32<br />

6<br />

Localization and<br />

Biopsy Process<br />

• In case the marker-lesion axis should intersect with one of the three horizontal<br />

struts of the slat lateral plate and thus prevent direct access to the lesion, or the<br />

penetration should not be made in the direction of the thorax, the angulation angle<br />

of the marker (and thus of the subsequent needle guidance) can be readjusted to<br />

-30°, -15°, + 15°, +30°.Your toolbox will then give you the new distance between<br />

marker and lesion with the new angle as described in the following:<br />

• First, you must complete Steps 1 through 4 as described above (Steps 5 through<br />

7 are not relevant in this case).<br />

• Then you must mark the pivot point with a vertical line. This line is located 22<br />

mm behind the tip of the pointer. (See Ill. 8).<br />

8<br />

22 mm<br />

Mark pivot point<br />

<strong>BI160</strong>-<strong>GE</strong>