03.12.2012 Views

EVALUATION OF THE KONELAB 20i CLINICAL CHEMISTRY ...

EVALUATION OF THE KONELAB 20i CLINICAL CHEMISTRY ...

EVALUATION OF THE KONELAB 20i CLINICAL CHEMISTRY ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>EVALUATION</strong> <strong>OF</strong> <strong>THE</strong> <strong>KONELAB</strong> <strong>20i</strong><br />

<strong>CLINICAL</strong> <strong>CHEMISTRY</strong> ANALYZER<br />

Juutilainen, P. 1 , Seppälä, K. 2 , Lampinen, H. 3<br />

1Clinical Laboratory, Municipal Health Center, Pori, Finland<br />

2Clinical Laboratory, Joint Municipal Authority for Public Health Work, Kankaanpää, Finland<br />

3Labsystems Clinical Laboratory Division, Espoo, Finland<br />

Abstract<br />

Konelab <strong>20i</strong> is a new, selective clinical chemistry analyzer for routine clinical chemistry tests,<br />

specific proteins, DoA, TDM, toxicology tests and electrolytes. Methods applied are colorimetric<br />

and turbidimetric as well as direct potentiometry for electrolyte measurement. Theoretical<br />

throughput of the analyzer is up to 200 photometric and 180 ISE tests/hour. The analyzer is<br />

capable of handling both routine and stat requests.<br />

Konelab <strong>20i</strong> was evaluated according to guidelines given by ECCLS and modified by Labquality<br />

(Finland) to meet with local requirements. Nine tests (K, Na, Cl, Li, ALT, ALP, CRP, Chol and<br />

Uric Acid) were evaluated and results were compared to those obtained with Cobas Mira,<br />

Konelab 60i (Cl), KONE Microlyte (Li) and Eppendorf Efox 5053 (Li) analyzer. Results of all<br />

evaluated tests showed a good correlation.<br />

Calibrations and daily routines were performed according to recommendations of the<br />

manufacturer. A calibration interval of two weeks was recommended for CRP and Cholesterol<br />

assays, and one month for Uric Acid and electrolytes. During the whole calibration period, no<br />

significant long-term drift was found.<br />

Within-run imprecision (CV%) was calculated from patient samples measured as duplicates.<br />

Coefficients of variation were below 1% for K, Na, and Cl and Lithium at high level. For ALP,<br />

Cholesterol and Uric Acid CVs below 2% were obtained. CRP gave 2.6% for high level samples<br />

and 7.9% for low levels. For ALT results CVs between 9.6 and 3.4% were observed.<br />

Between-run imprecision (CV%) was below 3.5 % for all tests, except CRP.<br />

No sample or reagent carry-over was found.<br />

Instruments<br />

Konelab <strong>20i</strong> (Labsystems Clinical Laboratory Division, Espoo, Finland) is random access, selective<br />

clinical chemistry analyzer. The number of on-board sample positions is 84 plus 6 stat samples. Samples<br />

are placed on continuous loading segments. Integrated barcode reader and cup type recognition for<br />

samples are included. Sample tubes of 5 ml, 7 ml and 10 ml with tailored segments can be used, as well<br />

as 0.5 ml and 2 ml analyzer cups. Sample volume is from 1 to 120 µl, typically 2-15 µl for photometric<br />

tests and 50 µl fo Na + , K + and Cl - .<br />

The refrigerated reagent disk has continuous loading positions of 35 vessels that can be 10, 20 or 60 ml.<br />

The reagents can be identified with external barcode reader. The real time reagent status is seen on the<br />

workstation window. The reagent volumes are 2-250 µl, up to four reagent additions per test is possible.<br />

