9. Place the sample centric on the basic plate. Application of the sample depends on kind of different

substances by dropping or covering at the center of the basic plate (7).

Note: The gap must be completely filled.

10. Trim the sample. The sample can be trimmed with a spatula or a trimming tool (available on request) -

the excess material to be measured will be removed. The trimming tool is used, as shown on the

Figure 16.

Figure 16 Trimming of the Sample with a Trimming Tool

11. Lay the stripper gripper flat on the basic plate. The inward facing ends are applied to the measuring

plate. By simultaneously pressing the brackets and pulling forward the tool, the excess material is

removed.

12. To ensure all-sided trimming of the sample, the opening of the adapter can be rotated forward (see

Figure 15). After trimming, the window is to turn back again to make ready for measurement. After

trimming, the opening has to be turned back again to make ready for measurement.

13. After clicking OK, the program asks you to set the distance between the plates (e.g. 1,00 mm) and will

display:

14. After clicking OK the measurement starts. The progress of the measurement can be seen in the

information window. When the measured data came (see Information window), they can be also

displayed in the Chart and in the Table. The measured values can be viewed and / or evaluated in the

Report.

15. After the measurement is completed, loosen the knurled screw (2), simultaneously lift the quick

connecting coupling (3) and move the measuring drive upwards. Clean the measuring and basic

plates.

16. Subsequently, new sample is applied and after the plate was hitched again, a new measurement can

be started.