275 x 155 x 95 mm<br />

Approval CE<br />

** SC: Single current (only 230 V)<br />

➲ Please see pages 46 and 47 for further information<br />

on our reactors<br />

Highlights<br />

■ ST-COD sealed tubes ready for use<br />

■ Suppression of chloride interference up<br />

to 1000 mg/l<br />

■ 3 ranges:<br />

Low range: 0 - 150 mg/l,<br />

meets ISO 15705:2002<br />

Middle range: 0 - 1500 mg/l,<br />

meets ISO 15705:2002<br />

High range: 0 - 15000 mg/l<br />

17<br />