Characterizing <strong>Clinical</strong> <strong>Photon</strong><br />

Caveats (continued):<br />

<strong>Beams</strong> in 3DRTPS<br />

� One can always use a set <strong>of</strong> beam modeling parameters<br />

to get the best agreement between the computed and<br />

measured beam data in a phantom phantom. .<br />

�However, that would not be a sufficient condition for robust and<br />

accurate beam modeling .<br />

� The value or function used to describe a parameter<br />

should have some physical meaning.<br />

�each parameter used in the dose calculation algorithm should<br />

model the physical reality it represents even if there is less than<br />

perfect agreement between measure and computed data.<br />

� The observed differences <strong>of</strong>ten reflect limitations <strong>of</strong> the dose<br />

computation algorithm