Thermo Scientific <strong>NanoDrop</strong> SpectrophotometersNucleic AcidMe<strong>as</strong>urementsSample Me<strong>as</strong>urements2. Pipette an aliquot of the nucleic acid sample onto the lower me<strong>as</strong>urement pedestal and lowerthe sampling arm. Click Me<strong>as</strong>ure.Tip: If me<strong>as</strong>uring more than one sample at a time on the <strong>NanoDrop</strong> 8000, it is important to use amulti-channel pipettor to deliver the sample aliquots.3. After the me<strong>as</strong>urement is complete, use a dry, lint-free lab wipe to remove the sample fromboth the top and bottom me<strong>as</strong>urement surfaces.4. Change pipette tips for the next me<strong>as</strong>urement.Tip: If me<strong>as</strong>uring multiple replicates of the same sample, it is important to use a fresh aliquotfor every me<strong>as</strong>urement to ensure accurate results.5. Review spectral image to <strong>as</strong>sess sample quality.10mm Absorbance38 -36 -260nm34 -32 -30 -28 -26 -24 -22 -20 -18 -280nm16 -230nm14 -12 -10 -8 -6 -4 -2 -0 -220 230 240 250 260 270 280 290 300 310 320 330 340Wavelength nmMe<strong>as</strong>urementsTypical Nucleic Acid SpectrumRefer to the Purity RatiosandTroubleshootingsections on pages 14 and 17 for more information.11