ITBV TD (ml)<br />

Intrathoracaal bloedvolume<br />

ITBV = PBV + GEDV<br />

r = 0.96<br />

ITBV = 1.25 * GEDV – 28.4 [ml]<br />

RAEDV<br />

RVEDV PBV LAEDV LVEDV<br />

ITBV kan direct worden gemeten met de thermal dye dilution technique (COLD System) en uit studies<br />

is gebleken dat ITBV 25% groter is dan GEDV, gemeten met de s<strong>in</strong>gle thermodilution technique<br />

(PiCCO) ITBV = 1,25 x GEDV<br />

GEDV vs. ITBV <strong>in</strong> 57 <strong>in</strong>tensive care patients<br />

Sakka et al, Intensive Care Med 26: 180-187, 2000