5.2 Rheometer Basic Instrument

5.2.1 Rheometer RHEOTEST ® RN with an automatic distance adjustment (Figure 1)

The rheometer (Figure 1) consists of a measuring drive (1), which is fastened by means of the knurled screw

(2) on the stand. The test drive is equipped with a quick coupling (3) for measuring systems like rotors, cones

or plates. With this quick coupling measurement systems can be changed quickly and easily.

The coupling (4) is used for fixing the adapter (6), for cone and plate measuring systems (or for fixing

measuring cups of cylinder measuring systems). The mechanism of the adapter is described in Adapter for a

Rheometer with an automatic distance adjustment. The coupling of the adapter or measuring cup is performed

by turn of the ball handle (5).

For measuring of samples, a Peltier-temperated basic plate (12) including located exchangeable plate (7) is

used. The basic plate is firmly connected with the basic plate guide (10). The basic plate guide is fixed by a

knurled screw (9) at the guidance rod of the stand. The stand (11) allows a stepless adjustment of the working

height of the measuring drive.

In the case of a device with an automatic distance adjustment, a two-sided temperature control of the sample

is carried out. The plate-plate distance setting is a part of the job to perform the measurements and is done

automatically.

Figure 1 Basic Instrument with an Automatic Distance Adjustment