2013 Buyers' Guide 2013 Buyers' Guide - Filtration News
2013 Buyers' Guide 2013 Buyers' Guide - Filtration News
2013 Buyers' Guide 2013 Buyers' Guide - Filtration News
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
ISOCTD tests with four ~20µm cartridges<br />
from three vendors (plus an extra 10µm<br />
one from vendor B for comparison).<br />
These four cartridges were selected because<br />
they were an offering to two of our<br />
plants (at 1 gpm/10” flux) with high<br />
usage. Both IFTS and IBR were used along<br />
with the vendor’s tests (if available); and<br />
in one case (vendor B), an actual cartridge<br />
from our test lot was returned for their<br />
evaluation after the low DHC was discovered.<br />
Basically, Table III shows:<br />
• Vendor A’s 20µm spun cartridge<br />
rated at 45µm by the CFP (B.Pt./1.65)<br />
value compares reasonably well to 40µm<br />
values obtained by the vendor, IFTS, and<br />
IBR (except for two spurious results by<br />
IFTS at 40 ppm dust loading). Note also<br />
that the earlier 2005 CFP results showed<br />
much higher 70µm (Galwick) and<br />
72µm (IPA) values than the 45µm in<br />
2010 (e.g., this same discrepancy was<br />
found for vendor B cartridges in 2005 vs.<br />
2010).<br />
• Vendor A’s DHCs agree very well between<br />
testers, except that the 4 gpm/10”<br />
results are much lower (~85g) than the 1<br />
gpm/10” results (~194g), as expected.<br />
Also, the Dow 69g DHC at 1 gpm/10” flux<br />
is quite low for some unexplained reason.<br />
• Even though IFTS and IBR used different<br />
dust challenge and particle counter<br />
range protocols, the results for vendor A’s<br />
particle efficiencies are quite comparable,<br />
especially considering that a different cartridge<br />
was used in each test.<br />
• Vendor B’s 25µm cartridges show<br />
very low 18g DHC by both the vendor<br />
and by IFTS, but very good comparison<br />
to the 22µm (B.Pt./1.65) value. The low<br />
18g DHC was originally suspect such that<br />
a new cartridge (from this same test lot)<br />
was returned to B for their testing, and<br />
they also obtained 18g (vs. 35g as reported<br />
in their literature). As a result, the<br />
vendor suspected that this lot was mislabeled<br />
since their 10µm cartridges have a<br />
reported 55g DHC; and their 10µm ones<br />
tested here showed an equivalent 10µm<br />
rating by both their ISOCTD and by the<br />
10µm (B.Pt/1.65) value with new 2010<br />
cartridges. Note again that the 40µm<br />
value in 2005 was much higher than<br />
22µm value in 2010 for B’s 25µm rated<br />
cartridge. As a consequence of these poor<br />
results, vendor B’s cartridge wasn’t tested<br />
5-40<br />
www.filtnews.com • August 2012 • 27<br />
Table 3