VERICIS Cardiac Workstation R2.X
VERICIS Cardiac Workstation R2.X
VERICIS Cardiac Workstation R2.X
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
3.4 Information Module Definitions<br />
SC Image 3.4.5.3<br />
Overlay Plane Not used<br />
Modality LUT Not Used<br />
VOI LUT 3.4.5.4<br />
SOP Common 3.4.5.5<br />
Please refer to DICOM v3.0 Standard Part 3 for a description of each of the entities and modules<br />
contained within the Secondary Capture Information Object.<br />
3.4.1 Patient Entity Module<br />
3.4.1.1 Patient Module Attributes<br />
Attribute Name Tag Type Attribute Description<br />
Patient Name (0010,0010) 2 Use original image value<br />
Patient ID (0010,0020) 2 Use original image value<br />
Patient Birthdate (0010,0030) 2 Use original image value<br />
Patient Sex (0010,0040) 2 Use original image value<br />
Referenced Patient Sequence (0008,1120) 3 Not used<br />
>Referenced SOP Class UID (0008,1150) 1C Not used<br />
>Referenced SOP Instance UID (0008,1155) 1C Not used<br />
Patient’s Birth Time (0010,0032) 3 Use original image value<br />
Other Patient Ids (0010,1001) 3 Use original image value<br />
Ethnic Group (0010,2160) 3 Not used<br />
Patient Comments (0010,4000) 3 Use original image value<br />
3.4.2 Study Entity Module<br />
3.4.2.1 General Study Module Attributes<br />
Attribute Name Tag Type Attribute Description<br />
Camtronics, Ltd. <strong>Cardiac</strong> WorkStation Conformance Statement Version 1.0<br />
Document. No. 09610-0056 Page 33 of 50 Copyright© Camtronics, Ltd. 1997, 1998, 1999, 2000, 2001