<strong>NSF</strong>/<strong>ANSI</strong> <strong>St<strong>and</strong>ard</strong> <strong>46</strong><br />

• Ultraviolet disinfection devices<br />

– Performance <strong>and</strong> evaluation<br />

• End <strong>of</strong> UV lamp life disinfection test<br />

– Evaluates disinfection capability <strong>of</strong> the device with the UV<br />

lamp output adjusted to 70% <strong>of</strong> initial<br />

– UV lamp output is measured after sampling on day 90 – if<br />

the measurement is 70% or less <strong>of</strong> initial output, no end <strong>of</strong><br />

lamp life test is required<br />

– If the measurement is greater than 70% lamp output is<br />

adjusted to 70%<br />

– Run in a 2-day period following the 90 day disinfection test<br />

under same test conditions, collecting samples 30 minutes<br />

after introduction <strong>of</strong> challenge water on day 2<br />

• 385 – Test omitted