17.12.2012 Views

8 Hour Stability

8 Hour Stability

8 Hour Stability

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Application Note:<br />

AN_E0637<br />

Key Words<br />

• ICP-AES<br />

• Environmental<br />

Analysis<br />

• SW-846 Method<br />

6010B<br />

• US EPA<br />

• SW-846 Method<br />

3050B<br />

US EPA SW-846 Method 6010B using the IRIS<br />

Intrepid II ICP-AES<br />

1. Introduction<br />

This Application Note describes the use of the Thermo<br />

Electron Corporation IRIS Intrepid II ICP-AES to carry<br />

out SW-846 Method 6010B compliant analysis. The<br />

method is suitable for the determination of 31 elements<br />

in ground waters, TCLP and EP extracts, industrial and<br />

organic wastes, soils, sludges and sediments. This note<br />

provides data showing compliance with each of the<br />

requirements and highlights the integrated system tools<br />

specifically designed to aid compliance.<br />

2. Background<br />

EPA Historical Facts<br />

In 1970, the United States government established the<br />

Environmental Protection Agency (EPA) in response to<br />

growing public demand for cleaner water, air and land.<br />

Prior to this, the national government was not structured<br />

to deal with pollution that caused harm to human health<br />

and degraded the environment. The EPA was tasked with<br />

repairing the damage already done and moving towards<br />

a cleaner environment. Its mission is to protect human<br />

health and to safeguard the natural environment. The<br />

Agency provides leadership in the nation's<br />

environmental science, research, education and<br />

assessment efforts and works closely with other federal<br />

agencies and local government to develop and enforce<br />

regulations under existing environmental law. The<br />

Agency is responsible for researching and setting national<br />

standards for a variety of environmental programs and<br />

delegates the responsibility for issuing permits, and<br />

monitoring and enforcing compliance, to local<br />

government. Where national standards are not met, the<br />

EPA can issue sanctions and take other steps to assist<br />

local government in reaching the desired levels of<br />

environmental quality. The Agency also works with<br />

industries and all levels of government in a wide variety<br />

of voluntary pollution prevention programs and energy<br />

conservation efforts. Its protocols have also been<br />

adopted as the basis for national or local standards in<br />

many countries around the world.<br />

Office of Solid Waste<br />

The EPA's Office of Solid Waste (OSW) regulates all waste<br />

under the Resource Conservation and Recovery Act<br />

(RCRA). The RCRA's goals are to:<br />

1. Protect the public from the hazards of waste disposal<br />

2. Conserve energy and natural resources by recycling and<br />

recovery<br />

3. Reduce or eliminate waste, and<br />

4. Clean up waste, which may have spilled, leaked, or<br />

been improperly disposed of.<br />

The RCRA tightly regulates all hazardous waste from<br />

production to disposal. The RCRA also controls domestic<br />

and industrial waste. Industrial waste is process waste that<br />

comes from a broad range of operations. Some wastes are<br />

managed by other federal agencies or state laws. Examples<br />

of such wastes are animal waste, radioactive waste, and<br />

medical waste.<br />

The EPA publication SW-846, entitled Test Methods for<br />

Evaluating Solid Waste, Physical/Chemical Methods, is the<br />

OSW's official compendium of over 200 analytical and<br />

sampling methods that have been evaluated and approved<br />

for use for analysis relating to the RCRA regulations. SW-<br />

846 functions primarily as a guidance document setting<br />

forth acceptable, although not required, methods for the<br />

regulated and regulatory communities to use in responding<br />

to RCRA-related sampling and analysis requirements. SW-<br />

846 is a multi-volume document that changes over time as<br />

new information and data are developed. It was first issued<br />

by the EPA in 1980 and is currently in its fourth edition.<br />

Advances in analytical instrumentation and techniques are<br />

continually reviewed by the OSW and incorporated into<br />

periodic updates to SW-846 to support changes in the<br />

regulatory program and to improve method performance<br />

and cost effectiveness. To date, the EPA has finalized<br />

Updates I through IIIA of the SW-846 manual, and the<br />

updated and fully integrated manual contains approximately<br />

3500 pages. The Methods Team of OSW has also made<br />

Draft Updates IVA and IVB available for public discussion.<br />

The SW-846 Method 6010B describes the use of ICP-<br />

AES instrumentation for determining a variety of metallic<br />

elements in aqueous and solid media. In draft edition IVA of<br />

November 2000, updated Method 6010C is under<br />

consideration.


