The electrode-holder cord must be placed so as to<br />
avoid any contact with the pati<strong>en</strong>t or other<br />
conductors.<br />
Active electrodes that are not being used temporarily<br />
must be kept well away from the pati<strong>en</strong>t.<br />
The pati<strong>en</strong>t should never <strong>com</strong>e into contact with<br />
metal parts connected to the earth (operating table,<br />
support, etc.).<br />
Contact betwe<strong>en</strong> the practitioner's skin and that of<br />
the pati<strong>en</strong>t should be avoided.<br />
Simultaneous use on a pati<strong>en</strong>t of a monitoring device<br />
equipped with needle electrodes and an electrosurgical<br />
device is not re<strong>com</strong>m<strong>en</strong>ded; if the need<br />
arises, the electrodes in the monitoring device must<br />
be placed as far as possible from the electro-surge<br />
electrode.<br />
Monitoring systems that include high-frequ<strong>en</strong>cy<br />
curr<strong>en</strong>t limiters are re<strong>com</strong>m<strong>en</strong>ded.<br />
During surgical procedures in which the highfrequ<strong>en</strong>cy<br />
curr<strong>en</strong>t could flow through parts of the<br />
body with a relatively small straight surface section,<br />
the use of bipolar methods may be desirable in order<br />
to avoid unwanted coagulation.<br />
3. 4 CAPACITIVE COUPLER<br />
Make sure that the <strong>en</strong>tire surface of the capacitive<br />
coupler is in contact with the pati<strong>en</strong>t's clothed body.<br />
The capacitive coupler should be placed betwe<strong>en</strong> the<br />
d<strong>en</strong>tal chair and the clothed pati<strong>en</strong>t.<br />
The <strong>en</strong>tire surface of the capacitive coupler should<br />
be connected as near as possible to the operating<br />
area, so that the curr<strong>en</strong>t's shortest path betwe<strong>en</strong> the<br />
active electrode and the capacitive coupler is not via<br />
the heart.<br />
3. 5 OUTPUT POWER<br />
The output power must be the lowest possible for the<br />
purpose required.<br />
Abnormally low power can be caused by a bad<br />
capacitive coupler connection. Before increasing the<br />
power, the operator should check the connections of<br />
the differ<strong>en</strong>t electrodes.<br />
The power level delivered by the SERVOTOME ® may vary<br />
according to numerous parameters (operating speed,<br />
histological variables, thickness and material of<br />
clothing, etc.).<br />
A malfunction may cause an abnormal increase in the<br />
output power.<br />
3. 6 ENVIRONMENT<br />
Due to the high-frequ<strong>en</strong>cy rays, there may be some<br />
disturbance caused to other electrical equipm<strong>en</strong>t<br />
wh<strong>en</strong> the SERVOTOME ® is in use.<br />
The SERVOTOME ® is not designed to function in the<br />
pres<strong>en</strong>ce of an anesthetic gas.<br />
The operator must be made aware of the risks<br />
involved in the use of high-frequ<strong>en</strong>cy equipm<strong>en</strong>t<br />
wh<strong>en</strong> flammable gases are used (explosion hazard).<br />
For pati<strong>en</strong>ts and/or users carrying a pacemaker or<br />
any other active implant, there is a pot<strong>en</strong>tial risk due<br />
to possible interfer<strong>en</strong>ce affecting the pacemaker or<br />
damage to it. If there is any doubt, seek the advice<br />
or approval of suitably qualified persons.<br />