This Planning Document was developed to enhance your knowledge of the Dräger portfolio and to provide guidance when designing a new Hybrid OR. This document contains an overview of Dräger products and their corresponding technical data. The last pages of this document show several exemplary Dräger solutions, including detailed technical drawings of the individual workplaces.

This Planning Document was developed to enhance your knowledge of the
Dräger portfolio and to provide guidance when designing a new Hybrid OR.
This document contains an overview of Dräger products and their corresponding
technical data.
The last pages of this document show several exemplary Dräger solutions,
including detailed technical drawings of the individual workplaces.

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

PLANNING DOCUMENTS F<strong>OR</strong> HYBRID <strong>OR</strong> | GENERAL PLANNING | 47<br />

This setup is optimized <strong>for</strong> a ceiling-mounted C-Bow system. All ceiling fixtures are out of the moving range of the C-Bow.<br />

It is not possible to have a central axis <strong>for</strong> <strong>OR</strong> lights over the <strong>OR</strong> table. Because of the moving range of the C-Bow, it is<br />

necessary to attach mounts <strong>for</strong> the <strong>OR</strong> lights on the sides. The light heads require long swivel arms to reach the operating<br />

field. The difference between solution 2 and 3 involves the docking of the anesthesia device and a MovaCart at the surgery<br />

supply unit. The supply units <strong>for</strong> surgery should have the same media outlets to enable rapid switching between anesthesia<br />

and surgery. This ensures a high degree of flexibility.<br />

12. ANESTHESIA<br />

A triple ceiling mount is selected to attach an<br />

X-Ray protection shield, the liftable anesthesia<br />

supply unit and long swivel arm <strong>for</strong> light head<br />

(satellite: 110.000 lux). The anesthesia device<br />

is lifted and docked to the supply unit (via<br />

MediaDocking). The supply unit has a long<br />

column. At this column an infusion pole is<br />

mounted.<br />

Devices at supply unit<br />

Monitoring (or at AN-device)<br />

IT/Documentation (or at AN-device)<br />

Pacemaker<br />

Defibrillator<br />

Infusions (or at AN-device)<br />

Additional devices supplied by unit<br />

Anesthesia device<br />

Blood gas analysis<br />

Sonographer<br />

Cell saver<br />

Endotracheal aspiration (often at AN-device)<br />

Further infusions<br />

Patient warmer<br />

13. SURGERY AND CARDIO<br />

There is a double ceiling mount to attach<br />

two supply units <strong>for</strong> surgery and cardio. The<br />

surgery supply unit is a long, liftable column<br />

with MediaDocking to lift a Mova-Cart<br />

(trolley with 4 shelves, 1 drawer). The cardio<br />

supply unit is a small liftable column with<br />

four separate swiveling display arms (load<br />

capacity <strong>for</strong> each arm: 15 kg).<br />

SURGERY<br />

Devices at supply unit<br />

HF surgery device<br />

ECG<br />

CARDIO<br />

Devices at supply unit<br />

Displays (4 x)<br />

Additional devices supplied by unit<br />

Surgical suction device<br />

Contrast agent injector<br />

Additional devices supplied by unit<br />

Heart lung machine HLM<br />

Hypothermia device<br />

7. DISPLAY HOLDER AND <strong>OR</strong> LIGHT<br />

For placing the display holder and the <strong>OR</strong><br />

light, a double ceiling mount is used. The<br />

display holder can carry 8 single displays<br />

(size: up to 21"). The light head (160.000 lux)<br />

is brought over the operating field by a long<br />

swivel arm.<br />