Method 6010B (and draft 6010C)<br />

Both methods are broadly similar and provide guidelines<br />

on general laboratory practices such as sample<br />

preparation, instrument setup, calibration of analytes,<br />

and interference correction equations. They also provide<br />

specific rules on various analytical practices that must be<br />

followed, including elements covered, quality control<br />

practices and instrument validation. Since both methods<br />

are well established and readily available in the public<br />

domain they have become widely adopted as templates<br />

for methodologies used by a host of laboratories<br />

undertaking environmental analysis world-wide. The<br />

aim of the protocol is to ensure a consistently high<br />

quality of analytical data by enforcing compliance with a<br />

variety of stringent instrument and analytical<br />

performance checks.<br />

NOTE: Method 6010C, along with all Draft Update IVA SW-846 methods,<br />

may be downloaded from the EPA OSW website:<br />

http://www.epa.gov/epaoswer/hazwaste/test/up4a.htm<br />

3. Experimental<br />

3.1 Equipment<br />

An IRIS Intrepid II Model XDL ICP-AES (Thermo<br />

Electron Corporation, Franklin, Mass., USA) was used<br />

for the analysis of soil and sludge samples. It was fitted<br />

with a dual view optical design - axial view was used for<br />

elements expected at low concentration where the best<br />

sensitivity is required while radial view was used for<br />

elements at higher concentration or for those elements<br />

that suffer from EIE (Easily Ionised Element) interference.<br />

Internal standards were added on-line using the Internal<br />

Standard Mixing Kit (P/N 13670800).<br />

The instrument conditions used for the analysis are<br />

shown in Table 1 below.<br />

PARAMETER VALUE<br />

RF Forward power 1200 watts<br />

Torch Orientation Dual View<br />

Torch Gas Flow 16 L/min<br />

Auxiliary Gas Flow 0.5 L/min<br />

Mixing Chamber Glass Cyclone<br />

Nebulizer Glass Concentric<br />

Nebulizer Gas 0.65 L/min<br />

Sample Pump Winding 3-stop Orange Tygon<br />

Sample Uptake 1.7 mL/min<br />

Purge Gas Argon<br />

Tank Flow 4 L/min<br />

Nozzle Flow 4 L/min<br />

Camera Flow 1 L/min<br />

Table 1. Intrepid II XDL Parameters<br />

Details of the wavelengths and plasma view<br />

orientations used, as well as the internal standards<br />

employed, are shown in Table 2.<br />

ELEMENT WAVELENGTH PLASMA INTERNAL<br />

(nm) VIEW STANDARD<br />

Ag 328.289 Axial Y 361<br />

Al 308.215 Radial Y360<br />

As 189.042 Axial Y224<br />

B 208.959 Axial Y224<br />

Ba 455.403 Radial Y360<br />

Be 313.107 Radial Y360<br />

Ca 317.933 Radial Y360<br />

Cd 214.438 Axial In230<br />

Co 228.616 Axial In230<br />

Cr 205.552 Axial Y224<br />

Cu 324.754 Axial Y361<br />

Fe 271.441 Radial Y360<br />

In 230.606 Axial Int Std<br />

K 766.491 Radial Y360<br />

Li 670.784 Radial Y360<br />

Mg 279.079 Radial Y360<br />

Mn 257.610 Axial Y361<br />

Mo 202.030 Axial Y224<br />

Na 589.592 Radial Y360<br />

Ni 231.604 Axial In230<br />

P 178.287 Axial In230<br />

Pb 220.353 Axial In230<br />

Sb 206.833 Axial Y224<br />

Se 196.090 Axial Y224<br />

Si 251.612 Radial Y360<br />

Sn 189.989 Axial In230<br />

Ti 334.941 Axial Y361<br />

Tl 190.864 Axial In230<br />

V 292.402 Axial Y361<br />

Y 224.306 Axial Int Std<br />

Y 360.073 Radial Int Std<br />

Y 361.105 Axial Int Std<br />

Zn 206.200 Axial In230<br />

Table 2. Wavelengths and Plasma View used.<br />

3.2 Calibration Solutions<br />

Calibration Solutions<br />

High purity reagents were used throughout. Calibration<br />

standards were prepared in 5%v/v HCl and 1%v/v HNO3<br />

from a range of SPEX ICP standards at concentrations<br />

selected to cover the linear range for each element.<br />

Mixed calibration standards were prepared according<br />

to the conditions laid down in Method 6010B.<br />

SOLUTION ELEMENTS<br />

I Be, Cd, Mn, Pb, Se and Zn<br />

II Ba, Co, Cu, Fe and V<br />

III As and Mo<br />

IV Al, Ca, Cr, K, Na, Ni, Li and Sr<br />

V Ag, Mg, Sb and Tl<br />

VI P<br />

A mixture of 20ppm In and 10ppm Y was used as the<br />

internal standard.


