The electrode-holder cord must be placed so as to<br />

avoid any contact with the pati<strong>en</strong>t or other<br />

conductors.<br />

Active electrodes that are not being used temporarily<br />

must be kept well away from the pati<strong>en</strong>t.<br />

The pati<strong>en</strong>t should never <strong>com</strong>e into contact with<br />

metal parts connected to the earth (operating table,<br />

support, etc.).<br />

Contact betwe<strong>en</strong> the practitioner's skin and that of<br />

the pati<strong>en</strong>t should be avoided.<br />

Simultaneous use on a pati<strong>en</strong>t of a monitoring device<br />

equipped with needle electrodes and an electrosurgical<br />

device is not re<strong>com</strong>m<strong>en</strong>ded; if the need<br />

arises, the electrodes in the monitoring device must<br />

be placed as far as possible from the electro-surge<br />

electrode.<br />

Monitoring systems that include high-frequ<strong>en</strong>cy<br />

curr<strong>en</strong>t limiters are re<strong>com</strong>m<strong>en</strong>ded.<br />

During surgical procedures in which the highfrequ<strong>en</strong>cy<br />

curr<strong>en</strong>t could flow through parts of the<br />

body with a relatively small straight surface section,<br />

the use of bipolar methods may be desirable in order<br />

to avoid unwanted coagulation.<br />

3. 4 CAPACITIVE COUPLER<br />

Make sure that the <strong>en</strong>tire surface of the capacitive<br />

coupler is in contact with the pati<strong>en</strong>t's clothed body.<br />

The capacitive coupler should be placed betwe<strong>en</strong> the<br />

d<strong>en</strong>tal chair and the clothed pati<strong>en</strong>t.<br />

The <strong>en</strong>tire surface of the capacitive coupler should<br />

be connected as near as possible to the operating<br />

area, so that the curr<strong>en</strong>t's shortest path betwe<strong>en</strong> the<br />

active electrode and the capacitive coupler is not via<br />

the heart.<br />

3. 5 OUTPUT POWER<br />

The output power must be the lowest possible for the<br />

purpose required.<br />

Abnormally low power can be caused by a bad<br />

capacitive coupler connection. Before increasing the<br />

power, the operator should check the connections of<br />

the differ<strong>en</strong>t electrodes.<br />

The power level delivered by the SERVOTOME ® may vary<br />

according to numerous parameters (operating speed,<br />

histological variables, thickness and material of<br />

clothing, etc.).<br />

A malfunction may cause an abnormal increase in the<br />

output power.<br />

3. 6 ENVIRONMENT<br />

Due to the high-frequ<strong>en</strong>cy rays, there may be some<br />

disturbance caused to other electrical equipm<strong>en</strong>t<br />

wh<strong>en</strong> the SERVOTOME ® is in use.<br />

The SERVOTOME ® is not designed to function in the<br />

pres<strong>en</strong>ce of an anesthetic gas.<br />

The operator must be made aware of the risks<br />

involved in the use of high-frequ<strong>en</strong>cy equipm<strong>en</strong>t<br />

wh<strong>en</strong> flammable gases are used (explosion hazard).<br />

For pati<strong>en</strong>ts and/or users carrying a pacemaker or<br />

any other active implant, there is a pot<strong>en</strong>tial risk due<br />

to possible interfer<strong>en</strong>ce affecting the pacemaker or<br />

damage to it. If there is any doubt, seek the advice<br />

or approval of suitably qualified persons.<br />