- 28 -<br />

MESSAGES<br />

The following messages list is general: some of them could not be used in your chamber. Note that messages are<br />

associated to bits, so the term “2052.3” means “bit number 3 (counting from 0) at address 2052”<br />

ADDRESS MEANING<br />

2052.0 Protection switches<br />

2052.1 Maximum temperature User<br />

2052.2 Minimum temperature User<br />

2052.3 Low Stage Massimum pressure<br />

2052.4 Maximum pressure High Stage<br />

2052.5 Thermal protection low stage <strong>com</strong>pressor<br />

2052.6 Thermal protection high stage <strong>com</strong>pressor<br />

2052.7 Maximum temperature service<br />

2052.8 Water lack<br />

2052.9 Safety temperature switches<br />

2052.10 General power lack<br />

2052.11 Equipment off<br />

2052.12 Low stage oil pressure switch<br />

2052.13 High stage oil pressure switch<br />

2052.14 Maximum temperature glycol<br />

2052.15 Minimum level glycol<br />

2053.0 Maximum temperature Steam generator<br />

2053.1 Inverter Alarm<br />

2053.2 Air lack<br />

2053.3 Flow switch<br />

2053.4 Minimum temperature glycol<br />

2053.5 Minimum pressure pump glycol<br />

2053.6 Maximum temperature discharge low stage<br />

2053.7 Maximum temperature discharge high stage<br />

2053.8 Door open<br />

2053.9 Minimum level water<br />

2053.10 Air differential pressure switch<br />

2053.11 Smoke detector<br />

2053.12 Salty solution lack<br />

2053.13<br />

2053.14<br />

2053.15<br />

2054.0<br />

2054.1<br />

2054.2<br />

2054.3<br />

2054.4<br />

2054.5<br />

2054.6<br />

2054.7<br />

2054.8<br />

2054.9<br />

2054.10<br />

2054.11<br />

2054.12<br />

2054.13<br />

2054.14<br />

2054.15<br />

2055.0 Critical alarm (The ladder writes)<br />

2055.1 Power lack (The ladder writes)<br />

2055.2 Resume from critical alarm (The firmware writes)<br />

2055.3<br />

2055.4 programme abort into execution (The firmware writes)<br />

2055.5 Natural end of the programme into execution (The firmware writes)<br />

2055.6 Resume time elapsed<br />

2055.7 Critical deviation<br />

2055.8 Not critical deviation<br />

2055.9<br />

2055.10<br />

<strong>Serial</strong> <strong>com</strong>munications protocol vers. 2.02 Angelantoni Industrie Spa Massa Martana (PG) Italy