ASSAY METHOD:<br />
OPERATING MODE:<br />
SAMPLING SPEED:<br />
DISPENSING SYSTEM:<br />
LIQUID DETECTION SYSTEM:<br />
SAMPLE DILUTION:<br />
SAMPLE/REAGENT WORK CAPACITY:<br />
SAMPLE RACK:<br />
REAGENTS:<br />
REACTION TRAY:<br />
INCUBATION MODE:<br />
HISTORIC ARCHIVE:<br />
PATIENT ID:<br />
SOFTWARE:<br />
CALIBRATION:<br />
HOST CONNECTION:<br />
ERROR RECOVERY:<br />
POWER SUPPLY:<br />
POWER CONSUMPTION:<br />
SAFETY STANDARDS:<br />
ELECTROMAGNETIC COMPATIBILITY:<br />
IVD:<br />
DIMENSIONS AND WEIGHT:<br />
RIA<br />
TECHNICAL FEATURES<br />
STRUMENTI<br />
Open system for radioimmunoassay tests<br />
Random <strong>access</strong> mode. Multi-batch<br />
Up to 750 samples/hr<br />
2 independent needles; sample syringe capacity: 2500 µl; minimum volume:<br />
10 µl; reagent syringe capacity: 2500 µl; pipetting volume: 20-2000 µl<br />
(with 1 µl steps) in multipipetting<br />
Capacitive<br />
Programmable predilution, selectable for volume and ratios, serial dilution<br />
options, also for standard and controls<br />
96 samples with 5 tests simultaneous or 192 samples with 3 tests simultaneous<br />
Up to 96 samples (12x75 mm or 12x100 mm) or up to 54 samples (16x100 mm)<br />
dedicated vials directly fitting on the reaction tray<br />
5 removable racks for 72 tubes for a total of 360 tests<br />
R.T. - 45°C; shaking: 100-250 rpm<br />
Search per name, ID test or date, on screen or printer output.<br />
From keyboard, host computer<br />
Operating System: Windows;<br />
monitor: min. 800x600; 2 serial RS 232; 1 parallel port or USB for optional<br />
printer (Windows compatible)<br />
Complete standard curve or on 2 calibrators when using stored curves<br />
Bidirectional mode: serial (xmodem protocol), network<br />
Automated<br />
100 V – 60 Hz or 220 V – 50 Hz<br />
500W (compressor included)<br />
Directive 2006/95/EC<br />
Directive 2004/108/EC<br />
Directive 98/79/EC<br />
94 L. x 60 W. x 44 H. cm; 70 kg<br />