Aim: Solid-phase antibody (Ab) detection assays are among the most important tools <strong>for</strong> determiningunacceptable antigens (UA) currently used in HLA labs. We report here a comparison of Ab identificationamong the Michigan HLA labs aiming to evaluate the extent of inter-laboratory variability in the solidphaseplat<strong>for</strong>m.Methods: Four sera were tested by six labs using the Luminex-based bead assays. Data collected includedtest per<strong>for</strong>med, reagent vendor and lot used, Ab specificities identified and the MFI values, UA assigned,and thresholds used <strong>for</strong> reporting.Results: While strong Ab with a mean MFI of >6<strong>00</strong>0 were concordantly reported as UAs by all labs, muchvariability was observed in the identification of moderate (MFI 2<strong>00</strong>0-6<strong>00</strong>0) and weak (MFI