The Clinical Chemistry photometer ScreenPlus is an instrument that can be easily placed in any type<br />

of clinical laboratory. It effects readings of end point, kinetic, fixed time and differential reaction, in<br />

absorbance and concentration, storage of 199 methods with possibility of programming each method<br />

with type of reaction, wavelength, working temperature, reading unit, normal values, incubation time<br />

and reading intervals. Reading is effected by single cuvette 10x10 mm., macro or semi micro, in<br />

mono and bi-chromatism. The instrument is completed with a liquid crystal display, 9 position<br />

thermostatic plate for cuvettes, a touch pad alphanumerical keyboard and an incorporated graphic<br />

printer for the print out of data, calibration and kinetic curves.<br />

71<br />

cod. 01060381SM<br />

TECHNICAL FEATURES<br />

READING METHODS: Absorbance, end point, kinetic, fixed time and differential reaction with standard<br />

(from 1 to 9) or K Factor<br />

OPTICAL SYSTEM: Halogen Lamp 20 W; 6 interferential filters, 340, 405, 505, 546, 578, 630 nm, with an accuracy<br />

of +/- 2 nm, band width of 8 nm and two free positions for optional filters;<br />

Solid state optical sensor; Photometric linearity from - 0.3000 to 2.9999 O.D. ;<br />

Photometric accuracy:+/-2%at 0.700O.D.<br />

READING CELL: 80 µL flow through cell; Optical path 10 mm; Minimum reading volume: 400 µL;<br />

Automatic withdrawal and wash cycle by means of peristaltic pump;<br />

Temperature regulation from 20°C to 45°C with 1°C resolution, with Peltier effect element;<br />

possibility to directly read open macro and semi-micro cuvettes;<br />

9 wells for open cuvette incubation<br />

SOFTWARE: All the photometer functions are managed by a FLASH memory microprocessor which can be<br />

up-dated through the serial output.; 199 programs;<br />

Automatic print out of results; Print out of the calibration and kinetic curves;<br />

Fitting types: linear, point to point and cubic; Pathological warnings;<br />

Functionality autodiagnosis<br />

USER INTERFACE: Functional alphanumerical keyboard; Alphanumerical liquid crystal display with four, twenty<br />

character lines; Thermal graphical printer with 192 dots per line RS 232 serial port<br />

POWER SUPPLY: 230/115 VAC, 50 - 60 Hz, 83 VA<br />

SAFETY STANDARDS: Directive 2006/95/EC<br />

ELECTROMAGNETIC COMPATIBILITY: Directive 2004/108/EC<br />

IVD: Directive 98/79/EC<br />

DIMENSIONS AND WEIGHT: L. 30 x W. 30 x H. 36 cm; 9,5 Kg