Figure 3.7: The IP cassette.<br />

both manufactured by FUJI for computed ra<strong>di</strong>ography. This in order to increase the amo<strong>un</strong>t<br />

of X-ray absorption and to reduce the photon noise. In general, increasing the thickness of the<br />

phosphor layer reduces sharpness. However the FUJI IP compensates this effect by minimizing<br />

the phosphor grain size, which enhances sharpness [10]. A comparison between the single side<br />

rea<strong>di</strong>ng and dual side rea<strong>di</strong>ng IP layer structure is shown in Fig. 3.8.<br />

Figure 3.8: The IP layer structure for the FUJI High Resolution single and dual side rea<strong>di</strong>ng<br />

plates.<br />

For all the measurements done, described in Chapter 4, the same IP has been used. Never-<br />