32<br />
6<br />
Localization and<br />
Biopsy Process<br />
• In case the marker-lesion axis should intersect with one of the three horizontal<br />
struts of the slat lateral plate and thus prevent direct access to the lesion, or the<br />
penetration should not be made in the direction of the thorax, the angulation angle<br />
of the marker (and thus of the subsequent needle guidance) can be readjusted to<br />
-30°, -15°, + 15°, +30°.Your toolbox will then give you the new distance between<br />
marker and lesion with the new angle as described in the following:<br />
• First, you must complete Steps 1 through 4 as described above (Steps 5 through<br />
7 are not relevant in this case).<br />
• Then you must mark the pivot point with a vertical line. This line is located 22<br />
mm behind the tip of the pointer. (See Ill. 8).<br />
8<br />
22 mm<br />
Mark pivot point<br />
<strong>BI160</strong>-<strong>GE</strong>