Figure 6: Rubber stopper in place to create seal with interface (<strong>Agilent</strong> splash guard still<br />
in place)<br />
Insert the rubber stopper into the waste pumping port as shown in Figure 6 and remove<br />
the <strong>Agilent</strong> splash guard (pictured below in Figure 7) before closing the <strong>Vapur</strong> <strong>Interface</strong>.<br />
Note: If may be necessary to adjust the length of the <strong>Agilent</strong> clasp used to secure the<br />
<strong>Vapur</strong> <strong>Interface</strong> to the front of the instrument. Make sure that the <strong>Vapur</strong> <strong>Interface</strong> makes<br />
a good seal against the front of the instrument otherwise there will be vacuum instability.<br />
Figure 7: <strong>Agilent</strong> splash guard removed<br />