Film Orientation (2010,0040) U User Configurable:<br />
PORTRAIT - default<br />
LANDSCAPE<br />
Film Size ID (2010,0050) U Not used<br />
Magnification Type (2010,0060) U Not used<br />
Smoothing Type (2010,0080) U Not used<br />
Border Density (2010,0100) U User Configurable:<br />
WHITE – default, BLACK<br />
Empty Image Density (2010,0110) U User Configurable:<br />
Minimum Density (2010,0120) U Not used<br />
Maximum Density (2010,0130) U Not used<br />
WHITE – default, BLACK<br />
Trim (2010,0140) U User Configurable:<br />
Configuration Information (2010,0150) U Not used<br />
Illumination (2010,015E) U Not used<br />
Reflected Ambient Light (2010,0160) U Not used<br />
Requested Resolution ID (2020,0050) U Not used<br />
YES - default, NO<br />
4.3 Image Box Pixel Presentation Module Attributes<br />
Polarity<br />
Attribute Name Tag Type Attribute Description<br />
(2020,0020)<br />
U User Configurable:<br />
NORMAL – default<br />
REVERSE<br />
Magnification Type (2010,0060) U Not used<br />
Smoothing Type (2010,0080) U Not used<br />
Configuration Information (2010,0150) U Not used<br />
Requested Image Size (2020,0030) U Not used<br />
Camtronics, Ltd. <strong>Cardiac</strong> WorkStation Conformance Statement Version 1.0<br />
Document. No. 09610-0056 Page 41 of 50 Copyright© Camtronics, Ltd. 1997, 1998, 1999, 2000, 2001