Photometry<br />
Technical data<br />
Optics<br />
temperature-compensated LEDs and<br />
photosensor amplifier in water proof<br />
sample chamber<br />
Ranges 0 – 150 mg / l O 2 ±3.5% *)<br />
0 – 1500 mg / l O 2 ±3.5% *)<br />
0 –15000 mg / l O 2 ±3.5% *)<br />
Power<br />
supply<br />
Auto - OFF<br />
Display<br />
Storage<br />
Additional<br />
feature<br />
Calibration<br />
9 V powerpack battery<br />
providing 40 hours operation<br />
automatic switch-off<br />
backlit LCD (on keypress)<br />
internal ring memory for 16 data sets<br />
real time clock<br />
and date<br />
factory calibration and<br />
user calibration. Reset to factory<br />
calibration possible<br />
Dimensions<br />
190 x 110 x 55 mm<br />
(L x W x H)<br />
Weight<br />
0.4 kg (base unit)<br />
Environmental Temperature: 0 – 40 °C<br />
conditions rel. humidity: 30 – 90%<br />
(non condensing)<br />
CE- DIN EN 50081-1<br />
conformity VDE 0839 part 81-1: 1993-03<br />
DIN EN 50082-2<br />
VDE 0839 part 82-2: 1996-02<br />
* based on the use of potassium-hydrogenephthalate standards (DIN 38 409)<br />
COD VARIO tube tests<br />
The AQUALYTIC ® COD VARIO tube tests are available<br />
for the measuring ranges 0-150 mg/l O 2 , 0-1500 mg/l O 2 and<br />
0-15000 mg/l O 2 . Their chemical properties and a 16 mm tube<br />
diameter is suitable also for use with Hach photometers.<br />
Tube tests Quantity Order code<br />
0 - 150 mg/l O 2 (25 pc.) mercury free 420710<br />
(25 pc.) 420720<br />
(150 pc.) 420725<br />
0 - 1500 mg/l O 2 (25 pc.) mercury free 420711<br />
(150 pc.) mercury free 420716<br />
(25 pc.) 420721<br />
(150 pc.) 420726<br />
0 - 15000 mg/l O 2 (25 pc.) mercury free 420712<br />
(25 pc.) 420722<br />
(150 pc.) 420727<br />
Standard solutions<br />
Standard solutions are solutions with a defined concentration<br />
and are provided to check the operation methods and<br />
devices of the cuvette tests as well as the condition of optical<br />
filters and the instrument.<br />
Standard solutions Quantity Order code<br />
100 mg/l COD 30 ml 420803<br />
500 mg/l COD 30 ml 420804<br />
5000 mg/l COD 10 ml 420805<br />
COD - reactors<br />
ET 108 / ET 125 / ET 125 SC**<br />
Chemical digestion of the samples is a prerequisite for the<br />
calculation of COD levels.<br />
The required temperatures and reaction time are programmed<br />
using the membrane keypad on the front of the<br />
AQUALYTIC ® reactors ET 108 / ET 125 / ET 125 SC**. All units are<br />
equipped with four temperature ranges (70/100/120/150 °C).<br />
At the end of the reaction time, the reactors automatically<br />
switch off and emit an acoustic alarm.<br />
Reactor<br />
Code<br />
COD-reactor ET 108 418930<br />
COD-reactor ET 125 418938<br />
COD-reactor ET 125 SC** 418939<br />
Technical data for COD-reactors<br />
Temperatur ET 108 / ET 125 / ET 125 SC**:<br />
70/100/120/150 °C ± 0,3 °C<br />
Temperature ET 108 / ET 125 / ET 125 SC**:<br />
70/100/120/150 °C ± 0.3 °C<br />
Timer<br />
30, 60, 120 minutes and<br />
continuous operation, automatic<br />
switch-off with alarm<br />
Heating ET 108: 270 Watts<br />
ET 125/ET 125 SC**: 400 Watts<br />
electronically controlled,<br />
with protection against overheating<br />
Heating-up time approx. 10 minutes<br />
Power supply ET 108: 230/115 V / 50-60 Hz<br />
ET 125 / ET 125 SC**:<br />
230 V / 50-60 Hz<br />
Weight ET 108: approx. 2.15 kg<br />
ET 125 / ET 125 SC**:<br />
approx. 3.60 kg<br />
Dimensions ET 108: 214 x 133 x 110 mm<br />
(L x W x H) ET 125 / AL25 SC**:<br />
275 x 155 x 95 mm<br />
Approval CE<br />
** SC: Single current (only 230 V)<br />
➲ Please see pages 46 and 47 for further information<br />
on our reactors<br />
Highlights<br />
■ ST-COD sealed tubes ready for use<br />
■ Suppression of chloride interference up<br />
to 1000 mg/l<br />
■ 3 ranges:<br />
Low range: 0 - 150 mg/l,<br />
meets ISO 15705:2002<br />
Middle range: 0 - 1500 mg/l,<br />
meets ISO 15705:2002<br />
High range: 0 - 15000 mg/l<br />
17<br />
General Catalogue January 2010