Blanks<br />

Two types of blanks are required for the analysis of<br />

samples prepared by any method other than EPA Method<br />

3050B. The calibration blank is used in establishing the<br />

analytical curve and the method blank is used to identify<br />

possible contamination resulting from either the reagents<br />

(acids) or the equipment used during sample processing<br />

including filtration.<br />

The calibration blank is prepared by acidifying<br />

reagent water to the same concentrations of the acids<br />

found in the standards and samples. A sufficient quantity<br />

must be prepared to flush the system between standards<br />

and samples. The calibration blank will also be used for<br />

all initial (ICB) and continuing calibration blank (CCB)<br />

determinations.<br />

The method blank must contain all of the reagents in<br />

the same volumes as used in the processing of the samples.<br />

The method blank must be carried through the complete<br />

preparation procedure and contain the same acid<br />

concentration in the final solution as the sample solution<br />

used for analysis<br />

Check Solutions<br />

The initial calibration verification (ICV) standard is<br />

prepared by combining compatible elements from a<br />

standard source different from that used to prepare the<br />

calibration standard, and at concentration near the<br />

midpoint of the calibration curve. This standard may also<br />

be purchased commercially.<br />

The continuing calibration verification (CCV)<br />

standard is prepared in the same acid matrix using the<br />

same standards used for calibration, at a concentration<br />

near the mid-point of the calibration curve.<br />

The interference check solution (ICS) is prepared to<br />

contain known concentrations of interfering elements that<br />

will provide an adequate test of the correction factors.<br />

The sample is spiked with the elements of interest,<br />

particularly those with known interferences at the 0.5 to 1<br />

mg/L level.<br />

3.3 Method Development<br />

Interelement Corrections<br />

Due to the complex nature of the matrix experienced with<br />

these types of samples, significant interferences from<br />

spectral overlaps can be observed. As part of the method<br />

development process, high purity solutions of the major<br />

matrix elements, such as Al, Ca, Fe, Mg, Si and P, were<br />

checked for their contribution to the signals of other<br />

analyte elements. After the interfering elements were<br />

identified, their solutions were measured and the<br />

instrument software was used to automatically calculate<br />

the appropriate interference correction factors.<br />

Table 3 lists the major interferences found during this<br />

procedure. A dash indicates that no significant interference<br />

was encountered at the wavelength concerned.<br />

ELEMENT WAVELENGTH (nm) INTERFERING ELEMENT<br />

Ag 328.289 Ca, Fe<br />

Al 308.215 -<br />

As 189.042 -<br />

B 208.959 Al, Fe<br />

Ba 455.403 -<br />

Be 313.107 -<br />

Ca 315.887 -<br />

Cd 214.438 Fe<br />

Co 228.616 Fe<br />

Cr 205.552 Fe<br />

Cu 324.754 Ca, Fe<br />

Fe 271.441 -<br />

In 230.606 -<br />

K 766.491 -<br />

Li 670.784 -<br />

Mg 279.079 -<br />

Mn 257.610 Al, Mg<br />

Mo 202.030 -<br />

Na 589.592 -<br />

Ni 231.604 Al, Fe<br />

P 178.287 -<br />

Pb 220.353 Al, Fe<br />

Sb 206.833 Al, Fe<br />

Se 196.090 Al, Fe<br />

Si 251.612 -<br />

Sn 189.989 -<br />

Ti 334.941 -<br />

Tl 190.864 Mg<br />

V 292.402 Fe<br />

Y 224.306 -<br />

Y 360.073 -<br />

Y 361.105 -<br />

Zn 206.200 Fe, Mg<br />

Table 3. Interelement interferences.<br />

Method 6010B states that interference effects must be<br />

evaluated for each individual instrument and the<br />

interelement corrections should be verified daily by<br />

running an ICS. The measured values must be within 20%<br />

for five consecutive days. They must also be verified and<br />

updated every six months or when changes to the<br />

instrumentation have been made (i.e. changing torch or<br />

nebuliser or plasma conditions).<br />

Method Detection Limits (MDL)<br />

Method Detection Limits were established by measuring a<br />

solution containing all the analytes at concentration levels<br />

which are 3 - 5x the published Instrument Detection<br />

Limits for the IRIS Intrepid II. The solution was analysed<br />

ten times on three separate days and the mean of the three<br />

sigma values from all runs was calculated. Table 4 lists the<br />

MDL’s in mg/L measured in the test and Method 6010B<br />

calls for them to be re-confirmed on a quarterly basis.


