Average temperature measurement with VITO temperature probes<br />
___________________________________________________________________________<br />
2.4 Troubleshooting<br />
2.4.1 Temperature error request (item EM)<br />
This item helps you configuring correctly by the most recent temperature error encountered by the option board.<br />
xx00 No error, value at initialisation<br />
xx11 No reply on initial HART commands Communication error. Check connection level gauge – 762 VITO interface, or<br />
change 762 VITO interface or option board.<br />
xx50 Pt100 (or spot 0) error Brown wire connected to CN1 A-1 (or orange wire connected to CN2B-1) not<br />
connected .<br />
xx51 Spot 1 (or Pt100) error Blue wire to CN2B-1 (or brown wire to CN1A-1) not connected<br />
xx52 Spot 2 error Blue wire to CN2A-2 not connected<br />
xx64 Spot 14 error Blue wire to CN5A-2 not connected<br />
xx65 Spot 15 error Yellow wire to CN5A-1 not connected<br />
xx70 Sub-system error Data from 762 VITO not valid. Reset instrument; if error persist, replace<br />
762 VITO interface<br />
or, open Pt100 wiring (red – brown wires)<br />
xx71 Wrong probe connection Thermocouple wires of VITO sensor connected wrongly (blue and yellow wires)<br />
or, isolation resistance VITO sensor too low (replace VITO probe)<br />
xx77 Wrong Pt100 connection Brown and red wires of VITO sensor connected wrongly<br />
xx80 R test error Measurement on test resistor failed. Reset instrument; if error persist<br />
replace 762 VITO interface or VITO probe<br />
xx84 Standard deviation too large EMC influence (check cable shields) or defective 762 VITO interface<br />
xx92 No data available Communication error between 762 VITO interface and level gauge. Check<br />
wiring; replace 762 VITO interface or option board.<br />
xx98 No temperature sources found No VITO temperature probe (and no RTD) connected<br />
xx: 22 for MPU emulation With error codes xx50 to xx65, connector numbers are mentioned<br />
24 for HPU emulation Refer to Installation guide 762 VITO Interfaces & VITO, MTT, LT and MRT probes<br />
30 for HCU emulation for the location of these connectors.<br />
2.4.2 Temperature status request (item MQ)<br />
Temperature status request item contains four status bytes (Byte 0, Byte 1, Byte 2, Byte 3) of the option board.<br />
For decoding, refer to the ASCII table in appendix A.<br />
Status byte 0:<br />
Contains the characters ‘0’ ÷ ‘F’, indicating the highest submersed spot element of the temperature probe.<br />
At start up: ‘I’.<br />
Status byte 1: Status byte 2:<br />
Bit 0: General temperature fail<br />
Bit 0: Last valid level used<br />
1: Fail in average product temperature 1: Manual level used<br />
2: Fail in average gas temperature 2: Level time out<br />
3: Level exceeds lowest spot element 3: Device not calibrated<br />
4: Level exceeds highest spot element 4: Exceeding differential temp, range<br />
5: Spot element fail (one or more spots defect) 5: Out of specified temperature range<br />
6: 1 6: 1<br />
7: 0 7: 0<br />
Status byte 3:<br />
Bit 0: No previous store command<br />
1-5: 0<br />
6: 1<br />
7: 0<br />
Note:<br />
Only the bits which are set to ‘1’ have an active status.<br />
Instruction manual VITO Page 17