Is depth ionization data depth dose?<br />

YES!!!<br />

With the caveat,<br />

�� TCPE exists at the point <strong>of</strong> measurement<br />

measurement.<br />

� the energy spectrum <strong>of</strong> incident photons does not change with the<br />

depth.<br />

�� fluence across the detector remains the same same.<br />

These conditions are met at depths beyond the range <strong>of</strong><br />

contaminant charged particles<br />

However at shallow depth, The contaminants and<br />

secondary electrons have energy spectra that change<br />

rapidly with depth.<br />

� Results in a variation <strong>of</strong> ~10% in restricted mass stopping power<br />

ratio data for water and air.<br />

�Translates into a spatial uncertainty <strong>of</strong> less than 1.5 mm in dose in the<br />

build up region