ASSAY METHOD:<br />

PRODUCTIVITY:<br />

WALK AWAY:<br />

OPERATING MODE:<br />

DISPENSING SYSTEM:<br />

WASHING SYSTEM:<br />

INCUBATION / REACTION:<br />

ALIQUOTE SYSTEM:<br />

THERMOSTAT:<br />

READER:<br />

SAMPLES:<br />

REAGENTS:<br />

CALIBRATION:<br />

QUALITY CONTROL:<br />

PATIENT FILE:<br />

SOFTWARE:<br />

HOST CONNECTION:<br />

OPERATING CONDITIONS:<br />

POWER SUPPLY:<br />

CONSUMPTION:<br />

SAFETY STANDARDS:<br />

ELECTROMAGNETIC COMPATIBILITY:<br />

IVD:<br />

DIMENSIONS AND WEIGHT:<br />

FIA<br />

TECHNICAL FEATURES<br />

Fully automated Random Access System in fluoroimmunoassays<br />

Up to 120 tests/hr<br />

400 tests<br />

Continuous loading of reagents, samples, calibrators, controls, pre-dilution cuvettes, reaction cuvettes and<br />

substrate. Simultaneous management of all the methods. Automatic substitution of the finished common<br />

reagents by continuous <strong>access</strong>. Separated solid and liquid waste station<br />

XYZ type with 2 independent needles and 2 500 µL syringes. (Optional) disposable tips. Capacitive<br />

coagulation level sensor (sensitivity 150 µL). Precision CV < 3% volume from 5 to 100 µL<br />

Two washer stations used for sedimentation, washing of the cuvettes and dispensing of the substrate<br />

180 positions’ station where the incubation and the reaction take place in orbital shake and at controlled<br />

temperature (37°C)<br />

On board for STAT<br />

Dry, at 37°C with ± 0.5°C precision. Heating up time 5’<br />

Fluorescence reading through led source with excitation at 370nm and revelation at 450nm<br />

Up to 120 samples on board (6 racks, 20 positions each). Continuous loading for routine and STA<br />

Primary, secondary tubes (12/16 mm) and paediatric cuvettes handling Positive identification by barcode<br />

reader Automatic predilution (56 positions) Automatic aliquote system on board Automatic dilution of outside<br />

range samples Reflex test<br />

Dedicated reagent pack for 50,100,150 or 200 tests 12 weeks stability on board; 20 different methods on<br />

board (up to 200 determinations each) 20 barcode identifiable positions for reagents Refrigerated reagent<br />

plate (4-8°C) Continuous loading without interruption to workflow<br />

Single use calibrator pack Positive calibrator identification via barcode reader; Two point calibration on Master<br />

Curve plus two control points Up to 4 weeks calibration stability<br />

Internal Quality Control management applying Westgard rules and Levy-Jennings charts<br />

Search for patients by ID, name, data and test Printing session report Printing patient report<br />

Multitask and Multilanguage interface Operating System: Windows XP or superior Touch Screen Monitor<br />

Bidirectional with serial port (RS232 interface), Web, ASTM<br />

15 – 32°C Temperature while operating; Maximum relative humidity 80%<br />

100 – 240 VAC; 50 – 60 Hz<br />

600 VA<br />

Directive 2006/95/EC<br />

Directive 2004/108/EC<br />

Directive 98/79/EC<br />

L.143 X W.78 X H.145 cm - 200 Kg<br />