Series Number (0020,0011) 2 Use original image value<br />

Laterality (0020,0060) 2C Not used<br />

Series Date (0008,0021) 3 Not used<br />

Series Time (0008,0031) 3 Not used<br />

Performing Physician’s Name (0008,1050) 3 Use original image value<br />

Protocol Name (0018,1030) 3 Not used<br />

Series Description (0008,103E) 3 Not used<br />

Operator’s Name (0008,1070) 3 Not used<br />

Referenced Study Component<br />

Sequence<br />

(0008,1111) 3 Not used<br />

>Referenced SOP Class UID (0008,1150) 1C Not used<br />

>Referenced SOP Instance UID (0008,1155) 1C Not used<br />

Body Part Examined (0018,0015) 3 Not used<br />

Patient Position (0018,5100) 2C Not used<br />

Smallest Pixel Value in Series (0028,0108) 3 Not used<br />

Largest Pixel Value in Series (0028,0109) 3 Not used<br />

3.4.4 Equipment Entity Module<br />

3.4.4.1 General Equipment Module Attributes<br />

Attribute Name Tag Type Attribute Description<br />

Manufacturer (0008,0070) 2 Use original image value<br />

Institution Name (0008,0080) 3 Use original image value<br />

Institution Address (0008,0081) 3 Use original image value<br />

Station Name (0008,1010) 3 Use original image value<br />

Institutional Department Name (0008,1040) 3 Not used<br />

Manufacturer’s Model Name (0008,1090) 3 Use original image value<br />

Device Serial Number (0018,1000) 3 Not used<br />

Software Version (0018,1020) 3 Not used<br />

Camtronics, Ltd. <strong>Cardiac</strong> WorkStation Conformance Statement Version 1.0<br />

Document. No. 09610-0056 Page 35 of 50 Copyright© Camtronics, Ltd. 1997, 1998, 1999, 2000, 2001