gas chromatography<br />

gc instruments<br />

SRI Inlet Systems<br />

Every SRI <strong>GC</strong> is equipped with an unheated on-column inlet<br />

system. On a Model 310 chassis, you can replace that with<br />

any one of the other four inlet systems available—headspace,<br />

split/splitless, purge and trap, and flash. On a Model 8610C<br />

chassis, you can mount any of the inlet systems to a total<br />

of five.<br />

On-Column Inlet Systems<br />

Suitable for 0.53mm capillary and 1/8" packed columns.<br />

Includes EPC carrier gas controller and molecular sieve filter.<br />

On-Column Inlet Systems<br />

Description<br />

Part No.<br />

Additional On-Column Injector 8690023<br />

Heated Static Headspace Inlet System<br />

Ideal for the analysis of volatiles, especially where the sample<br />

matrix is dirty. Sample is heated and injected via a sample<br />

loop and gas sampling valve. Vial sleeve can be heated to<br />

90°C. Requires 10–20mL of sample for operation.<br />

Heated Static Headspace Inlet Systems<br />

Description<br />

Part No.<br />

Heated Static Headspace Injector 8690045<br />

SRI Detectors<br />

The SRI <strong>GC</strong> allows you to choose the detector(s) you want<br />

installed. The chassis can accommodate as many as four<br />

detectors to give you maximum flexibility.<br />

Table 1 – Detector Specifications<br />

Detector Selectivity Sensitivity (Approx.)<br />

CCD Hydrogen and hydrocarbons 500ppm<br />

TCD Universal 200-500ppm<br />

FID Hydrocarbons 1ppm<br />

DELCD Chlorinated and brominated 10ppb<br />

molecules<br />

FID/DELCD Hydrocarbons, chlorinated and 10ppm<br />

brominated molecules<br />

HID Universal with exception of neon 10ppm<br />

PID<br />

Aromatics and molecules with double 10ppb<br />

carbon bonds<br />

NPD Nitrogen and phosphorus 10ppm<br />

NPD/DELCD Nitrogen, phosphorus, chlorinated 100ppb<br />

and brominated<br />

TID Nitro functional groups (TNT etc.) 1ppb, 50ppb<br />

ECD Electronegative compounds 10ppb<br />

RGD Reducing gases (such as H2 and CO) 10ppb<br />

FPD Sulfur and phosphorus 200ppb and 10ppb<br />

FPD/FID Sulfur, phosphorus, and<br />

hydrocarbons<br />

200ppb, 10ppb, and<br />

100ppm<br />

Dual FPD Sulfur and phosphorus<br />

simultaneously<br />

200ppb and 10ppb<br />

FID Dual FPD<br />

Hydrocarbons, sulfur, phosphorus<br />

simultaneously<br />

100ppm, 200ppb,<br />

and 10ppb<br />

Heated Split/Splitless Injector<br />

Permits the use of narrow-bore capillary columns (0.32mm i.d.<br />

and smaller) in either split or splitless modes. Allows 0.53mm<br />

capillary columns and 1/8" packed columns to be used in<br />

split, splitless or on-column modes. Injector temperature is<br />

adjustable or programmable to 300°C.<br />

Heated Split/Splitless Injector<br />

Description<br />

Part No.<br />

Heated Split/Splitless Injector 8690034<br />

Heated Split/Splitless Injector with PTV 8611303<br />

Heated Flash Vaporization Injector<br />

Vaporizes the sample prior to the column. Ideal for desorption<br />

of SPME fibers or injection of extremely dirty samples. The<br />

injector heater block can be heated to 300°C.<br />

Heated Flash Vaporization Injector<br />

Description<br />

Part No.<br />

Heated Flash Vaporization Injector 8690025<br />

SRI <strong>GC</strong> Detectors (choose up to four per chassis)<br />

Description<br />

Part No.<br />

CCD (Catalytic Combustion Detector) 86902007<br />

TCD (Thermal Conductivity) 8690107<br />

FID (Flame Ionization) 8690010<br />

DELCD (Dry Electrolytic Conductivity) 86901026<br />

FID/DELCD Simultaneous 8692026<br />

HID (Helium Ionization Detector) 8690030<br />

PID (Photo-Ionization) 8690040<br />

NPD (Nitrogen-Phosphorus) 8690015<br />

NPD/DELCD (Series Detector) 8611299<br />

TID (Thermionic Ionization Detector) 8690017<br />

ECD (Electron Capture) 8690020<br />

RGD (Reduction Gas Detector) 8611293<br />

FPD (Flame Photometric) 8690080<br />

FPD/FID Simultaneous 8691080<br />

DFPD (Dual Flame Photometric) 8690085<br />