The reaction cuvette for photometric tests is a discrete multicell cuvette with 12 square reaction cells in<br />

a row. Cuvettes can be loaded continuously. The on-board capacity is 50 multicell cuvettes, which<br />

means 600 reaction positions. Reaction end volume is from 100 to 250 µl.<br />

Automatic sample pre- and re-dilutions between 1:1 and 1:120 is available. Calibration with separate<br />

calibrator samples or with automatically diluted series from a stock calibrator is possible. Linear, nonlinear<br />

and bias calibrations as well as several blank possibilities can be used.<br />

Real time QC function is available including multiple and variable (Westgard) rules and chart printouts<br />

both daily and cumulative basis. The control intervals are programmable.<br />

The photometer is a single channel system with beam splitting reference and interference filters between<br />

340 and 700 nm.<br />

For comparison with the Konelab <strong>20i</strong> were used the following analyzers:<br />

Cobas Mira (Roche Diagnostics, Basel, Switzerland)<br />

Eppendorf EFOX 5053 (Eppendorf-Netheler-Hinz GmbH, Hamburg, Germany)<br />

Konelab 60i (Labsystems Clinical Laboratory Division, Espoo, Finland)<br />

KONE Microlyte (Labsystems Clinical Laboratory Division, Espoo, Finland)


Materials and methods<br />

The methods and reagents used in this study are summarized in Table 1.<br />

Reagents used in Konelab <strong>20i</strong> were prepared according to the instructions of the manufacturer. They<br />

were stored in the analyzer’s refrigerated reagent compartment.<br />

Reagents for Cobas Mira were stored between analysis sections in the refrigerator at + 4 o C.<br />

Calibrators and recommended calibration intervals are given in Table 2.<br />

Materials used for between-run imprecision study are given in Table 4.<br />

For the entire evaluation, all control samples were dissolved at the same time and stored in 1 ml aliquots<br />

at –20 o C.<br />

Patient samples, minimum 81 for each analyte representing low, medium and high levels of measuring<br />

range were collected and stored at –20 o C. They were analyzed as duplicates in Konelab <strong>20i</strong>. A single<br />

determination was performed at the same time in Cobas Mira, or on the same day in Konelab 60i or<br />

KONE Microlyte. Lithium samples were measured first in Eppendorf EFOX, then stored at –20 o C until<br />

analyzed on the same day in KONE Microlyte and Konelab <strong>20i</strong>.<br />

Double determinations in Konelab <strong>20i</strong> were used to calculate the within-run imprecision. In the<br />

comparative study with other instruments the first of duplicate values was used in regression analysis.<br />

The sample carry-over was studied with low and high level patient sera using Uric Acid (low level<br />

sample 105 µmol/l and high level sample 675 µmol/l) and C-Reactive Protein (low level sample 21 mg/l<br />

and high level sample 207 mg/l) tests.<br />

The reagent carry-over was studied using Lactate Dehydrogenase (240 U/l) and Alanine<br />

