Call 1-877-853-9755<br />

36.98 36.5 37.0<br />

37.01 38.9 37.0<br />

I<br />

O<br />

2-Channel Economy Low-noise Heating<br />

Controller, with digital RS232 interface<br />

TC-1-100<br />

Low electrical noise, heating and cooling temperature controller <strong>for</strong> microscope stages. A simplified ver-<br />

sion of TC2-80-150 model. Flexible controls <strong>for</strong> stable operation. Multiple temperature probes to choose<br />

<strong>for</strong> feedback. Can be used with objective heaters. Can be used with flow control and perfusion systems.<br />

Most heating and cooling stages work as inline solution pre-heaters. Can be used with flow control and<br />

perfusion systems. The second channel is usually used <strong>for</strong> objective heaters, or incubator lids. Requires<br />

connecting cables. An external temperature probe might be needed, to monitor bath temperature <strong>for</strong> example.<br />

External probes are plastic-encapsulated: no metal ions leakage into solutions. Requires an external<br />

power supply: 12V is suitable <strong>for</strong> most small heating elements and objective heaters, 18V is required <strong>for</strong><br />

miniature incubators, 24V is required <strong>for</strong> cooling stages.<br />

• Range from -80 to 150°C with accuracy<br />

0.1°C<br />

• Temperature Set manually or externally<br />

• Built-in overheating protection<br />

• Temperature probes: miniature 0.87mm<br />

(fits small volume chambers)<br />

• Feedback: Stage sensor or External probe<br />

• RS232 port <strong>for</strong> programmed temperature<br />

changes<br />

• Analog Input to set temperature changes<br />

• Analog Output to monitor temperature<br />

• Automatically switches from Heating to<br />

Cooling (<strong>for</strong> BTC-1-100)<br />

• Standby mode activated manually or by<br />

external TTL signal<br />

• No vibrations during imaging and recording<br />

- no internal fan<br />

• Dimensions: 6.5 x 3 x 9in.<br />

• Stability: 0.01°C<br />

• Settings: 3-tier flexible, allow to stabilize<br />

temperature in different sample volumes and<br />

heating stage sizes<br />

• Output: 5A max per channel, 165W with<br />

35V input<br />

• <strong>High</strong> Power Mode: 54A outtput, 5.4kW with<br />

100 V input (installed on request)<br />

2-Channel Economy Temperature Controller, low noise<br />

Catalog No. Description Price<br />

TC-1-100 2-Channel Temperature Controller, high stability, no electrical<br />

noise<br />

$495<br />

TC-TP 1-Channel Temperature Controller, range room - 50°C $95<br />

TC1-TCR Replacement Cable Assembly <strong>for</strong> TC-1-100 controller $95<br />

TC2-18V External Power Supply, 18V required <strong>for</strong> miniature incubators<br />

$95<br />

TC2-12V External Power Supply, 12V suitable <strong>for</strong> small heating<br />

stages<br />

$95<br />

TC2-24V External Power Supply, 24V suitable <strong>for</strong> cooling stages $95<br />