ELEMENT WAVELENGTH MDL MAXIMUM<br />

(nm) (µg/L) LINEAR RANGE (mg/L)<br />

Ag 328.289 0.5 10<br />

Al 308.215 11.4 1000<br />

As 189.042 2.1 50<br />

B 208.959 2.2 10<br />

Ba 455.403 0.7 50<br />

Be 313.107 0.1 10<br />

Ca 317.933 2.6 1000<br />

Cd 214.438 0.1 20<br />

Co 228.616 0.2 100<br />

Cr 205.552 0.3 50<br />

Cu 324.754 0.5 50<br />

Fe 271.441 52 1000<br />

K 766.491 30 500<br />

Li 670.784 1.0 50<br />

Mg 279.079 40 1000<br />

Mn 257.610 0.1 20<br />

Mo 202.030 0.5 20<br />

Na 589.592 10 200<br />

Ni 231.604 0.4 50<br />

P 178.287 3.2 20<br />

Pb 220.353 2.1 50<br />

Sb 206.833 2.4 50<br />

Se 196.090 2.8 50<br />

Si 251.612 7.2 50<br />

Sn 189.989 1.8 50<br />

Ti 334.941 0.2 20<br />

Tl 190.864 1.6 100<br />

V 292.402 0.6 50<br />

Zn 206.200 0.2 20<br />

Table 4. Wavelengths, Method Detection Limits and Maximum Linear Range.<br />

Linear Concentration Range<br />

The upper limit of the linear dynamic range must be<br />

established for each analyte and instrument configuration.<br />

The upper range limit should result in an observed signal<br />

no more than 10% below the level extrapolated from the<br />

normal calibration curve. This check should be carried<br />

out whenever instrument changes occur and at six<br />

monthly intervals as a minimum.<br />

This evaluation is very important when using interelement<br />

correction since, if the linear range is exceeded for<br />

the interfering element measurement, the calculated<br />

correction factor will be incorrect and the analytical result<br />

will be in error.<br />

For some elements the plasma is viewed in radial<br />

mode using a weaker line to improve the linear range.<br />

This results in less chance of some samples being overrange<br />

so that more accurate data is achieved. Table 4<br />

shows the maximum linear range figures in mg/L.<br />

Background Correction Points<br />

Method 6010B requires that off-peak background<br />

correction is used for trace element determinations. The<br />

background was measured simultaneously by monitoring<br />

the signal in pixels adjacent to the analyte peak position.<br />

The ideal position of background correction points was<br />

determined by examining spectra in subarrays for a<br />

calibration blank, calibration standard and spiked matrix<br />

solution. The points selected for analysis should take into<br />

account any spectral interferences that may be present.<br />

The documented data for all decisions related to<br />

wavelength choice (where different to the EPArecommended<br />

lines), background correction points,<br />

interelement corrections and MDL’s must be kept on file<br />

and be available for review by the data user or auditor.<br />

4. Results and Discussion<br />

4.1 Initial (ICV) and Continuing (CCV) Calibration<br />

Verification<br />

Method 6010B requires that a very strict quality control<br />

procedure is followed to ensure validity of sample data.<br />

Quality control checks are carried out following<br />

instrument calibration, during sample analysis and at the<br />

end of the analytical run. For the sample data to be<br />

acceptable all checks must meet the required criteria.<br />

The instrument was set up using the parameters shown in<br />

Table 1 and allowed to stabilise for 30 minutes prior to<br />

calibration.<br />

Immediately after calibration, an Initial Calibration<br />

Verification (ICV) solution, calibration blank and<br />

Continuing Calibration Verification (CCV) solution were<br />

run. The calibration blank value must be within 3 times<br />

the Method Detection Limit (MDL) while the calibration<br />

verification solution results must be within 10% of their<br />

calibrated value. In addition, the standard deviation for a<br />

minimum of 2 resamples must be less than 5% for the<br />

data to be acceptable. Typical results are shown in Table 5<br />

below.<br />

Analysis of the CCV solution and calibration blank<br />

was then repeated every 10 samples to ensure that the<br />

instrument remained in calibration.<br />

4.2 Interference Check Solutions<br />

Prior to the start of each sample analysis, interference<br />

check solutions were run to verify that the inter-element<br />

correction factors and background correction points were<br />

still valid.<br />

Interference Check Sample A (ICSA) was prepared<br />

containing 500mg/L Al, Ca, Mg and 200mg/L Fe.<br />

Interference Check Solution AB was then prepared by<br />

spiking this solution at concentrations of 0.05 to 1mg/L<br />

for the analyte elements. The values measured for ICSAB<br />

must be within 20% of the true value for the data to be<br />

acceptable. Data is shown in Table 6 below.


