38<br />

For horizontal biopsies (no angulations, 0°) the handling of the <strong>NORAS</strong><br />

Mammotome® MR adaptor is equal to the standard <strong>NORAS</strong> positioning<br />

devices.<br />

For angulated biopsies please note that the center of rotation varies from the<br />

standard <strong>NORAS</strong> positioning devices. The center is located 68mm (not the usual<br />

22mm) behind the head of the pointer, which is visible in the MR image. In order<br />

to calculate the coordinates you have to subtract 68mm from the position of the<br />

head of the pointers line.<br />

For angulations use, turn the adjustment screw at the front of the fi ducial holder<br />

until you are able to change the angle. Anterior as well as posterior angulations<br />

are possible for 15° and 30°. After the angulations adjustment close the adjustment<br />

screw slowly until the new position of the fi ducial holder is fi xated.<br />

Mammotome ® MR Adapter Set for <strong>GE</strong> 8 ch Breast Coil<br />

MR10062-ET-<strong>GE</strong><br />

Targeting Set mit<br />

Halterung (Ethicon)<br />

In case of a very anterior lesion position the guide fork (Ethicon) can be set on a<br />

higher position, which enables another 17mm anterior access. Please note that<br />

in this procedure the holster must be held upside down so that tissue samples<br />

will fall downwards.<br />