Dispensing neutralising agent via DOS 3Setting the dosage concentrationProgrammable functionsThe dosage concentration for the neutralising agent must be set according tothe manufacturer's recommendations or the official standards required.Step^ Press 0 until E02 appears.^ Turn the programme selector to CHEM60°C-5'.^ Press ß as many times as necessary orhold pressed in until the required valueappears in the display, e.g. 0.20.Display / ResultE02The dosage concentration whichhas been programmed in willappear in the display,e.g. 0.10 = 0.1 %0.20The dosage concentration level hasnow been set.Dosage concentration set:Venting the dispensing systemThe dispensing system must be vented:– when the dispensing system is used for the first time,– if the chemical agent container runs out and is not refilled in time.Step^ Press 0 until E02 appears.Display / ResultE02^ Turn the programme selector to 6. d§3^ Press ß.^ Exit from programming mode.Then start the Pre-rinse D programme.The dispensing system will bevented automatically. 0 will appearin the display when the ventingprocess has finished.Any agents which are in the washcabinet after the venting processwill be diluted before being drainedaway.43