4 Print Management Specific Modules<br />

4.0 Introduction<br />

The following sections specify the default and User optional attributes that can be set in the<br />

modules used for Print Management.<br />

4.1 Basic Film Session Presentation Module Attributes<br />

Attribute Name Tag Type Attribute Description<br />

Number of Copies (2000,0010 U User Configurable:<br />

1 (default ) - 10<br />

Print Priority (2000,0020) U Not used<br />

Medium Type (2000,0030) U Not used<br />

Film Destination (2000,0040) U Not used<br />

Film Session Label (2000,0050) U Not used<br />

Memory Allocation (2000,0060) U Not used<br />

Owner ID (2100,0160) U Not used<br />

4.2 Basic Film Box Presentation Module Attributes<br />

Attribute Name Tag Type Attribute Description<br />

Image Display Format (2010,0010) U STANDARD\1,1 - default<br />

STANDARD\1,2<br />

STANDARD\2,1<br />

STANDARD\2,2<br />

STANDARD\3,2<br />

STANDARD\3,3<br />

STANDARD\3,4<br />

STANDARD\4,3<br />

STANDARD\4,4<br />

Annotation Display Format ID (2010,0030) U Not used<br />

Camtronics, Ltd. <strong>Cardiac</strong> WorkStation Conformance Statement Version 1.0<br />

Document. No. 09610-0056 Page 40 of 50 Copyright© Camtronics, Ltd. 1997, 1998, 1999, 2000, 2001