ELEMENT ICV (µg/g) CCV (µg/g)<br />

ACTUAL MEASURED ACTUAL MEASURED<br />

Ag 1 0.96 0.5 0.51<br />

Al 200 202 100 99.7<br />

As 1 0.99 0.5 0.48<br />

B 1 0.99 0.5 0.5<br />

Ba 10 10.3 5 4.97<br />

Be 1 1.02 0.5 0.52<br />

Ca 200 201 100 98.8<br />

Cd 1 1.03 0.5 0.49<br />

Co 10 10.5 5 5.02<br />

Cr 10 10.2 5 5.05<br />

Cu 10 9.80 5 4.98<br />

Fe 200 203 100 101<br />

K 100 98.7 50 49.9<br />

Mg 100 101 50 50.5<br />

Mn 10 9.85 5 4.92<br />

Mo 1 1.02 0.5 0.48<br />

Na 100 102 50 51.2<br />

Ni 10 10.1 5 5.10<br />

P 10 10.2 5 5.07<br />

Pb 10 9.93 5 5.04<br />

Sb 1 0.98 0.5 0.48<br />

Se 1 1.05 0.5 0.53<br />

Ti 10 10.1 5 5.02<br />

Tl 1 0.97 0.5 0.48<br />

V 10 10.4 5 4.89<br />

Zn 10 10.1 5 5.02<br />

Table 5. Calibration QC Checks.<br />

4.3 Correction Factor <strong>Stability</strong> Evaluation<br />

The IRIS Intrepid II XDL spectrometer has a highly<br />

regulated temperature controlled optical system. This<br />

ensures that peak analytical performance is maintained<br />

over extended time periods, despite any fluctuations in the<br />

general laboratory conditions.<br />

An interference correction was set up for Fe on Cd<br />

and then a high purity iron solution was measured over<br />

an 8 hour period in order to assess the amount of drift on<br />

the Cd signal in this time frame. A plot of the data is<br />

shown in Figure 1.<br />

To provide additional evidence of the instrument<br />

stability, a nominal 100 ppb multielement standard was<br />

analysed over an 8 hour period, without any calibration<br />

updates. The plots are shown in Figure 2 and illustrate the<br />

inherent stability of the system.<br />

ELEMENT ICSA ICSAB TARGET VALUE %<br />

(µg/L) (µg/L) (µg/L) RECOVERY<br />

Ag < 0.5 98.7 100 98.7<br />

As -1.3 100.7 100 100.7<br />

B < 2 499 500 99.8<br />

Ba < 0.7 486 500 97.2<br />

Be < 0.1 495 500 99.0<br />

Cd < 0.1 103 100 103<br />

Co < 0.2 489 500 97.8<br />

Cr < 0.3 476 500 95.2<br />

Cu < 0.5 478 500 95.6<br />

Mn < 0.1 512 500 102<br />

Mo < 0.5 478 500 95.6<br />

Ni < 0.4 952 1000 95.2<br />

P < 3 475 500 95<br />

Pb < 2 48.9 50 97.8<br />

Sb < 2 503 500 101<br />

Se < 3 58 50 116<br />

Si < 7 47 50 94<br />

Ti < 0.2 507 500 101<br />

Tl < 2 88.5 100 88.5<br />

V < 0.6 487 500 97.4<br />

Zn < 0.2 1012 1000 101<br />

Table 6. Results for Interference Check Solutions.<br />

4.4 Sample Results<br />

After the successful analysis of the specified quality<br />

control solutions, the actual sample analysis sequence can<br />

begin. With the exception of groundwater samples, all<br />

aqueous and solid matrices require acid digestion prior to<br />

analysis. Recommended dissolution procedures for real<br />

samples are given in EPA Method 3050B or 3051A (the<br />

latter being a microwave-assisted version of the former).<br />

Groundwater samples that have been pre-filtered and<br />

acidified do not need acid digestion.<br />

It is general practice to include at least one Laboratory<br />

Control Sample (LCS) in each run, the LCS normally<br />

being a Standard Reference Material with a similar matrix<br />

to the samples being analysed. In this particular analysis, a<br />

sample of Buffalo River Sediment (NBS SRM2704) was<br />

prepared and used as the LCS.<br />

This sample can also be prepared and analysed in<br />

duplicate, at least once per analysis, in order to check for<br />

contamination problems. The sample was also spiked and<br />

analysed to monitor spike recoveries.<br />

Method 6010B states that duplicate samples must<br />

agree within 20% of each other and spiked samples must<br />

be recovered within 25% of the actual value.