Aminotransferase (31 U/l) tests.<br />

Table 1. Tests, methods and reagents used in evaluation<br />

Konelab <strong>20i</strong> Cobas Mira<br />

Test Method Method<br />

Potassium ISE, direct measurement ISE, direct measurement<br />

Sodium ISE, direct measurement ISE, direct measurement<br />

Chloride ISE, direct measurement ISE, direct measurement /Konelab 60i<br />

Lithium ISE, direct measurement ISE, direct measurement /KONE Microlyte<br />

ISE, direct measurement Flame emission / Eppendorf EFOX 5053<br />

ALP SCE, DEA-buffer<br />

Kinetic<br />

Konelab 981359<br />

ALT IFCC/ECCLS<br />

Kinetic<br />

Konelab 981361<br />

C-Reactive<br />

Protein<br />

Immunoturbidimetry<br />

End-point with sample blank<br />

Konelab 981666<br />

Cholesterol CHOD<br />

End-point with reagent blank<br />

Konelab 981368<br />

Uric acid Uricase, Trinder (AOX)<br />

End-point with reagent blank<br />

Konelab 981391<br />

SCE, DEA-buffer<br />

Kinetic<br />

Medix Biochemica 0054<br />

IFCC/ECCLS<br />

Kinetic<br />

Medix Biochemica 0558<br />

Immunoturbidimetry<br />

End-point with sample blank<br />

Orion Diagnostica 67393 and 67179<br />

CHOD<br />

End-point with reagent blank<br />

Konelab 981368<br />

Uricase, PAP<br />

End-point with reagent blank<br />

Roche 0736813


Table 2. Calibrators and recommended calibration intervals of the analytes<br />

Konelab <strong>20i</strong> Cobas Mira<br />

Analyte Calibrator Cal. Interval Calibrator Cal. Interval<br />

Potassium ISE Cal 1<br />

Every assay ISE Standard 1, 07 2568 4 Every assay<br />

Sodium Konelab 984031<br />

ISE Standard 2, 07 2569 2 240 min interval<br />

Chloride<br />

See Konelab <strong>20i</strong> *<br />

See Konelab <strong>20i</strong> *<br />

Standard Solution 1, 980796<br />

Standard Solution 2, 980797<br />

Every assay<br />

Lithium<br />

Standard Solution 3, 980798 ** 240 min interval **<br />

ISE Cal 2 ja 3<br />

One month System Blank Sol. 0030 340.000 At the start and<br />

Konelab 984035<br />

Serum Standard 0030 311.000 *** then every 10 min ***<br />

ALP Factor (molar absorptivity) - Factor (molar absorptivity) -<br />

ALT Factor (molar absorptivity) - Factor (molar absorptivity) -<br />

C-Reactive CRP Calibration Set Two weeks CRP Calibrator Set<br />

When the antisera lot<br />

Protein Konelab 981674<br />

Orion Diagnostica 68090 changes<br />

Cholesterol Calibrator 2<br />

Three weeks Seronorm Lipid<br />

Two weeks<br />

Konelab 980502<br />

Sero<br />

Uric acid Calibrator 1<br />

One month Biocal<br />

Every day<br />

Konelab 980501<br />

Bioclin<br />

* = Konelab 60i<br />

** = Kone Microlyte<br />

*** = Eppendorf EFOX 5053<br />

Results<br />

Imprecision<br />

The within-run imprecision was better than 1 % in ISE measurements with exception of Lithium at low levels.<br />

For Alkaline Phosphatase, Cholesterol and Uric Acid CVs below 2 % were obtained. C-Reactive Protein gave<br />

2.6 % for medium and high level samples and 7.9 % for low level. For Alanine Aminotransferase results CVs<br />

between 9.6 and 3.4 % were observed.<br />

Between-run imprecision (CV%) was below 3.5 % for all tests, except C-Reactive Protein.<br />

The results are presented in Tables 3 and 4.<br />

Carry-over<br />

No sample or reagent carry-over was found.<br />

Correlation<br />

The regression analysis of the results are presented in Figures 1-10.<br />

The correlations were very good. A bias of 5 to 15 % was found in the results of Lithium assays. Also a<br />

significant bias of 10 to 20 % was found for Alkaline Phosphatase results.


