Chapter 3<br />

Ex<strong>per</strong>imental setup<br />

The entire set of measurements <strong>per</strong>formed in order to evaluate the physical <strong>per</strong>formances of a<br />

complete system of CR mammography, based on a FUJI computed ra<strong>di</strong>ography system, have<br />

been taken at the CSPO (“Centro Stu<strong>di</strong> Prevenzione Oncologica”) in Florence (Italy) with the<br />

su<strong>per</strong>vision of the Physical Health Staff of the “Azienda Ospedaliera Careggi”.<br />

In this chapter the characteristics of each hardware element will be described, with particular<br />

emphasis devoted to the CR <strong>di</strong>gital acquisition system FUJI FCR 5000 MA and to the X-ray<br />

source, based on a mammography facility manufactured by Instrumentarium, in daily clinical<br />

use at CSPO.<br />

3.1 The mammography <strong>un</strong>it<br />

The mammographic <strong>un</strong>it used as X-ray equipment in the present work is an Instrumentarium<br />

model Diamond, with a molybdenum anode and an internal molybdenum filter 25 �m thick (a<br />

25 �m rho<strong>di</strong>um filter and a 500 �m aluminum filter are also available but have not been used in<br />

this investigation). The exit window isa1mmthick beryllium one. Great care has been taken to<br />

<strong>per</strong>form all the measurements in a clinical <strong>di</strong>agnostic-like environment. The X-ray tube voltage<br />

was set to the nominal 28.0 kVp, but a set of measurements have been <strong>per</strong>formed to check the<br />

precision of the control <strong>un</strong>it. A comparison of the value set on the <strong>un</strong>it control and of the same<br />

as measured by a High Voltage Voltmeter (Victory Model Nero) is shown in Tab. 3.1.<br />

The other parameter which can be adjusted by the o<strong>per</strong>ator is the mAs value, that is strictly<br />