Setting the Source Type<br />

The source setting in the MassHunter software is most likely set to “E<strong>SI</strong>,” which is the<br />

default setting if the system has not detected an i-button present.<br />

Change the source type to “Dual-E<strong>SI</strong>”.<br />

The capillary (in the case of E<strong>SI</strong>) is at high temperature. You can increase/decrease the<br />

temp inside the MassHunter software window (shown below in Figure 9).<br />

Enter Dual-<br />

E<strong>SI</strong> source<br />

Figure 9: MassHunter Control program window<br />

Optimal Parameters<br />

<strong>IonSense</strong> recommends using the following operational settings:<br />

Capillary Voltage: ~1000V<br />

Drying Gas: 8.5 L/min (Important to keep H 2 O cluster formation low)<br />

Fragmentor Voltage: 150-175V<br />

Octapole Voltage:<br />

o 250V for masses above 100 Da<br />

o 100V for masses below 100 Da<br />

If “Dual-E<strong>SI</strong> is not an option:<br />

On older-version instruments, “Dual-E<strong>SI</strong>” is not an option in the software.<br />

In this case:<br />

1. Set the source type to “E<strong>SI</strong>”.<br />

2. On the start button select:<br />

Programs → <strong>Agilent</strong>- MassHunterWorkstation → Acquisition tools →<br />

Instrument Configuration<br />

3. Deselect LC<br />

4. On the source button select "Do not wait for set-points to equilibrate"<br />

<strong>DART</strong>-<strong>Agilent</strong> <strong>Vapur</strong> <strong>Interface</strong> <strong>Manual</strong> (<strong>SI</strong>-<strong>300</strong>-A <strong>Rev</strong> 6) 10