Concentration (ug/L)<br />

110<br />

105<br />

100<br />

95<br />

90<br />

85<br />

80<br />

10:38<br />

11:15<br />

11:44<br />

8 <strong>Hour</strong> Run, drift of Corrected Cadmium with Iron IEC<br />

12:16<br />

12:48<br />

13:14<br />

13:40<br />

EPA Acceptable Range<br />

EPA Acceptable Range<br />

13:57<br />

8 <strong>Hour</strong> <strong>Stability</strong><br />

1 21 41 61 81 101 121<br />

Sample number<br />

ELEMENT MEASURED CONCENTRATION CERTIFIED VALUE<br />

As 23.1µg/g 23.4+/- 0.8<br />

Ba 407µg/g 414 +/- 12<br />

Ca 2.55% 2.60 +/- 0.03<br />

Cd 3.50µg/g 3.45 +/- 0.22<br />

Co 14.3µg/g 14 +/- 0.6<br />

Cr 132µg/g 135 +/- 5<br />

Cu 100.2µg/g 98.6 +/- 5.0<br />

Fe 4.03% 4.11 +/- 0.10<br />

K 2.02% 2.00 +/- 0.04<br />

Mg 1.21% 1.20 +/- 0.02<br />

Mn 547µg/g 555 +/- 19<br />

Na 0.53% 0.547 +/- 0.014<br />

Ni 43.8µg/g 44.1 +/- 3.0<br />

Pb 163µg/g 161 +/- 17<br />

Sb 3.61µg/g 3.79 +/- 0.15<br />

Ti 0.46% 0.457 +/- 0.018<br />

V 93µg/g 95 +/- 4<br />

Zn 440µg/g 438 +/- 12<br />

Table 7. NBS 2704 Buffalo River Sediment Data<br />

14:25<br />

15:12<br />

15:40<br />

15:55<br />

16:25<br />

16:50<br />

17:18<br />

17:42<br />

18:00<br />

2<br />

1.5<br />

1<br />

0.5<br />

0<br />

-0.5 -<br />

-1 -<br />

-1.5 -<br />

-2 -<br />

PPB<br />

As1890<br />

Ba4554<br />

Cd2288<br />

Pb2203<br />

Se1960<br />

Zn2138<br />

Figure 1. <strong>Stability</strong> of<br />

the Fe interelement<br />

correction factor for<br />

Cd<br />

Figure 2. <strong>Stability</strong> of a<br />

100ppb multielement<br />

standard over an 8<br />

hour period<br />

Table 7 compares the measured and certified results<br />

for the SRM2704 Buffalo River Sediment sample,<br />

demonstrating the excellent agreement between them.


