Download this publication as PDF - NanoDrop
Download this publication as PDF - NanoDrop
Download this publication as PDF - NanoDrop
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Thermo Scientific <strong>NanoDrop</strong> SpectrophotometersNucleic AcidTroubleshootingConnection ErrorsIf your instrumentoperates properlymost of the time, butconnection errorsappear intermittently,the instrument maynot be receivingpower, or recognizingthe USB connection.• Ensure that the USB and power cables are plugged into the backof the instrument, and that the instrument is receiving power.• Many instrument issues can be addressed by a simple powerrestart.1. Exit the software.2. Disconnect the instrument power cord and USB cable.3. Reconnect the instrument power cord first, then the USB cable.4. Restart the software.If the error persists, contact Technical Support. Outside of theUS and Canada, ple<strong>as</strong>e contact your local <strong>NanoDrop</strong> productsdistributor.Signal ErrorsSome errormessages aretriggered whenlittle to no lightreaches the detectorduring initializationor a me<strong>as</strong>urement.• Refer to the cleaning directions under the Best Practicessection on page 6.• Run the Intensity Check diagnostic. Refer to the model-specificuser guide for additional information.23