BOD-Measurement<br />

Technical data<br />

Meas. principle<br />

Inductive stirring system<br />

Manometric; mercury-free;<br />

electronic pressure sensor<br />

Ranges 0 - 40, 0 - 80, 0 - 200,<br />

[mg/l O 2 ] 0 - 400, 0 - 800,<br />

0 - 2000, 0 - 4000 mg/l<br />

Accuracy* 0.5 % full scale at 20°C<br />

Applications BOD 5 , BOD 7 , OECD 301 F ...<br />

Result display<br />

Measurement<br />

parameter display<br />

Measurement<br />

period<br />

Auto result<br />

storage<br />

Storage<br />

interval<br />

Automatic<br />

start function<br />

Power<br />

supply<br />

BOD [mg/l]; 4 - digits ; 7 - segment LED<br />

BOD-range, volume, duration,<br />

time of measurement<br />

User-selectable, between<br />

1 and 28 days<br />

Up to 28 results, depending<br />

on measurement period<br />

– hourly (1 day)<br />

– every 2 hours (2 days)<br />

– daily (3-28 days)<br />

– After temperature<br />

equalisation of samples<br />

– Can be switched off<br />

3 alkaline-manganese<br />

batteries (“Baby” cells/size “C”)<br />

Battery life 1 year (normal use as BOD 5 meter –<br />

max. one reading a day); early warning<br />

before battery fails<br />

Interface<br />

RS 232 for printer or PC connection<br />

Clock<br />

Real-time clock<br />

Protection class IP 54 (sensor head)<br />

Dimensions<br />

375 x 195 x 230 mm<br />

(L x W x H) including stirring unit<br />

Weight<br />

3850 g, unit with bottles<br />

5750 g, complete with stirring unit<br />

Housing<br />

ABS<br />

Approval<br />

CE<br />

* No standard is available to check the accuracy of respirometric oxygen uptake<br />

measurement. Tests with a glucose-glutamic acid solution having a known theoretical<br />

BOD have shown that the variation is approximately 5% in the range of 50…100 mg/l BOD,<br />

and 3% for higher range. Minimum Response or sensitivity of respirometric systems is about<br />

0.05 … 1 mg/l.<br />

The microprocessor-controlled AQUALYTIC ® inductive stirring<br />

system is non-wearing and maintenance-free. In other words,<br />

there are no moving parts in the system.<br />

At regular intervals, the magnetic stirring rods are accelerated<br />

and slowed down again, taking them up to maximum speed<br />

and back down again. This ensures the centralization of the<br />

stirring rods.<br />

Stirring rods that move away from the centre of the bottle are<br />

re-centered quickly and reliably.<br />

The inductive actuation system guarantees maintenance-free<br />

operation (no need to replace drive belts or burnt-out drive<br />

motors) for many years.<br />

Highlights<br />

■ Maintenance-free and non-wearing<br />

■ Regular change in stirring speed<br />

■ Automatic centering of stirring rods<br />

■ No mechanical components in<br />

the stirring system<br />

Accessories<br />

Item<br />

Test set for AL606<br />

We also supply a test set to check for correct operation of the<br />

AL606 BOD meter. The set contains 8 BOD CM1 test tablets that<br />

cause a defined oxygen consumption.<br />

The tablets are easy to use. Simply place a tablet in the BOD<br />

bottle, start the measurement process, read off the BOD value<br />

after 5 days, and then compare with the defined value. If this<br />

value is within the quoted tolerance, this means that the BOD<br />

measuring system is functioning correctly.<br />

Temperature equalisation during BOD<br />

measurement<br />

Order code<br />

Sensor head 2444430<br />

BOD sample bottle 418644<br />

Brown glass, 500 ml<br />

BOD sample bottles, Brown glass, 418645<br />

500 ml, set of 6 bottles<br />

Cable for connection to a PC 2444440<br />

serial 9-pins<br />

Inductive stirring system 2444451<br />

for 6 samples, 110-240 V / 50-60 Hz<br />

Stirring rod 418637<br />

Stirring rod remover 418638<br />

Rubber gasket 418636<br />

Chemicals:<br />

Potassium hydroxide solution 2418634<br />

45 %, 50 ml<br />

Nitrification inhibitor (N-ATH) 2418642<br />

50 ml<br />

Overflow flask, 21.7 ml 418664<br />

Overflow flask, 56 ml 418655<br />

Overflow flask, 94 ml 418656<br />

Overflow flask, 157 ml 418657<br />

Overflow flask, 244 ml 418658<br />

Overflow flask, 360 ml 418659<br />

Overflow flask, 428 ml 418660<br />

Complete set 418654<br />

Overflow flasks<br />

Test set, BOD CM test tablets, 418328<br />

box with 8 tablets<br />

Temperature equalisation is essential prior to biological testing,<br />

as temperature has a major effect on biological activity. BOD<br />

measurements, for example, are always performed in a thermostatically<br />

controlled cabinet at a temperature of 20°C.<br />

For temperature equalisation, we recommend AQUALYTIC ®<br />

thermostatically controlled cabinets with a user-selectable<br />

temperature from 2°C to 40°C.<br />