Table 8 shows the results for a duplicate analysis test<br />

with the LCS sample. Agreement between duplicates is<br />

generally within 1%, well within the EPA limit of 20%.<br />

ELEMENT ANALYSIS 1 ANALYSIS 2 % DIFFERENCE<br />

As 23.1µg/g 23.3µg/g 0.8<br />

Ba 407µg/g 403µg/g 1.0<br />

Ca 2.55% 2.54% 0.4<br />

Cd 3.50µg/g 3.46µg/g 1.1<br />

Co 14.3µg/g 14.4µg/g 0.7<br />

Cr 132µg/g 131µg/g 0.8<br />

Cu 100.2µg/g 101µg/g 0.8<br />

Fe 4.03% 4.04% 0.3<br />

K 2.02% 2.03% 0.6<br />

Mg 1.21% 1.20% 0.3<br />

Mn 547µg/g 543µg/g 0.8<br />

Na 0.53% 0.54% 1.8<br />

Ni 43.8µg/g 44.1µg/g 0.7<br />

Pb 163µg/g 163µg/g 0.2<br />

Sb 3.61µg/g 3.65µg/g 1.1<br />

Ti 0.46% 0.46% 0.4<br />

V 93.0µg/g 92.3µg/g 0.8<br />

Zn 440µg/g 435µg/g 1.1<br />

Table 8. NBS 2704 Buffalo River Sediment Duplicate Analysis Test<br />

Table 9 shows the recoveries from spiked samples of<br />

the LCS, again all well within the EPA 25% limit.<br />

ELEMENT CERTIFIED VALUE SPIKE CONC MEASURED VALUE % REC<br />

www.thermo.com/elemental<br />

µg/L µg/L µg/L<br />

As 230 50 276 92<br />

Ba 4140 100 4234 94<br />

Cd 34.5 10 43.6 91<br />

Co 140 10 151 110<br />

Cr 1350 50 1402 104<br />

Cu 986 50 1033 94<br />

Mn 5550 100 5652 102<br />

Ni 441 10 450 90<br />

Pb 1610 50 1656 92<br />

Sb 37.9 10 47.0 91<br />

V 950 50 1004 108<br />

Zn 4380 100 4482 102<br />

Table 9. NBS 2704 Buffalo River Sediment Spike Sample Recovery Data<br />

In order to check that the method works for different<br />

sample matrices, a reference sample of a sewage sludge<br />

(BCR146) was analysed using exactly the same procedure.<br />

Table 10 compares the measured and certified results for<br />

the sediment sample, demonstrating the excellent<br />

agreement. Values in brackets are advisory figures only.<br />

USA: 5225 Verona Road<br />

Madison, WI 53711-4495<br />

Tel: +1 800 201 8132, +1 608 276 6100<br />

Fax: +1 608 273 5046<br />

Email: spectroscopy@thermo.com<br />

ELEMENT MEASURED CONCENTRATION CERTIFIED VALUE<br />

Ca 10.2% (10.14%)<br />

Cd 75.8µg/g 77.7 +/- 2.6<br />

Co 11.9µg/g 11.8 +/- 0.7<br />

Cu 928µg/g 934 +/- 24<br />

Fe 1.80% (1.86%)<br />

K 0.49% (0.48%)<br />

Mg 2.01% (1.99%)<br />

Mn 570µg/g 588 +/- 24<br />

Na 0.21% (0.22%)<br />

Ni 276µg/g 280 +/- 18<br />

P 2.52% (2.58%)<br />

Pb 1260µg/g 1270 +/- 28<br />

Zn 4040µg/g 4059 +/- 90<br />

Table 10. BCR146 Sewage Sludge Data<br />

5. Conclusions<br />

The IRIS Intrepid II demonstrates SW-846 Method 6010B<br />

compliant analysis for a range of sample types and easily<br />

copes with the stringent interference checks and AQC<br />

requirements stipulated for use in this method.<br />

The IRIS Intrepid II is the latest generation of IRIS<br />

