For horizontal biopsies (no angulations, 0°) the handling of the <strong>NORAS</strong><br />
Mammotome® MR adaptor is equal to the standard <strong>NORAS</strong> positioning<br />
devices.<br />
For angulated biopsies please note that the center of rotation varies from the<br />
standard <strong>NORAS</strong> positioning devices. The center is located 68mm (not the usual<br />
22mm) behind the head of the pointer, which is visible in the MR image. In order<br />
to calculate the coordinates you have to subtract 68mm from the position of the<br />
head of the pointers line.<br />
For angulations use, turn the adjustment screw at the front of the fi ducial holder<br />
until you are able to change the angle. Anterior as well as posterior angulations<br />
are possible for 15° and 30°. After the angulations adjustment close the adjustment<br />
screw slowly until the new position of the fi ducial holder is fi xated.<br />
Mammotome ® MR Adapter Set for <strong>GE</strong> 8 ch Breast Coil<br />
MR10062-ET-<strong>GE</strong><br />
Targeting Set mit<br />
Halterung (Ethicon)<br />
In case of a very anterior lesion position the guide fork (Ethicon) can be set on a<br />
higher position, which enables another 17mm anterior access. Please note that<br />
in this procedure the holster must be held upside down so that tissue samples<br />
will fall downwards.<br />
<strong>BI160</strong>-<strong>GE</strong>