Table 3. Within-run imprecision<br />

Calculated from duplicate determinations of patient samples<br />

Analyte Mean SD CV% n<br />

Potassium 3,73 0,013 0,4 28<br />

(mmol/l) 4,35 0,032 0,7 29<br />

5,49 0,030 0,6 28<br />

Sodium 138,7 0,435 0,3 29<br />

(mmol/l) 143,4 0,447 0,3 30<br />

147,3 0,465 0,3 30<br />

Chloride 92,2 0,316 0,3 30<br />

(mmol/l) 99,8 0,835 0,8 28<br />

107,2 0,381 0,4 31<br />

Lithium 0,49 0,008 1,6 34<br />

(mmol/l) 0,74 0,007 0,9 33<br />

1,00 0,005 0,5 33<br />

ALP 109,9 0,972 0,9 27<br />

(U/l) 180,7 2,639 1,5 27<br />

525,6 7,185 1,4 27<br />

ALT 7,7 0,736 9,6 36<br />

(U/l) 22,7 0,779 3,4 28<br />

59,3 2,899 4,9 26<br />

CRP 5,7 0,451 7,9 27<br />

(mg/l) 42,0 1,086 2,6 28<br />

148,4 3,824 2,6 28<br />

Cholesterol 4,36 0,060 1,4 29<br />

(mmol/l) 5,80 0,091 1,6 29<br />

7,62 0,149 2,0 31<br />

Uric acid 221,4 1,824 0,8 29<br />

(µmol/l) 379,0 3,284 0,9 30<br />

522,8 6,927 1,3 31


Konelab <strong>20i</strong> (mmol/l)<br />

8,0<br />

7,5<br />

7,0<br />

6,5<br />

6,0<br />

5,5<br />

5,0<br />

4,5<br />

4,0<br />

3,5<br />

3,0<br />

2,5<br />

2,0<br />

Table 4. Between run imprecision<br />

Analyte Sample Mean SD CV% n<br />

Potassium Pathonorm L 2,195 0,041 1,8 19<br />

(mmol/l) Pathonorm H 6,432 0,089 1,4 19<br />

Pooled sera C 4,616 0,083 1,8 19<br />

Sodium Pathonorm L 125,8 2,069 1,6 17<br />

(mmol/l) Pathonorm H 159,3 1,961 1,2 17<br />

Pooled sera C 150,3 2,443 1,6 17<br />

Chloride Pathonorm L 73,6 1,688 2,3 18<br />

(mmol/l) Pathonorm H 128,1 3,334 2,6 18<br />

Pooled sera C 110,8 2,243 2,0 19<br />

Lithium Nortrol 0,95 0,018 1,9 17<br />

(mmol/l) Abtrol 2,46 0,050 2,0 18<br />

Pathonorm H 1,61 0,037 2,3 18<br />

ALP Pathonorm L 167,6 5,512 3,3 16<br />

(U/l) Pathonorm H 630,9 21,020 3,3 16<br />

Pooled sera C 166,8 4,328 2,6 16<br />

ALT Pathonorm L 32,2 1,098 3,4 18<br />

(U/l) Pathonorm H 105,3 2,697 2,6 18<br />

Abtrol 124,5 3,365 2,7 18<br />

CRP CRP Control 29,8 1,437 4,8 17<br />

(mg/l) CRP C. High 82,8 3,192 3,9 17<br />

Pooled sera V 157,5 10,696 6,8 15<br />

Cholesterol Pathonorm L 3,19 0,025 0,8 16<br />

(mmol/l) Pathonorm H 6,91 0,209 3,0 17<br />

Pooled sera C 5,68 0,064 1,1 17<br />

Uric acid Pathonorm L 177,6 2,910 1,6 19<br />

(µmol/l) Pathonorm H 574,1 8,689 1,5 19<br />

Pooled sera C 343,2 5,105 1,5 19<br />

Where: 1) Pathonorm L (Sero), Lot 703078<br />

2) Pathonorm H (Sero), Lot MI8128x<br />

3) Pooled sera (PKS), Lot C, normal values<br />

4) Nortrol (Konelab Corp.), Lot 4165F<br />

5) Abtrol (Konelab Corp.), Lot 4205E<br />

6) CRP Control (Konelab Corp.), Lot 1586<br />

7) CRP Control High (Konelab Corp.), Lot 1576<br />

8) Pooled sera (PKS), Lot V, high CRP-level<br />

Potassium<br />

y = 1.069x – 0.242<br />

R 2 = 0.983 n = 85<br />

2,0 2,5 3,0 3,5 4,0 4,5 5,0 5,5 6,0 6,5 7,0 7,5 8,0<br />

Cobas Mira (mmol/l)<br />

Konelab <strong>20i</strong> (mmol/l)<br />

160<br />

155<br />

150<br />

145<br />

140<br />

135<br />

130<br />

125<br />

120<br />

Sodium<br />

y = 0.913x + 12.448<br />

R 2 = 0.888 n = 89<br />

120 125 130 135 140 145 150 155 160<br />

Cobas Mira (mmol/l)