ICP Spectrometers from Thermo Electron. It incorporates<br />

a new design of the exclusive charge injection device<br />

(CID) detector housed in a highly regulated optical<br />

system. The result of these product developments is an<br />

instrument that is capable of providing extremely low<br />

detection limits and excellent long term stability. This<br />

makes the IRIS Intrepid II ideally suited for cost-effective<br />

elemental analysis in the environmental field.<br />

References<br />

USEPA SW-846 Method 3050B, Revision 2, December 1996.<br />

USEPA SW-846 Method 3051A, Revision 1, January 1998.<br />

USEPA SW-846 Method 6010B, Revision 2, December 1996.<br />

USEPA SW-846 Method 6010C, Revision 3, November 2000.<br />

UK: 19 Mercers Row<br />

Cambridge, CB5 8BZ, UK<br />

Tel: +44 (0)1223 345410<br />

Fax: +44 (0)1223 345411<br />

Email: spectroscopy.uk@thermo.com<br />

Switzerland: En Vallaire Ouest C<br />

case postale, CH-1024, Ecublens<br />

Tel: +41 (0)21 694 71 11<br />

Fax: +41 (0)21 694 71 12<br />

Email: spectroscopy.ch@thermo.com<br />

Thermo Electron Corporation<br />

has direct subsidiary offices in<br />

North America, Europe and<br />

Japan. To complement these<br />

direct subsidiaries, we maintain<br />

a network of representative<br />

organizations throughout the<br />

world. Use this reference list or<br />

visit our Web site to locate the<br />

representative nearest you.<br />

Australia<br />

Tel: +61 (0)2 9898 1244<br />

Austria<br />

Tel: +43 (0)1 333 50 34 0<br />

Belgium<br />

Tel: +32 2482 3030<br />

Canada<br />

Tel: +1 905 890 1034<br />

China<br />

Thermo ARL:<br />

Tel: +86 10 6833 6715<br />

Thermo Elemental:<br />

Tel: +86 10 6592 0232<br />

Thermo Nicolet:<br />

Tel: +86 10 6597 3388 ext 2912<br />

France<br />

Tel: +33 (0)1 39 30 53 00<br />

Germany<br />

Tel: +49 (0)6102 3671 0<br />

Italy<br />

Tel: +39 02 6601 6351<br />

Japan<br />

Thermo Elemental:<br />

Tel: +81 774 201245<br />

Thermo Nicolet:<br />

Tel: +81 45 450 1112<br />

Netherlands<br />

Tel: +31 76 5724840<br />

Nordic<br />

Thermo Nicolet:<br />

Tel: +358 9 3291 00<br />

Thermo ARL:<br />

Tel: +46 (0)8 556 468 31<br />

South Africa<br />

Tel: +27 (0) 11 570 1840<br />

Spain<br />

Tel: +34 (91) 657 49 30<br />

©2003 Thermo Electron Corp.<br />

All rights reserved worldwide.<br />

We make no warranties,<br />

expressed or implied, in<br />

this product summary, and<br />

information is subject to<br />

change without notice.<br />

All product and company<br />

names are property of their<br />

respective owners.<br />

AN_E0637 06/03

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!