Thermo Scientific <strong>NanoDrop</strong> SpectrophotometersNucleic AcidIntroductionSample RetentionTechnologyIntroduction• Pipette 1 – 2 μL sample directly onto the me<strong>as</strong>urement pedestal.Tip: 0.5 μL aliquots may be used for samples that have 10 mm equivalent absorbance valuesof 3.0 or higher ( >150 ng/ μL dsDNA). <strong>NanoDrop</strong> 2000/2000c model only.• Lower the sampling arm and initiate a spectral me<strong>as</strong>urement using the software on the PC.• Surface tension is used to hold samples in place between two optical fibers.• Light from a xenon fl<strong>as</strong>hlamp p<strong>as</strong>ses through the top optical fiber, down through the liquidcolumn and is detected by the internal spectrometer.• When the me<strong>as</strong>urement is complete, raise the sampling arm and wipe the sample from boththe upper and lower pedestals using a dry, lint-free laboratory wipe.Using <strong>this</strong> technology, <strong>NanoDrop</strong> spectrophotometers have the capability to me<strong>as</strong>ure samplesbetween 50 and 200 times more concentrated than samples me<strong>as</strong>ured using a standard1 cm cuvette.5