9.2.2 Settings
Further job’s settings.
1 – Suppress Waittime in data: The first measured value is transferred from rheometer RHEOTEST ® to
computer after defined time depending upon method’s parameters used Trigger time and Time raster.
Measured values could be also presented / displayed starting from the moment when time = 0 with the help of
“Suppress Waittime in data”.
2 – Show Sample Collection: Data from the data list will be also shown at a plot in parallel to the actual data.
Data of the data list are set in chart settings. Data of the data list will be always shown with the line thickness
XS and appeared in legend with their set names. Colors and symbols correspond to colors and symbols of the
actual data.
3 – Osci-Correction: There is a possibility to correct the measuring data during data processing of oscillation
tests; in particular J-correction assumes inertia correction.
4 – Standard temperature: This temperature is set at the following conditions:
• if thermostat is used and it is ready for operation;
• if thermostat could assure this temperature;
• after finishing the measurement;
• after program start if it is indicated in the user settings.
5 – Copy Job to distant folder: The Software RN 5 Manager enables the user to observe / to monitor the
development of the measurement on a computer that different from the one for carrying out the measurement.
For this the function “Copy Job to distant folder” should be selected on the computer that controls the
measurement. Further an appropriate directory (distant folder) should be created and chosen, see for it
section User Settings. Furthermore the View distant job function is to mark on the distant computer.
6 – Show complete during analysis: This option is active in that case only when data list for this job will shown
also. This option acts during one measurement. Chart will be shown and scaled in accordance with the actual
data, if this option is inactive. This concerns X- and Y- axes.
Chart presentation and scaling is defined by the data from the data list before beginning of the measurements,
if this option is active. It means that, if the data list includes measuring results during 300 s the measuring
parameters will be shown completely during this 300 s. Actually, measured data will be added to this image
after analysis.
7 – Set temperature before analysis: For temperature-controlled tests this function enables to set the
necessary temperature for the analysis. If this function is marked, the window for temperature input arises
before measuring.
The temperature query (enter target temperature) is carried out before each test, so that a new temperature
can be set for each individual test. If the job name contains the components: “BINDER”, “RTFO”, “PAV”,
“PNST”, “T_SWEEP”, the query is omitted during the measurement before each individual test.
If the temperature control is activated in the method only in the first test (Const or Jump) and not in the
following tests (none), the query is omitted during the measurement before each individual test. The
temperature is however measured.