MAN0551.P65 Issue 06 - 03/042108M05016. CALIBRATION(17) Wait and the display will show:Remove gas, PressEnter: xx.x(18) Remove the gassing cover from the weather protection.Ensure the reading has returned to 00.0% LEL.(19) Press ↵ and the display will show:Press Enter ToRelease 4 - 20Releases the analog output from 2mA (inhibit signal)to normal operation (4mA).(20) Press ↵ and the display will show for 3 seconds:4 - 20 releasedThe analogue output has now returned to normaloperation.(21) Wait and the display will show:CalibrateCal SensorReturns to the Calibrate Menu.(22) Press esc twice to return to the Main Menu:MainDisplay72