ASSAY METHOD:<br />

OPERATING MODE:<br />

SAMPLING SPEED:<br />

DISPENSING SYSTEM:<br />

LIQUID DETECTION SYSTEM:<br />

SAMPLE DILUTION:<br />

SAMPLE/REAGENT WORK CAPACITY:<br />

SAMPLE RACK:<br />

REAGENTS:<br />

REACTION TRAY:<br />

INCUBATION MODE:<br />

HISTORIC ARCHIVE:<br />

PATIENT ID:<br />

SOFTWARE:<br />

CALIBRATION:<br />

HOST CONNECTION:<br />

ERROR RECOVERY:<br />

POWER SUPPLY:<br />

POWER CONSUMPTION:<br />

SAFETY STANDARDS:<br />

ELECTROMAGNETIC COMPATIBILITY:<br />

IVD:<br />

DIMENSIONS AND WEIGHT:<br />

RIA<br />

TECHNICAL FEATURES<br />

STRUMENTI<br />

Open system for radioimmunoassay tests<br />

Random <strong>access</strong> mode. Multi-batch<br />

Up to 750 samples/hr<br />

2 independent needles; sample syringe capacity: 2500 µl; minimum volume:<br />

10 µl; reagent syringe capacity: 2500 µl; pipetting volume: 20-2000 µl<br />

(with 1 µl steps) in multipipetting<br />

Capacitive<br />

Programmable predilution, selectable for volume and ratios, serial dilution<br />

options, also for standard and controls<br />

96 samples with 5 tests simultaneous or 192 samples with 3 tests simultaneous<br />

Up to 96 samples (12x75 mm or 12x100 mm) or up to 54 samples (16x100 mm)<br />

dedicated vials directly fitting on the reaction tray<br />

5 removable racks for 72 tubes for a total of 360 tests<br />

R.T. - 45°C; shaking: 100-250 rpm<br />

Search per name, ID test or date, on screen or printer output.<br />

From keyboard, host computer<br />

Operating System: Windows;<br />

monitor: min. 800x600; 2 serial RS 232; 1 parallel port or USB for optional<br />

printer (Windows compatible)<br />

Complete standard curve or on 2 calibrators when using stored curves<br />

Bidirectional mode: serial (xmodem protocol), network<br />

Automated<br />

100 V – 60 Hz or 220 V – 50 Hz<br />

500W (compressor included)<br />

Directive 2006/95/EC<br />

Directive 2004/108/EC<br />

Directive 98/79/EC<br />

94 L. x 60 W. x 44 H. cm; 70 kg<br />