04.12.2014 Views

Planning Documents for Hybrid OR

This Planning Document was developed to enhance your knowledge of the Dräger portfolio and to provide guidance when designing a new Hybrid OR. This document contains an overview of Dräger products and their corresponding technical data. The last pages of this document show several exemplary Dräger solutions, including detailed technical drawings of the individual workplaces.

This Planning Document was developed to enhance your knowledge of the
Dräger portfolio and to provide guidance when designing a new Hybrid OR.
This document contains an overview of Dräger products and their corresponding
technical data.
The last pages of this document show several exemplary Dräger solutions,
including detailed technical drawings of the individual workplaces.

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

PLANNING DOCUMENTS F<strong>OR</strong> HYBRID <strong>OR</strong> | GENERAL PLANNING | 49<br />

This setup is optimized <strong>for</strong> a floor-mounted multi-axis C-Bow system 30° arrangement. All ceiling fixtures are out of the<br />

moving range of the C-Bow. Here, it is possible to have a central axis <strong>for</strong> <strong>OR</strong> lights over the <strong>OR</strong> table. In this layout there is<br />

a third light head at the foot end. The supply units <strong>for</strong> surgery should have the same media outlets to enable rapid switching<br />

between anesthesia and surgery. This ensures a high degree of flexibility.<br />

8. ANESTHESIA AND CARDIO<br />

A triple ceiling mount is selected to attach<br />

an X-Ray protection shield, the liftable anesthesia<br />

supply unit and long swivel arm <strong>for</strong><br />

light head (satellite: 110.000 lux). The anesthesia<br />

device is lifted and docked to the supply<br />

unit (via MediaDocking). The supply unit<br />

has a long column. At this column, an infusion<br />

pole is mounted.<br />

Devices at supply unit<br />

Monitoring (or at AN-device)<br />

IT/Documentation (or at AN-device)<br />

Pacemaker<br />

Defibrillator<br />

Infusions (or at AN-device)<br />

CARDIO<br />

Devices at supply unit<br />

Displays (4 x)<br />

Additional devices supplied by unit<br />

Anesthesia device<br />

Blood gas analysis<br />

Sonographer<br />

Cell saver<br />

Endotracheal aspiration (often at AN-device)<br />

Further infusions<br />

Patient warmer<br />

Additional devices supplied by unit<br />

Heart lung machine HLM<br />

Hypothermia device<br />

9. <strong>OR</strong> LIGHT AND X-RAY PROTECTION<br />

For the <strong>OR</strong> light, a double ceiling fixation at<br />

the foot end <strong>for</strong> a third <strong>OR</strong> light and an X-Ray<br />

protection shield is used. The light head<br />

(110.000 lux) is brought over the operating<br />

field by a long swivel arm.<br />

10. SURGERY<br />

There is a single ceiling mount to attach the<br />

surgery supply unit. This unit is a long, liftable<br />

column with MediaDocking to lift a MovaCart<br />

(trolley with 4 shelves, 1 drawer).<br />

SURGERY<br />

Devices at supply unit<br />

HF surgery device<br />

ECG<br />

Additional devices supplied by unit<br />

Surgical suction device<br />

Contrast agent injector<br />

11. DISPLAY HOLDER<br />

A separate ceiling mount is used <strong>for</strong> placing<br />

the display holder. The display holder can<br />

carry a single display (size: 60”).<br />