Average temperature measurement with VITO temperature probes<br />

___________________________________________________________________________<br />

2.4 Troubleshooting<br />

2.4.1 Temperature error request (item EM)<br />

This item helps you configuring correctly by the most recent temperature error encountered by the option board.<br />

xx00 No error, value at initialisation<br />

xx11 No reply on initial HART commands Communication error. Check connection level gauge – 762 VITO interface, or<br />

change 762 VITO interface or option board.<br />

xx50 Pt100 (or spot 0) error Brown wire connected to CN1 A-1 (or orange wire connected to CN2B-1) not<br />

connected .<br />

xx51 Spot 1 (or Pt100) error Blue wire to CN2B-1 (or brown wire to CN1A-1) not connected<br />

xx52 Spot 2 error Blue wire to CN2A-2 not connected<br />

xx64 Spot 14 error Blue wire to CN5A-2 not connected<br />

xx65 Spot 15 error Yellow wire to CN5A-1 not connected<br />

xx70 Sub-system error Data from 762 VITO not valid. Reset instrument; if error persist, replace<br />

762 VITO interface<br />

or, open Pt100 wiring (red – brown wires)<br />

xx71 Wrong probe connection Thermocouple wires of VITO sensor connected wrongly (blue and yellow wires)<br />

or, isolation resistance VITO sensor too low (replace VITO probe)<br />

xx77 Wrong Pt100 connection Brown and red wires of VITO sensor connected wrongly<br />

xx80 R test error Measurement on test resistor failed. Reset instrument; if error persist<br />

replace 762 VITO interface or VITO probe<br />

xx84 Standard deviation too large EMC influence (check cable shields) or defective 762 VITO interface<br />

xx92 No data available Communication error between 762 VITO interface and level gauge. Check<br />

wiring; replace 762 VITO interface or option board.<br />

xx98 No temperature sources found No VITO temperature probe (and no RTD) connected<br />

xx: 22 for MPU emulation With error codes xx50 to xx65, connector numbers are mentioned<br />

24 for HPU emulation Refer to Installation guide 762 VITO Interfaces & VITO, MTT, LT and MRT probes<br />

30 for HCU emulation for the location of these connectors.<br />

2.4.2 Temperature status request (item MQ)<br />

Temperature status request item contains four status bytes (Byte 0, Byte 1, Byte 2, Byte 3) of the option board.<br />

For decoding, refer to the ASCII table in appendix A.<br />

Status byte 0:<br />

Contains the characters ‘0’ ÷ ‘F’, indicating the highest submersed spot element of the temperature probe.<br />

At start up: ‘I’.<br />

Status byte 1: Status byte 2:<br />

Bit 0: General temperature fail<br />

Bit 0: Last valid level used<br />

1: Fail in average product temperature 1: Manual level used<br />

2: Fail in average gas temperature 2: Level time out<br />

3: Level exceeds lowest spot element 3: Device not calibrated<br />

4: Level exceeds highest spot element 4: Exceeding differential temp, range<br />

5: Spot element fail (one or more spots defect) 5: Out of specified temperature range<br />

6: 1 6: 1<br />

7: 0 7: 0<br />

Status byte 3:<br />

Bit 0: No previous store command<br />

1-5: 0<br />

6: 1<br />

7: 0<br />

Note:<br />

Only the bits which are set to ‘1’ have an active status.<br />

Instruction manual VITO Page 17