IHE Patient Care Device Technical Framework
IHE Patient Care Device Technical Framework
IHE Patient Care Device Technical Framework
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>IHE</strong> <strong>Patient</strong> <strong>Care</strong> <strong>Device</strong> <strong>Technical</strong> <strong>Framework</strong>, Vol. 2: Transactions________________________________________________________________________D.8.3 Channel: DeliveryInfusor Source Channel ParametersName Term Code Data Type Units ValuesTotalCurrentRateTotalVolumeInfusedOperational StatusOperational ModeMDC_FLOW_FLUID_PUMP (2::26712) Numeric::FLOAT-Type MDC_DIM_MILLI_L_PER_HR(4::3122)MDC_VOL_INFUS_ACTUAL_TOTAL(2::26876)Numeric::FLOAT-Type MDC_DIM_MILLI_L (4::1618)MDC_PUMP_STAT (2::53436) Enumeration::TEXT 21 N/AMDC_PUMP_MODE (2::53432) Enumeration::TEXT 21 N/A“pump-statusinfusing”+“pump-statuskvo”+ “pumpstatus-ready”+”pump-statusstandby”+“pump-statuspaused”“pump-modenominal”+“pump-modesecondary”+“pump-modedrug-dosing”21 This parameter is specified as a set of bit flags, but for this technical framework, the enumerated textstrings shall be used.Rev. 1.1 TI: 2006-08-1582Copyright © 2005-2006: ACC, ACCE, HIMSS, and RSNA