MLX90615
MLX90615
MLX90615
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>MLX90615</strong>Infra Red Thermometer8.5.2 Request conditiont REQSCLFigure 13: Request (switch to SMBus) conditionIf PWM is enabled, the <strong>MLX90615</strong>’s SMBus Request condition is needed to disable PWM and reconfigureSDA/PWM pin before starting SMBus communication. Once disabled, PWM can only be enabled byswitching the supply Off-On. The <strong>MLX90615</strong>’s SMBus request condition requires forcing the SCL pin LOW fora period longer than the request time (t REQ ). The SDA/PWM line value is ignored in this case.8.5.3 PWM is disabledSMBus RequestIf PWM is disabled by means of EEPROM the SDA/PWM pin is directly used for the SMBus communicationafter POR.3901090615 Page 16 of 30 Data SheetRev 00128/Aug/2008