Konelab <strong>20i</strong> (mmol/l)<br />

Konelab <strong>20i</strong> (mmol/l)<br />

130<br />

120<br />

110<br />

100<br />

90<br />

80<br />

70<br />

1,6<br />

1,4<br />

1,2<br />

1,0<br />

0,8<br />

0,6<br />

0,4<br />

0,2<br />

0,0<br />

Chloride<br />

y = 0.957x + 3.738<br />

R 2 = 0.929 n = 89<br />

70 80 90 100 110 120 130<br />

Konelab 60i (mmol/l)<br />

Lithium<br />

y = 0.968x + 0.068<br />

R 2 = 0.986 n = 100<br />

0,0 0,2 0,4 0,6 0,8 1,0 1,2 1,4 1,6<br />

Microlyte (mmol/l)<br />

Konelab <strong>20i</strong> (mmol/l)<br />

1,6<br />

1,4<br />

1,2<br />

1,0<br />

0,8<br />

0,6<br />

0,4<br />

0,2<br />

0,0<br />

Lithium<br />

y = 0.966x + 0.080<br />

R 2 = 0.995 n = 83<br />

0,0 0,2 0,4 0,6 0,8 1,0 1,2 1,4 1,6<br />

Eppendorf EFOX 5053 (mmol/l)


Konelab <strong>20i</strong> (U/l)<br />

Konelab <strong>20i</strong> (mg/l)<br />

1200<br />

1000<br />

800<br />

600<br />

400<br />

200<br />

0<br />

400<br />

350<br />

300<br />

250<br />

200<br />

150<br />

100<br />

50<br />

0<br />

ALP<br />

y = 1.098x + 14.688<br />

R 2 = 0.994 n = 81<br />

0 200 400 600 800 1000 1200<br />

Cobas Mira (U/l)<br />

CRP<br />

y = 1.067x – 4.991<br />

R 2 = 0.991 n = 83<br />

0 50 100 150 200 250 300 350 400<br />

Cobas Mira (mg/l)<br />

Konelab <strong>20i</strong> (U/l)<br />

Konelab <strong>20i</strong> (mmol/l)<br />

200<br />

180<br />

160<br />

140<br />

120<br />

100<br />

80<br />

60<br />

40<br />

20<br />

0<br />

12<br />

11<br />

10<br />

9<br />

8<br />

7<br />

6<br />

5<br />

4<br />

3<br />

2<br />

ALT<br />

y = 1.037x – 1.180<br />

R 2 = 0.990 n = 90<br />

0 20 40 60 80 100 120 140 160 180 200<br />

Cobas Mira (U/l)<br />

Cholesterol<br />

y = 0.930x + 0.277<br />

R 2 = 0.998 n = 89<br />

2 3 4 5 6 7 8 9 10 11 12<br />

Cobas Mira (mmol/l)


Konelab <strong>20i</strong> (umol/l)<br />

700<br />

600<br />

500<br />

400<br />

300<br />

200<br />

100<br />

0<br />

Conclusions<br />

The Konelab <strong>20i</strong> demonstrated very good analytical performance throughout the evaluation. Correlations to<br />

routine methods in the laboratory were excellent as well precisions.<br />

The Konelab <strong>20i</strong> analyzer with Konelab reagents turned out to be a cost-efficient and user-friendly analyzing<br />

system with many features for easy and self-guiding operation. The main routines, loading of samples, reagents<br />

and reaction cuvettes can be performed without any workload interruptions. The analyzer demands very few<br />

maintenance operations daily.<br />

We concluded that Konelab <strong>20i</strong> is a reliable analyzer for small- to medium-sized laboratories.<br />

References<br />

Uric acid<br />

y = 0.979x + 21.808<br />

R 2 = 0.976 n = 90<br />

0 100 200 300 400 500 600 700<br />

Cobas Mira (umol/l)<br />

1. Kaihola, H.-L., Rintola, P. and Sandbacka, B: Pienten laboratorioiden kliinisen kemian analysaattorin<br />

koestusohjelma, Moodi 1990: 3: 176-185<br />

Laboratory Medicine 2000, XXVII Nordic Congress of Clinical Chemistry, Bergen, Norway, 4-8 June 2000

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!